LEIBINGER - LUHR SMALL ORTHOPEDIC BONE SCREWS
Device Facts
| Record ID | K963739 |
|---|---|
| Device Name | LEIBINGER - LUHR SMALL ORTHOPEDIC BONE SCREWS |
| Applicant | Howmedica Leibinger, Inc. |
| Product Code | HWC · Orthopedic |
| Decision Date | Jan 22, 1997 |
| Decision | SN |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
This submission describes a system of small bone screws. It is intended for use in internal fixation of small bones including the hand and foot secondary to trauma or for reconstruction.
Device Story
Small bone screw system for internal fixation of hand and foot bones; used in trauma or reconstruction procedures. Screws range 0.8-3.5 mm diameter, 2-40 mm length. Single-use metallic fixation appliances. Operated by surgeons in clinical/OR settings. Provides mechanical stabilization of bone fragments to facilitate healing.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Metallic bone fixation screws; material Ti6Al4V; diameters 0.8-3.5 mm; lengths 2-40 mm; single-use; mechanical fixation principle.
Indications for Use
Indicated for internal fixation of small bones in the hand and foot, secondary to trauma or for reconstruction.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Howmedica Luhr® Fixation System (K761228, K882454, K913355, K935448)
- Wurzburg® Bone Screws (K854886)
- Steinhauser® Bone Screws (K862482)
- Howmedica Leibinger Forefoot Reconstruction System (K961485)