DALL-MILES STAINLESS STEEL BEADED CABLE

K961569 · Howmedica Corp. · JDQ · Jun 24, 1996 · Orthopedic

Device Facts

Record IDK961569
Device NameDALL-MILES STAINLESS STEEL BEADED CABLE
ApplicantHowmedica Corp.
Product CodeJDQ · Orthopedic
Decision DateJun 24, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3010
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Dall-Miles Beaded Cable is intended to be used for spinal wiring, sternotomy applications, cerclage procedures, and trauma surgery of the shoulder, elbow, knee, hip, or ankle. In addition, these cables may be used for supplementary fixation when used with bone plates and screws and for the stabilization of bone graft material.

Device Story

Stainless steel cable with electron-beam-melted bead on one end; modification of existing cable design. Used with one-sided tensioner in space-constrained surgical sites where two-sided tensioners are unsuitable. Operates as bone fixation cerclage; provides mechanical stabilization for bone segments or grafts. Used by surgeons in OR settings for orthopedic and trauma procedures. Output is physical fixation of bone/graft; assists in fracture reduction and stabilization; facilitates healing by maintaining bone alignment.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Stainless steel cable; electron-beam-melted bead termination. Mechanical fixation device. No software or energy source.

Indications for Use

Indicated for patients requiring spinal wiring, sternotomy, cerclage procedures, trauma surgery of the shoulder, elbow, knee, hip, or ankle, supplementary fixation with bone plates/screws, or stabilization of bone graft material.

Regulatory Classification

Identification

A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} JUN 24 1996 K961569 510(k) Summary Proprietary Name: Dall-Miles Stainless Steel Beaded Cable Common Name: Bone Fixation Cerclage Classification Name and Reference: Bone Fixation Cerclage 21 CFR 888.3010 Proposed Regulatory Class: Class II Device Product Code: OR(87) JDQ For information contact: Frank Maas Manager, Regulatory Affairs Howmedica Inc. 359 Veterans Boulevard Rutherford, NJ 07070 (201) 507-7875 - Phone (201) 507-6870 - Fax Date Summary Prepared: April 17, 1996 The Dall-Miles Stainless Steel Beaded Cable is a modification of the existing stainless steel cable. A bead is formed on one end of the cable by melting it with an electron beam. This modification allows the cable to be used with a one-sided Tensioner in those applications where a conventional two-sided Tensioner will not fit. The Dall-Miles Beaded Cable is intended to be used for spinal wiring, sternotomy applications, cerclage procedures, and trauma surgery of the shoulder, elbow, knee, hip, or ankle. In addition, these cables may be used for supplementary fixation when used with bone plates and screws and for the stabilization of bone graft material. The substantial equivalence of the Beaded Cable is based on the equivalence in intended use, materials, design, operational principles, and indications and contraindications to Howmedica's Dall-Miles Trochanter Cable Grip System (K844068 & K900926) and Dall-Miles Suture/Beaded Cable (K953818), Smith & Nephew Richards' Ries Cable System (K875156), and Depuy's Control Cable System (K934557).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...