CANNULATED PLUS HIP SCREW (STERILE)

K955837 · Wrightmedicaltechnologyinc · HWC · Mar 4, 1996 · Orthopedic

Device Facts

Record IDK955837
Device NameCANNULATED PLUS HIP SCREW (STERILE)
ApplicantWrightmedicaltechnologyinc
Product CodeHWC · Orthopedic
Decision DateMar 4, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Cannulated PLUS™ Hip Screw (Sterile) is intended to be used in the fixation of cancellous bone, for treatment of conditions such as, but not limited to: 1. femoral neck fractures; 2. arthrodesis of the hip; 3. femoral condyle fractures; 4. tibial plateau fractures; 5. ankle malleolar fractures; or 6. elbow fractures

Device Story

Cannulated bone screw; optional beveled washer for osteoporotic bone; available in various sizes. Used by orthopedic surgeons for internal fixation of cancellous bone fractures or arthrodesis. Device provides mechanical stabilization of bone fragments to facilitate healing.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Cannulated bone screw and optional washer; material: stainless steel; sterile.

Indications for Use

Indicated for patients requiring fixation of cancellous bone, including femoral neck fractures, hip arthrodesis, femoral condyle fractures, tibial plateau fractures, ankle malleolar fractures, and elbow fractures.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K955837 ![img-0.jpeg](img-0.jpeg) # WRIGHT ## MEDICAL TECHNOLOGY, INC. 5677 AIR INF ROAD ARLINGTON, IN 38002 901 877 9971 **510(k) SUMMARY** Contact Person: Judy English Date Prepared: December 22, 1995 Trade Name: Cannulated PLUS™ Hip Screw (Sterile) Common Name: Hip Bone Screw Classification: II Predicate Device: Cannulated PLUS™ Hip Screw This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92. ## Description/Intended Use The Cannulated PLUS™ Hip Screw (Sterile) is a cannulated bone screw. An optional beveled washer is available for use with osteoporotic bone. The device is available in a range of sizes to fit varying anatomical requirements. The Cannulated PLUS™ Hip Screw (Sterile) is intended to be used in the fixation of cancellous bone, for treatment of conditions such as, but not limited to: 1. femoral neck fractures; 2. arthrodesis of the hip; 3. femoral condyle fractures; 4. tibial plateau fractures; 5. ankle malleolar fractures; or 6. elbow fractures ## Material The Cannulated PLUS™ Hip Screw (Sterile) is manufactured from stainless steel. The optional washer is also manufactured from stainless steel. 000040 71
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...