OSSIOfiber Cannulated Trimmable Fixation Nail, OSSIOfiber Trimmable Fixation Nail, OSSIOfiber Trimmable Fixation Nail, Cannulated Design

K203465 · OSSIO , Ltd. · HTY · Jan 6, 2021 · Orthopedic

Device Facts

Record IDK203465
Device NameOSSIOfiber Cannulated Trimmable Fixation Nail, OSSIOfiber Trimmable Fixation Nail, OSSIOfiber Trimmable Fixation Nail, Cannulated Design
ApplicantOSSIO , Ltd.
Product CodeHTY · Orthopedic
Decision DateJan 6, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

OSSIOfiber® Trimmable Fixation Nails are indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodesis, and bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace).

Device Story

Degradable bone fixation nail; composed of PLDLA polymer reinforced with continuous mineral fibers; provides mechanical fixation for bone fractures, osteotomies, arthrodesis, and bone grafts. Device degrades via hydrolysis into alpha-hydroxy acids and bone-natural minerals; load transfers to surrounding anatomy during healing; eliminates need for hardware removal surgery. Cannulated design; supplied sterile; single-patient use. Used by orthopedic surgeons in clinical settings; compatible with standard ISO 9714 orthopedic instrumentation. Provides structural support during bone healing; benefits patient by avoiding secondary removal procedures.

Clinical Evidence

Bench testing only. Performed flexural bending, pull-out, and shear testing to verify strength and fixation properties compared to predicate. Characterized in-vitro degradation profile. Biocompatibility established via reference to ISO 10993 data from predicate and reference devices (K181180, K190652).

Technological Characteristics

Material: PLDLA polymer reinforced with continuous mineral fibers. Form factor: Cannulated nail, 10-70 mm length, 2.80-4.60 mm diameter. Principle: Degradable fixation; load transfer via hydrolysis. Sterilization: Ethylene Oxide (EtO). Connectivity: None.

Indications for Use

Indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodesis, and bone grafts in patients requiring additional immobilization (e.g., rigid fixation implants, cast, brace).

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ January 6, 2021 Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. OSSIO Ltd. % David McGurl Director, Regulatory Affairs MCRA, LLC 1050 K Street NW. Suite 1000 Washington, District of Columbia 20001 Re: K203465 Trade/Device Name: OSSIOfiber Cannulated Trimmable Fixation Nail, OSSIOfiber Trimmable Fixation Nail, OSSIOfiber Trimmable Fixation Nail, Cannulated Design Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HTY Dated: November 24, 2020 Received: November 24, 2020 Dear David McGurl: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K203465 Device Name OSSIOfiber® Cannulated Trimmable Fixation Nail, OSSIOfiber® Trimmable Fixation Nail, OSSIOfiber® Trimmable Fixation Nail, Cannulated Design Indications for Use (Describe) OSSIOfiber® Trimmable Fixation Nails are indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodesis, and bone grafts in the presence of appropriate additional immobilization implants, cast, brace). Type of Use (Select one or both, as applicable) 区 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. {3}------------------------------------------------ ## K203465 # 510(k) SUMMARY # OSSIOfiber® Trimmable Fixation Nail, Cannulated Design ### Submitter # Ossio Ltd. 8 HaTochen Street, Caesarea, Israel, 3079861 Phone: +972-4-9986600 Facsimile: +972-4-9986601 Contact Person: Taly Lindner Date Prepared: November 24, 2020 Name of Device: OSSIOfiber® Cannulated Trimmable Fixation Nail, OSSIOfiber® Trimmable Fixation Nail, OSSIOfiber® Trimmable Fixation Nail, Cannulated Design Common or Usual Name: fixation, pin, smooth Classification Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II, 21 C.F.R. § 888.3040 Product Code: HTY # Primary Predicate Device OSSIO® Pin Product Family (K181180) ### Reference Devices OSSIOfiber® Hammertoe Fixation System (K190652) OSSIOfiber® Compression Screw (K193660) ### Device Description OSSIOfiber® Cannulated Trimmable Fixation Nails are indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodesis, and bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace). The OSSIOfiber® Cannulated Trimmable Fixation Nail is a fixation device made of degradable poly (L-lactide-co-D.L-lactide) (PLDLA) reinforced with continuous mineral fibers. The polymer content degrades by hydrolysis into alpha-hydroxy acids that are metabolized by the fibers are made from materials that are found in natural bone. As the OSSIOfiber® implants degrade, the load transfers to the surrounding anatomy throughout the healing period of the osteotomy, fusion, or fracture. Substantial degradation takes place within approximately 18 months as shown in pre-clinical studies, thus eliminating the requirement for future hardware removal surgery. {4}------------------------------------------------ The OSSIOfiber® Cannulated Trimmable Fixation Nails are cannulated, supplied sterile, for single patient use only, and non-pyrogenic. They are available in several sizes: 10-70 mm long, and 2.80-4.60 mm nominal diameter (core diameter of 2.4-4.00 mm). The OSSIOfiber® Cannulated Trimmable Fixation Nails are designed to be used with commonly available orthopedic surgical tools such as ISO 9714 compatible instrumentations. # Indications for Use OSSIOfiber® Trimmable Fixation Nails are indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodesis, and bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace). # Summary of Technological Characteristics The OSSIOfiber® Cannulated Trimmable Fixation Nails have the same intended use, indications for use, material composition, and principles of operation, and similar design characteristics as the predicate device OSSIO® Pin Product Family (K181180) and the reference device OSSIOfiber® Hammertoe Fixation System (K190652). The OSSIOfiber® Cannulated Trimmable Fixation Nails' material as well as the manufacturing methods are the same as that of the cleared predicate device OSSIO® Pin Product Family (K181180). Both the device and the predicate are supplied sterilized by EtO. Although there are differences with regards to shape and size and cannulation as compared to the predicate, mechanical testing demonstrated at least equivalent performance both initially and after in vitro degradation. Any differences between OSSIOfiber® Cannulated Trimmable Fixation Nail and its predicate device do not raise different questions of safety and effectiveness. ### Non-Clinical Data Flexural bending, pull-out, and shear testing were performed to verify the strength and fixation properties of the OSSIOfiber® Cannulated Trimmable Fixation Nail, and to compare them to those of the predicate device (K181180). Testing was done initially and following in-vitro degradation. The invitro degradation profile (i.e., change in material properties) was characterized. Biocompatibility for the implants was established primarily based on the referenced ISO 10993 data from the previously cleared predicate and reference devices (K181180, K190652) as well as a rationale. A rationale was provided to support the MR safe labeling of the device. {5}------------------------------------------------ # Conclusions The OSSIOfiber® Cannulated Trimmable Fixation Nail is substantially equivalent to its predicate device, OSSIO® Pin Product Family (K181180). The OSSIOfiber® Cannulated Trimmable Fixation Nails have the same intended use, indications for use, material composition, and principles of operation, and similar design characteristics as the predicate device OSSIO® Pin Product Family (K181180) and the reference device OSSIOfiber® Hammertoe Fixation System (K190652). The minor differences do not alter the intended surgical use of the device and do not affect its substantial equivalence when used as labeled. Non-clinical testing data demonstrate that the OSSIOfiber® Cannulated Trimmable Fixation Nail has a substantially equivalent safety and effectiveness profile as the predicate device. Thus, the OSSIOfiber® Cannulated Trimmable Fixation Nail is substantially equivalent.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...