Integra DigiFuse Cannulated Intramedullary Fusion System

K152527 · Ascension Orthopedics · HWC · Oct 29, 2015 · Orthopedic

Device Facts

Record IDK152527
Device NameIntegra DigiFuse Cannulated Intramedullary Fusion System
ApplicantAscension Orthopedics
Product CodeHWC · Orthopedic
Decision DateOct 29, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The DigiFuse® implant is indicated for the fixation of osteotomies and reconstruction of the lesser phalanges during procedures to correct deformities of the toes and fingers. Indications include: · Hammer toe deformity - · Claw toe deformity - · Mallet toe deformity - · Other deformities of the feet and hands The DigiFuse® implants are intended for single use only.

Device Story

Integra® DigiFuse® is a one-piece threaded cannulated intramedullary fusion implant; used for fixation of osteotomies and reconstruction of lesser phalanges in toes and fingers. Implanted into medullary bone by orthopedic surgeons during corrective procedures. Available in various diameters (2.0mm, 2.5mm), angles (0°, 10°), and lengths. Device provides mechanical stabilization of bone segments to facilitate fusion. Single-use only.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical evaluation and dimensional analysis.

Technological Characteristics

Material: Titanium Alloy (Ti 6Al-4V ELI). Form factor: One-piece threaded cannulated intramedullary implant. Variations: 2.0mm/2.5mm diameters, 0°/10° angles, standard/short lengths. Mechanical fixation principle.

Indications for Use

Indicated for fixation of osteotomies and reconstruction of lesser phalanges in patients with toe and finger deformities, including hammer toe, claw toe, mallet toe, and other deformities.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 29, 2015 Ascension Orthopedics % Ms. Jayana Kenana Regulatory Associate Integra LifeSciences 311 Enterprise Drive Plainsboro, New Jersey 08536 Re: K152527 Trade/Device Name: Integra® Digifuse® Cannulated Intramedullary Fusion System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: August 19, 2015 Received: September 3, 2015 Dear Ms. Kenana: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ Page 2 - Ms. Jayana Kenana forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES. Food and Drug Administration ### Indications for Use 510(k) Number (if known) K152527 Device Name Integra® DigiFuse® Cannulated Intramedullary Fusion System Indications for Use (Describe) The DigiFuse® implant is indicated for the fixation of osteotomies and reconstruction of the lesser phalanges during procedures to correct deformities of the toes and fingers. Indications include: · Hammer toe deformity - · Claw toe deformity - · Mallet toe deformity - · Other deformities of the feet and hands The DigiFuse® implants are intended for single use only. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of infornation is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 6. 510(k) Summary | Submitter Information – 807.92(a)(1) | | |---------------------------------------------------------------------------------------------|----------------------------------------------------| | Sponsor Name | Integra LifeSciences Corporation | | Address | 311 Enterprise Drive<br>Plainsboro, NJ 08536 | | Name of Contact Person | Jayana Kenana | | Phone Number | 609-936-2657 | | Fax Number | 609-275-9445 | | Establishment Registration Number | 1651501 | | Date Prepared | October 21, 2015 | | Name of Device – 807.92(a)(2) | | | Trade of Proprietary Name | Integra® DigiFuse® | | Common or Usual Name | Cannulated Intramedullary Fusion System | | Classification Name | Smooth & threaded metallic bone fixation fasteners | | Classification Panel | Orthopedics | | Regulation | Class II, 21 CFR 888.3040 | | Product Code | HWC | | Material | Titanium Alloy (Ti 6A1-4V ELI) | | Legally marketed device(s) to which equivalence is claimed – 807.92(a)(3) | | | Metasurg DigiFuse Implant – K111536 | | | OsteoMed Extend 2.0/2.4 Cannulated Screw System – K062863 | | | NewDeal K-Wire – K022599 | | | Device Description – 807.92(a)(4) | | | The Integra® DigiFuse® Implant is a one-piece threaded device intended to be implanted into | | the medullary bone of the lesser toes. The implant is offered in several variations consisting of the combination of 2.0mm and 2.5mm, 0° and 10° blades, as well as standard and short blade options. {4}------------------------------------------------ ## Integra LifeSciences Corporation Traditional 510(k) Integra DigiFuse Cannulated Intramedullary Fusion System ## Intended Use The DigiFuse® implant is indicated for the fixation of osteotomies and reconstruction of the lesser phalanges during procedures to correct deformities of the toes and fingers. Indications include: - · Hammer toe deformity - · Claw toe deformity - Mallet toe deformity - Other deformities of the feet and hands The DigiFuse® implants are intended for single use only. ## Conclusion: The Integra® DigiFuse® Implant is composed as the same material as the predicate device from MetaSurg (K111536) and the only difference is the proposed device will be adding two new lengths. The two new lengths are also included in the length range of the predicate device passed by Osteomed (K062863). There are no design, material, or indication differences between the proposed device and predicate devices. Non-clinical evaluation and dimensional analysis was done to confirm substantial equivalence.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...