Exeter V40 Femoral Stem Hip System, Orthinox V40 Femoral Heads

K173499 · Stryker Orthopaedics · JDG · Mar 19, 2018 · Orthopedic

Device Facts

Record IDK173499
Device NameExeter V40 Femoral Stem Hip System, Orthinox V40 Femoral Heads
ApplicantStryker Orthopaedics
Product CodeJDG · Orthopedic
Decision DateMar 19, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3360
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Exeter® V40™ Femoral Stem Hip System is intended for use in total or hemi hip replacement. It is intended for cemented use only. The Exeter® V40TM Femoral Stem Hip System is indicated for: · noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; - · rheumatoid arthritis: - · correction of functional deformity; · revision procedures where other treatments or devices have failed; and, · treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

Device Story

Exeter V40 Femoral Stem Hip System and Orthinox V40 Femoral Heads are sterile, single-use orthopedic implants for cemented total or hemi-hip arthroplasty. Stems feature tapered distal design with V40 neck taper, available in various offsets and sizes. Used by orthopedic surgeons in clinical/OR settings to replace damaged femoral components. Implants restore joint function and mobility in patients with degenerative disease or fractures. Output is physical structural support for the hip joint. Device is implanted via surgical procedure; clinical decision-making relies on surgeon assessment of patient anatomy and pathology. Benefits include pain relief and improved joint function.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including range of motion analysis, femoral stem neck and body fatigue testing (ISO 7206-4:2010, ISO 7206-6:2013), and bacterial endotoxin testing (ANSI/AAMI ST72:2011). MRI compatibility established via reference to K153345.

Technological Characteristics

Cemented femoral hip prosthesis. Materials, design, and operational principles identical to predicate Exeter V40 system. Features tapered distal stem and V40 neck taper. Bench testing performed per ISO 7206-4 and ISO 7206-6. Sterilization and pyrogenicity (Endotoxin < 20EU/Device) verified per ANSI/AAMI ST72:2011.

Indications for Use

Indicated for patients requiring total or hemi hip replacement due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, failed prior treatments/devices, or proximal femur fractures (nonunion, femoral neck, trochanteric) with head involvement.

Regulatory Classification

Identification

A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Stryker Orthopaedics Kirsten Lehmuller Staff Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430 March 19, 2018 ## Re: K173499 Trade/Device Name: Exeter V40 Femoral Stem Hip System, Orthinox V40 Femoral Heads Regulation Number: 21 CFR 888.3360 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metallic Cemented Or Uncemented Prosthesis Regulatory Class: Class II Product Code: JDG, JDI, KWY, LZO Dated: February 16, 2018 Received: February 20, 2018 Dear Kirsten Lehmuller: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K173499 #### Device Name Exeter V40 Femoral Stem Hip System, Orthinox V40 Femoral Heads Indications for Use (Describe) Exeter® V40TM Femoral Stem Hip System (includes Orthinox V40 Femoral heads) The Exeter® V40™ Femoral Stem Hip System is intended for use in total or hemi hip replacement. It is intended for cemented use only. The Exeter® V40TM Femoral Stem Hip System is indicated for: · noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; - · rheumatoid arthritis: - · correction of functional deformity; · revision procedures where other treatments or devices have failed; and, · treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ ## 510(k) Summary | Sponsor | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430 | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Kirsten Lehmuller<br>Staff Regulatory Affairs Specialist<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>201-831-5903 | | Alternate Contact | Kristen Meany<br>Sr. Regulatory Affairs Manager<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>201-972-9164 | | Date Prepared: | November 9, 2017 | | Proprietary Name: | Exeter® V40™ Femoral Stem Hip System | | Common Name: | Artificial Hip Replacement | | Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis 21 CFR §888.3353 | | | Hip joint metal/polymer semi-constrained cemented prosthesis 21 CFR<br>§888.3350 | | | Hip joint femoral (hemi-hip) metallic cemented or uncemented<br>prosthesis 21 CFR §888.3360 | | | Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented<br>prosthesis 21 CFR §888.3390 | | Product Codes: | JDI, LZO, JDG, KWY | ## Legally Marketed Device to Which Substantial Equivalence is Claimed: Exeter V40 Femoral Stems – K011623, K110290, K121308, K153345 Accolade HFx Femoral Stems - K051741, K121308 {4}------------------------------------------------ Biomet Sirius Femoral Hip Stem - K130610, K142295 Device Description: The Exeter® V40™ Femoral Stems/Orthinox Femoral Heads are sterile, single-use components intended for cemented total and hemi-hip arthroplasty. The femoral stems have a tapered distal portion with a V40™ neck taper. The stems are available in various offsets and body sizer. The Exeter Hip System continues to be manufactured from the same materials, the same design and the same manufacturing process as the previously cleared Exeter® V40™ femoral stems. Intended Use: The subject devices are intended for use in total or hemi hip arthroplasty. The subject devices are intended for cemented use only. Indications: Exeter® V40™ Femoral Stem Hip System (includes Orthinox V40 Femoral heads) The Exeter® V40™ Femoral Stem Hip System is intended for use in total or hemi hip replacement. It is intended for cemented use only. The Exeter® V40TM Femoral Stem Hip System is indicated for: · noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis: - rheumatoid arthritis: - · correction of functional deformity; - · revision procedures where other treatments or devices have failed; and, - · treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Summary of Technological Characteristics: The technological characteristics (material, design, and operational principles) of the Exeter® V40™ Hip System are similar or identical to the predicate devices. The purpose of this submission is to introduce additional stem lengths to the Exeter V40 Hip System that are within the cleared stem lengths range, add additional Orthinox femoral heads, and update compatibility for compatible usage with unipolar and bipolar femoral heads. Non-Clinical Testing: The following non-clinical laboratory testing was performed to determine substantial equivalence: {5}------------------------------------------------ - 0 Range of Motion Analysis - Femoral Stem Neck and Body Fatigue, ISO 7206-4 (2010)and ISO 7206-6 (2013) o - Bacterial endotoxin testing (BET) as specified in ANSVAAMI ST72:2011 and European 0 Pharmacopoeia Chapter 2.6.14 was used for pyrogenicity testing to achieve an Endotoxin limit of < 20EU/Device and < 20 IU/Device respectively. - Magnetic Resonance Imaging Analysis: The Exeter Hip System has previously established MRI compatibility in pre-market notification K153345 where worst case combinations were presented. The additional femoral stems do not present a new worst case condition. Clinical Testing: Clinical testing was not required as a basis for substantial equivalence. Conclusion: Based upon a comparison of intended use, materials, summary of technological characteristics and performance testing, the Exeter® V40™ Femoral Stem Hip System is substantially equivalent to the predicate devices identified in this premarket notification.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...