I.B.S SNAP-OFF SCREW / SOLAFIX SNAP-OFF SCREW

K132911 · In2bones Sas · HWC · Apr 4, 2014 · Orthopedic

Device Facts

Record IDK132911
Device NameI.B.S SNAP-OFF SCREW / SOLAFIX SNAP-OFF SCREW
ApplicantIn2bones Sas
Product CodeHWC · Orthopedic
Decision DateApr 4, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The I.B.S.® Snap-off screws are indicated for bone reconstruction, osteotomy and fracture fixation of bones appropriate for the size of the device. Examples include: Mono-cortical fixation of small bone fragments Weil osteotomy Osteotomies and fractures fixation in the foot and hand.

Device Story

I.B.S. Snap-off screw is a single-piece, self-drilling, self-tapping metallic bone fixation fastener. Device consists of a screw head and a drill shank; shank inserts into a driver for bone insertion. Upon reaching target depth, drill shank twists off and breaks cleanly from screw head. Used in hospital or outpatient surgery centers by surgeons for bone reconstruction, osteotomy, and fracture fixation. Implanted for >30 days. Benefits include stable fixation of small bone fragments in foot and hand procedures. Supplied sterile via gamma irradiation.

Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F543-07 to assess torque to failure, insertion torque, axial pullout strength, and self-tapping performance. Engineering/dimensional comparison to predicates performed. Results indicated device performed as expected for all tests.

Technological Characteristics

Material: Titanium alloy TA6V (ISO 5832-3, ASTM F136). Design: Solid, non-cannulated, self-drilling, self-tapping screw with snap-off mechanism. Sterilization: Gamma irradiation. Single-use.

