K060914 · Johnson & Johnson Company · HWC · Jun 9, 2006 · Orthopedic
Device Facts
Record ID
K060914
Device Name
TRIDENT ANCHOR
Applicant
Johnson & Johnson Company
Product Code
HWC · Orthopedic
Decision Date
Jun 9, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Trident Anchor is indicated for use in the following: Shoulder: Rotator Cuff Repair, Biceps Tenodesis Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis; Joint Capsule Closure Elbow: Biceps Tendon Reattachment
Device Story
Trident Anchor is a sterile, metallic bone fixation fastener used for soft tissue-to-bone attachment. Device consists of an anchor pre-mounted on an inserter shaft with an anvil, optionally including a threader tab and #2 sutures. System deployed via reusable Deployment Gun. Used by surgeons in operating room settings for orthopedic repairs including rotator cuff, biceps tenodesis, and ligament reconstructions. Provides secure fixation of soft tissue to bone, facilitating healing and structural stability in orthopedic procedures.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Metallic bone fixation fastener; supplied sterile; pre-mounted on inserter shaft with anvil; includes #2 suture options; deployed via reusable mechanical deployment gun. Class II device under 21 CFR 888.3040.
Indications for Use
Indicated for patients requiring soft tissue-to-bone fixation in shoulder (rotator cuff repair, biceps tenodesis), knee (MCL, LCL, posterior oblique ligament repair, iliotibial band tenodesis, joint capsule closure), and elbow (biceps tendon reattachment) procedures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
K140855 — ARTHREX SUTURETAK SUTURE ANCHORS · Arthrex, Inc. · Dec 18, 2014
K063478 — VERSALOK ANCHOR WITH FREE STRANDS OF ORTHOCORD SUTURE (210808), ETHIBOND SUTURE (210809), AND PANACRYL SUTURE (210810) · Depuy Mitek, A Johnson & Johnson Company · Dec 15, 2006
K092133 — ARTHROCARE TITANIUM SUTURE ANCHOR, MODEL 22-9055,22-9055N, 22-9065, 22-9065N · Arthrocare Corp. · Nov 5, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K060914 Page 1 of 2
## SECTION 2 – 510(k) SUMMARY
| | Trident Anchor | | JUN - 9 2006 |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|--------------|
| Submitter's Name and<br>Address: | DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767 | | |
| Contact Person | Ruth C. Forstadt<br>Project Management Lead, Regulatory Affairs<br>DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br>Telephone:<br>508-977-3988<br>Facsimile:<br>508-828-3750<br>e-mail: | rforstad@dpyus.jnj.com | |
| Name of Medical Device | Classification Name: | Fastener, Fixation, Nondegradable, Soft<br>Tissue Smooth or threaded metallic bone<br>fixation fasteners | |
| | Common/Usual Name:<br>Proprietary Name: | Bone Anchor<br>Trident Anchor | |
| Substantial Equivalence | Trident Anchor is substantially equivalent to:<br>ROC EZ Fastener (K970089 & K971922); the Arthrex Pushlock<br>Anchor (K051219) and the ArthroCare Opus Magnum Implant<br>(K042914). | | |
| Device Classification | This device carries an FDA product code MBI and HWC, and is<br>classified as Fastener, Fixation, Nondegradable, Soft Tissue Smooth or<br>threaded metallic bone fixation fasteners under 21 CFR 888.3040. | | |
| Device Description | The Trident Anchor System includes the Trident Anchor, which will be<br>presented sterile, pre-mounted on an inserter shaft w/anvil with or<br>without a threader tab and suture. The System will be deployed with<br>the use of a reusable Deployment Gun. The Anchor could be provided<br>with a variety of #2 suture options. | | |
| Premarket Notification: Traditional | | | Confidential |
Premarket Notification: Traditional Trident Anchor
.
:
{1}------------------------------------------------
| Indications for Use | Trident Anchor is indicated for use in the following:<br>Shoulder: Rotator Cuff Repair, Biceps Tenodesis<br>Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament<br>Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis;<br>Joint Capsule Closure<br>Elbow: Biceps Tendon Reattachment |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety and Performance | Results of performance and safety testing have demonstrated that the<br>modified device is substantially equivalent to the predicate devices.<br>Based on the indications for use, technological characteristics, and<br>comparison to predicate devices, the Trident Anchor has been shown to<br>be substantially equivalent to predicate devices under the Federal Food,<br>Drug and Cosmetic Act. |
.
Confidential
.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 9 2006
DePuy Mitek a Johnson & Johnson Co. % Ms. Ruth C. Forstadt Project Management Lead, Regulatory Affairs 325 Paramount Drive Raynham, Massachusetts 02767
Re: K060914
Trade/Device Name: Trident Anchor Regulation Code: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulation Class: II Product Code: HWC, MBI Dated: April 3, 2006 Received: April 4, 2006
Dear Ms. Forstadt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 -- Ms. Ruth C. Forstadt
CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page Of
510(k) Number (if known): K660914
Device Name: Trident Anchor
Trident Anchor is indicated for use in the following:
Shoulder: Rotor Cuff Repair, Biceps Tenodesis Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis; Joint Capsule Closure Elbow: Biceps Tendon Reattachment
Prescription Use
OR
Over-the -Counter Use No
(Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Milliken
Division Sign-Off) Division of General, Restorative and Neurological Devices
Number K06 09/4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.