PERIARTICULAR LOCKING PLATES AND SCREWS, MODEL 2358 SERIES

K051098 · Zimmer, Inc. · HWC · Jul 7, 2005 · Orthopedic

Device Facts

Record IDK051098
Device NamePERIARTICULAR LOCKING PLATES AND SCREWS, MODEL 2358 SERIES
ApplicantZimmer, Inc.
Product CodeHWC · Orthopedic
Decision DateJul 7, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Zimmer Periarticular Locking Plate System is indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, nonunions, and malunions.

Device Story

The Zimmer Periarticular Locking Plate System is a metallic bone fixation device consisting of anatomically contoured, low-profile plates and locking screws. The system is designed for internal fracture fixation; threaded holes in the plate accept locking screws to create a stable, fixed-angle construct. It is used by orthopedic surgeons in a clinical or surgical setting to stabilize bone fractures or osteotomies. By providing rigid fixation, the device supports bone healing and stabilization, particularly in complex cases like comminuted or osteopenic bone fractures. The device is implanted during surgery to provide temporary support until the bone has sufficiently healed.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Metallic bone fixation plate and screw system. Features low-profile, anatomically contoured plates with threaded holes for locking screws to create fixed-angle constructs. Materials and manufacturing processes are similar to predicate devices.

Indications for Use

Indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted, supracondylar, intra-articular, and extra-articular condylar fractures, as well as fractures in osteopenic bone, nonunions, and malunions in patients requiring surgical fracture management.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the word "zimmer" in a bold, sans-serif font. To the left of the word is a large, stylized letter "Z" enclosed in a circle. The "Z" is also in a bold font and appears to be the first letter of the word "zimmer." JUL 7 - 2005 K051098 p 1/2 Traditional 510(k) Premarket Notification # Summary of Safety and Effectiveness | Submitter: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 | |---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Laura D. Williams, RAC<br>Manager, Corporate Regulatory Affairs<br>Telephone: (574) 372-4523<br>Fax: (574) 372-4605 | | Date: | April 29, 2005 | | Trade Name: | Zimmer® Periarticular Locking Plate System | | Classification Name: | Plate, Fixation, Bone<br>Screw, Fixation, Bone | | Classification Reference: | 21 CFR § 888.3030,3040 | | Predicate Device: | Zimmer ECT® Internal Fracture Fixation System,<br>Preamendment Device | | | Zimmer Periarticular Locking Plate System,<br>K040593, cleared April 12, 2004 | | Device Description: | The Zimmer Periarticular Locking Plate System is a<br>plate and screw system intended for internal<br>fracture fixation. The low-profile periarticular<br>locking plate is anatomically contoured and has<br>threaded holes which accept locking screws to<br>create a stable, fixed angle construct. | | Intended Use: | The Zimmer Periarticular Locking Plate System is<br>indicated for temporary internal fixation and<br>stabilization of osteotomies and fractures, including<br>comminuted fractures, supracondylar fractures,<br>intra-articular and extra-articular condylar fractures,<br>fractures in osteopenic bone, nonunions, and<br>malunions. | | Comparison to Predicate Device: | The Zimmer Periarticular Locking Plate System has<br>the same intended use, has similar performance<br>characteristics, is manufactured from similar | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image contains the word "zimmer" with a stylized "Z" in a circle to the left of the word. The "Z" is bold and black, while the word "zimmer" is in a smaller, sans-serif font. The overall impression is of a logo or brand mark. materials using similar processes, and is similar in design to the predicate devices. Performance Data: The results of non-clinical analysis demonstrate that the device is safe and effective. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized eagle with three lines extending from its head, resembling a flag. JUL 7 - 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Laura D. Williams, RAC Manager, Corporate Regulatory Affairs Zimmer Incorporated P.O. Box 708 Warsaw, Indiana 46581-0708 Re: K051098 Trade/Device Name: Zimmer® Periarticular Locking Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HWC Dated: April 29, 2005 Received: May 2, 2005 Dear Ms. Williams: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ### Page 2 - Ms. Laura D. Williams, RAC This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Hypt Clurke m C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use #### 510(k) Number (if known): #### Device Name: Zimmer® Periarticular Locking Plate System #### Indications for Use: The Periarticular Locking Plate System is indicated for temporary internal fixation and stabilization of osteotomies and fractures, including: - Comminuted fractures . - . Supracondylar fractures - Intra-articular and extra-articular condylar fractures . - . Fractures in osteopenic bone - . Nonunions - Malunions . . Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) 0017 (Please do not write below this line - Continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (DDE) | Page 1 of 1 | | |---------------|------------------------------------------------------------| | | (Division Sign-Off) | | | Division of General, Restorative, and Neurological Devices | | 510(k) Number | K051088 |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...