Indications for Use

Indicated for bone reconstruction, osteotomy, and fracture fixation of bones appropriate for device size, including mono-cortical fixation of small bone fragments, Weil osteotomy, and foot/hand fracture fixation.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows a logo for "In2Bones". The logo is an oval shape that is split into two parts by a white line. The text "In2Bones" is written in black letters on the right side of the oval. The logo is simple and modern. ﺔ 23, charren du Va: " 20120 Eculiv " 21 + 7210 - 72 20 outar solar hots #### K132911 page 1 of 4 ### 004_Revised 510(k) Summary for I.B.S. Snap-off Screw (K132911) Response to September 17, 2013 RTA Checklist, Section A4a.pdf # 510(k) SUMMARY In2Bones I.B.S." Snap-off screw | Sponsor identification | In2Bones SAS<br>28 chemin du Petit Bois<br>69130 Ecully - France<br>Phone: +33.4.72.29.26.26<br>Fax: +33.4.72.29.26.29 | |------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Establishment registration number | New company. Will register following FDA clearance | | Date of preparation | August 1st, 2013 | | Contact person | Norman Estrin<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Fax: (301) 294-0126<br>Email: estrin@yourFDAconsultant.com | | Authorized Agent in the<br>United States | Norman Estrin<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Fax: (301) 294-0126<br>Email: estrin@yourFDAconsultant.com | | Proprietary Name | I.B.S.® Snap-off screw | | Common name | Bone fixation screw | | Device classification regulation | 21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener<br>Class II | | Device Product<br>Code and Panel | HWC: screw, fixation, bone<br>87 orthopedics | | Device Description | The I.B.S.® Snap-off screw is a one-piece device made of Titanium<br>Alloy intended to be used as a screw for bone reconstruction,<br>osteotomy and fracture fixation of bones appropriate for the size of the<br>device. | | | The implant is a self-drilling and self-tapping snap-off screw. It is<br>machined on the end of a drill shank which inserts into a driver. After<br>insertion of the screw into the bone, the drill shank twists off and<br>breaks cleanly from the screw head. | | | The implant is supplied sterile and is available in various sizes<br>diameters and lengths. | | | Sizes:<br>The I.B.S.® Snap-off screws are available in various diameters and<br>lengths. | | | Material:<br>The I.B.S.® Snap-off screws are manufactured from titanium alloy<br>TA6V as per ISO 5832-3 and ASTM F136. They do not have any<br>coating. | | | Single use:<br>The I.B.S.® Snap-off screws are designed for single use only.<br>Sterilization: | | | The I.B.S.® Snap-off screws are supplied sterile, using gamma<br>irradiation. | | | Place of use:<br>The I.B.S.® Snap-off screws are indicated for use in a hospital, or<br>outpatient surgery center where sterile field may be created and<br>maintained. | | Predicate Devices | Integra / Newdeal Spin screw (K991477)<br>Arthrex Low Profile Screws (K103705)<br>Nexa Orthopedics / Nexa Bone screw (K053394) | | Indications for use | The I.B.S.® Snap-off screws are indicated for bone reconstruction,<br>osteotomy and fracture fixation of bones appropriate for the size of the<br>device. | | | Examples include:<br>Mono-cortical fixation of small bone fragments<br>Weil osteotomy<br>Osteotomies and fractures fixation in the foot and hand. | {1}------------------------------------------------ に K132911 page 2 of 4 {2}------------------------------------------------ K132911 page 3 of 4 #### COMPARISON OF THE INDICATIONS FOR USE WITH THE PREDICATE DEVICES: As with the predicate devices, the I.B.S.® Snap-off screws are indicated for surgical implantation longer than 30 days in the fixation of bone fractures or for bone reconstruction. | 510k number | K132911 | K991477 | K103705 | K053394 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------| | manufacturer | In2Bones | Newdeal -- an<br>Integra Lifesciences<br>company | Arthrex | Nexa Orthopedics<br>/ Tornier | | Name of<br>device | I.B.S.® Snap-off<br>screw | SPIN® screw | Low Profile Screw<br>- QuickFix | Nexa Bone screw<br>system / NexFix | | Use | Single use | Single use | Single use | Single use | | Fixation | Bone | Bone | Bone | Bone | | Indications<br>for use | The I.B.S.® Snap-<br>off screws are<br>indicated for bone<br>reconstruction,<br>osteotomy and<br>fracture fixation of<br>bones appropriate<br>for the size of the<br>device.<br>Examples include:<br>Mono-cortical<br>fixation of small<br>bone fragments<br>Weil osteotomy<br>Osteotomies and<br>fractures fixation in<br>the foot and hand. | The SPIN® Snap-off<br>screw is indicated<br>for fixing the<br>elective osteotomies<br>of the mid-foot<br>bones and the<br>metatarsal and<br>phalanges of the<br>foot only.<br>Examples include:<br>Weil osteotomy<br>Unicortical small<br>bone fixation | The Arthrex Low<br>Profile Screws<br>(2.0-3.0mm solid)<br>are intended to be<br>used as stand-<br>alone bone screws,<br>or in a plate-screw<br>system for internal<br>bone fixation for<br>bone fractures,<br>fusions,<br>osteotomies and<br>non-unions in the<br>ankle, foot, hand,<br>and wrist. | The Nexa bone<br>screw system<br>provides fixation<br>of fractures,<br>fusions, or<br>osteotomies of the<br>hand and foot. | {3}------------------------------------------------ #### K132911 page 4 of 4 | Comparison of<br>Technological<br>characteristics | The technological characteristics of the I.B.S.® Snap-off screw are the<br>same as the characteristics of predicates devices in terms of intended<br>use and design. All the screws have the following features:<br>- Solid / non-cannulated – the I.B.S.® Snap-off screws and all<br>predicate devices are solid / non-cannulated screws<br>- Self-tapping and self-drilling, with a snap-off design - the<br>I.B.S® Snap-off screws and all predicate devices are self-<br>tapping and self-drilling screws with a snap-off mechanism<br>- Material: made from Titanium alloys, with no new materials<br>being introduced in the product - the I.B.S.® Snap-off screws<br>and all predicate devices are manufactured from Titanium<br>Alloy TA6V<br>- Equivalent size range: the diameters and lengths covered by the<br>predicate devices enable to cover diameters and lengths of the<br>I.B.S.® Snap-off screws<br>- Indicated for surgical implantation longer than 30 days - the<br>I.B.S.® Snap-off screws and all predicate devices are indicated<br>for surgical implantation longer than 30 days | |---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Substantial Equivalence<br>Summary | The I.B.S.® Snap-off screw has similar technological characteristics<br>when compared to the predicate devices. | | Summary Performance<br>Data | Mechanical testing was performed according to ASTM F543-07. This<br>standard describes methods to assess the torque to failure, insertion<br>torque, axial pullout strength, and self-tapping performance of screws.<br>The specification standard provides requirements for materials, finish,<br>marking, care and handling, and the acceptable dimensions and<br>tolerances for metallic bone screws implanted into bone tissue. As the<br>specification standard states, the dimensions and tolerances are<br>applicable only to metallic bone screws described in the specification<br>itself. Additional information must be provided to document that the<br>design of the product will provide adequate mechanical properties for<br>the particular application.<br>In addition, an engineering / dimensional comparison to the predicates<br>was performed to ascertain substantial equivalence.<br>The results of the testing performed by the test lab indicated that the<br>I.B.S.® Snap-off screw performed as expected for each test. | | CONCLUSION | Based on the evaluations performed, the design and indications of<br>the I.B.S.® Snap-off screw are substantially equivalent to the<br>predicates identified in the 510(k) submission. No new materials or<br>processes are used in the development of this implant.<br><br>In addition, the results of the testing performed by the test lab<br>indicated that the screws performed as expected for each test.<br><br>The I.B.S.® Snap-off screws are acceptable for the application. | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service April 4, 2014 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002 In2Bones SAS % Norman Estrin, Ph.D. Managing Partner Estrin Consulting Group LLC 9109 Copenhaver Drive Potomac, Maryland 20854 Re: K132911 Trade/Device Name: I.B.S.® Snap-off screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: March 8, 2014 Received: March 11. 2014 Dear Dr. Estrin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {5}------------------------------------------------ Page 2 -Norman Estrin, Ph.D. (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, ### Mark N. Mellerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ K132911 page 1 of 1 #### Indications for Use 510(k) Number (if known): K132911 Device Name: I.B.S snap-off screw Indications For Use: The I.B.S snap-off screws are intended for bone reconstruction, osteotomy and fracture fixalinn of bones appropriate for the size of the device. Examples include : Mono-cortical fixation of small bone fragments Weil ostcolomy Osteotomies and fractures fixation in the foot and hand. Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ## Elizabet顾迹:rank -S Division of Orthopedic Devices
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...