PROFYLE TITANIUM HAND AND SMALL FRAGMENT SYSTEM
Device Facts
| Record ID | K961497 |
|---|---|
| Device Name | PROFYLE TITANIUM HAND AND SMALL FRAGMENT SYSTEM |
| Applicant | Howmedica, Inc. |
| Product Code | HRS · Orthopedic |
| Decision Date | Jun 28, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The system is intended for use in internal fixation of small bones including the hand, foot, and craniomaxillofacial skeleton.
Device Story
System consists of titanium plates and screws for internal bone fixation; plates available in various lengths, thicknesses, and configurations (straight, L, T, Y, Z, rectangular, condylar); screws (1.2-2.7 mm diameter) are fully threaded, self-tapping, with cross-slotted heads; used by surgeons for small bone stabilization; provides mechanical support to facilitate bone healing.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Titanium alloy plates and screws; 1.2-2.7 mm screw diameters; cross-slotted screw heads; self-tapping design; various plate geometries.
Indications for Use
Indicated for internal fixation of small bones in the hand, foot, and craniomaxillofacial skeleton.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Luhr® Fixation System (Howmedica)
- Synthes® Small Fragment and Mini Set (Synthes USA)
- Wurzburg® Titanium Mini Bone Plates and Screws (Howmedica Leibinger Inc.)