CSS CANNULATED SCREW SYSTEM

K060428 · Orthohelix Surgical Designs, Inc. · HWC · Mar 7, 2006 · Orthopedic

Device Facts

Record IDK060428
Device NameCSS CANNULATED SCREW SYSTEM
ApplicantOrthohelix Surgical Designs, Inc.
Product CodeHWC · Orthopedic
Decision DateMar 7, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The CSS Cannulated Screw System will be indicated for use in long and small bone fracture fixation, which includes the following; - Fractures of the tarsals and metatarsals - Fractures of the olecranon, distal humerous - Fractures of the radius and ulna - Patella fractures - Distal tibia and pilon fractures - Fractures of the fibula, medial malleolus, os calcis - Tarso-metatarsal and metatarsal phalangeal arthrodesis - Metatarsal and phalangeal osteotomies - Osteochondritis dissecans - Ligament fixation - Other small fragment, cancellous bone fractures and osteotomies

Device Story

CSS Cannulated Screw System consists of self-tapping cannulated screws, washers, and guide wires; used for bone fixation in orthopedic surgical procedures. Device provides mechanical stabilization of fractures and osteotomies. Surgeons select appropriate screw diameter, length, and thread configuration based on anatomical requirements. Washers (flat, contoured, partial contoured, domed) provide anatomical fit. System is used in clinical/surgical settings. Output is physical fixation of bone fragments, aiding healing and structural stability.

Clinical Evidence

No clinical data. Evidence consists of bench testing and Finite Element Analysis (FEA) comparing the subject device to the predicate.

Technological Characteristics

Materials: Titanium Ti-6Al-4V ELI Alloy (ASTM F-136). Components: Cannulated screws, washers, guide wires. Principle: Mechanical bone fixation via threaded fasteners. Design: Self-tapping, various diameters/lengths/thread configurations. Sterilization: Not specified.

Indications for Use

Indicated for long and small bone fracture fixation, including tarsal, metatarsal, olecranon, distal humerus, radius, ulna, patella, distal tibia, pilon, fibula, medial malleolus, and os calcis fractures; also indicated for arthrodesis, osteotomies, osteochondritis dissecans, and ligament fixation.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Premarket Notification CSS Cannulated Screw System ## 510(k) SUMMARY ## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared. OrthoHelix Surgical Designs, Inc. 1815 W. Market Suite 205 Akron, Ohio 44313 Phone: (330) 869-9563 Fax: (330) 869-9583 Contact Person: Lee A. Strnad Senior Development Manager OrthoHelix Surgical Designs, Inc. Date Prepared: February 13, 2006 #### Name of Device CSS Cannulated Screw System #### Common or Usual Name Cannulated Screws ## Classification Name Screw, Fixation, Bone Single/Multiple Component Metallic Fixation Appliances and Accessories #### Predicate Devices Howmedica Asnis III Cannulated Screw System (K000080) #### Intended Use The CSS Cannulated Screw System will be indicated for use in long and small bone fracture fixation, which includes the following; - · Fractures of the tarsals and metatarsals {1}------------------------------------------------ 510(k) Premarket Notification CSS Cannulated Screw System - · Fractures of the olecranon, distal humerous - · Fractures of the radius and ulna - · Patella fractures - · Distal tibia and pilon fractures - · Fractures of the fibula, medial malleolus, os calcis - · Tarso-metatarsal and metatarsal phalangeal arthrodesis - · Metatarsal and phalangeal osteotomies - · Osteochondritis dissecans - · Ligament fixation - · Other small fragment, cancellous bone fractures and osteotomies ## Device Description The CSS Cannulated Screw System consists of cannulated screws of various diameters, lengths and thread configurations. The system also includes correspondingly sized washers, the use of which are optional, as well as guide wire and various instruments. The cannulated screws are self tapping. They are made from made from Titanium Ti-6Al-4V ELI Alloy in conformance with ASTM F-136 Standard Specification for Wrought Titanium 6Al-4V ELI Alloy for Surgical Implant Applications. The CSS System includes (4) different types of washers. These are; Flat Washers, Contoured Washers, Partial Contoured Washers, and Domed Washers. The various styles are intended to offer the surgeon a variety of anatomical fit to the bone in certain applications. ## Performance Data A Finite Element Analysis (FEA) was performed on the CSS Cannulated Screw System, in comparison to its' predicate, the Asnis III. Performance testing was also conducted to confirm the results of the FEA. Results clearly demonstrate that the CSS System is substantially to its' predicate. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and head. The text "U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the bird symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 2006 MAR 7 ---- OrthoHelix Surgical Designs, Inc. C/o Mr. Lee A. Strnad Senior Development Manager 1815 West Market Street, Suite 205 Akron, Ohio 44313 Re: K060428 > Trade/Device Name: CSS Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Codes: HWC, HTN Dated: February 13, 2006 Received: February 21, 2006 Dear Mr. Strnad: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ #### Page 2 - Mr. Lee A. Strnad forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Qmee Mark N. Melkerson, M.S. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K060428 # Indications for Use 510(k) Number (if known): TBD Device Name: CCS Cannulated Screw System Indications for Use: The CSS Cannulated Screw System will be indicated for use in long and small bone fracture fixation, which includes the following: - Fractures of the tarsals and metatarsals . - Fractures of the olecranon, distal humerous - Fractures of the radius and ulna � - Patella fractures ● - Distal tibia and pilon fractures . - Fractures of the fibula, medial malleolus, os calcis ● - Tarso-metatarsal and metatarsal phalangeal arthrodesis . - Metatarsal and phalangeal osteotomies - Osteochondritis dissecans - Ligament fixation : 上一篇: 上一篇: - Other small fragment, cancellous bone fractures and osteotomies Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) | Page ___ of ___ | | |-----------------|--| |-----------------|--| Concurrence of CDRH, Office of Device Evaluation (ODE) ![Signature](signature.png) (Division Sign Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | K060428 | |---------------|---------| |---------------|---------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...