Arthrex BioSync® Bone Wedge

K151256 · Arthex, Inc. · PLF · Aug 12, 2015 · Orthopedic

Device Facts

Record IDK151256
Device NameArthrex BioSync® Bone Wedge
ApplicantArthex, Inc.
Product CodePLF · Orthopedic
Decision DateAug 12, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Arthrex BioSync Bone Wedge is intended to be used for internal bone fixation for bone fractures, fusions, or osteotomies, in the ankle and foot, such as: Cotton and Evans Wedges: - Opening wedge osteotomies of the bones of the foot including osteotomies for Hallux Valgus - Opening wedge of Medial Cuneiform or Cotton osteotomies - Lateral Column Lengthening (Evans Lengthening Osteotomy of Calcaneal Z Osteotomy) - Metatarsal/Cuneifrom arthrodesis Midfoot Wedges: - Opening wedge osteotomies of the bones of the foot including osteotomies for Hallux Valgus - Nonunion of arthrodesis of the Midfoot including Metatarsal/Cuneiform arthrodesis (TMT or Lapidus) This device is intended for use with ancillary fixation. The Arthrex BioSync Bone Wedge is not intended for use in the spine.

Device Story

Arthrex BioSync Bone Wedge is a family of pre-sized, porous titanium implants designed for angular correction of small bones in the ankle and foot. Used by orthopedic surgeons in clinical settings, the device acts as a structural spacer during osteotomies or arthrodesis procedures. It requires ancillary fixation (e.g., screws or plates) to stabilize the bone construct. By maintaining the desired bone geometry during the healing process, the wedge facilitates bone fusion and fracture repair. The device is provided as a sterile implant.

Clinical Evidence

Bench testing only. Mechanical testing data, including static compression, dynamic compression, and expulsion testing, was provided to demonstrate substantial equivalence.

Technological Characteristics

Porous titanium bone wedges. Available in varying widths and thicknesses. Designed for small bone applications in the ankle and foot. Requires ancillary fixation. Non-resorbable metallic implant.

Indications for Use

Indicated for internal bone fixation in ankle and foot fractures, fusions, or osteotomies, including Cotton, Evans, and midfoot procedures. Not for use in the spine.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 12, 2015 Arthrex, Incorporated Ms. Laura Medlin Regulatory Affairs Associate 1370 Creekside Boulevard Naples, Florida 34108-1945 Re: K151256 Trade/Device Name: Arthrex BioSync® Bone Wedge Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: PLF, HRS, HWC Dated: May 14, 2015 Received: May 22, 2015 Dear Ms. Medlin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ Page 2 – Ms. Laura Medlin forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerelv vours. ## Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image contains a logo and the text "TRADITIONAL 51". The logo is a stylized design with the word "Artistic" written in a script font. The text "TRADITIONAL 51" is written in a bold, sans-serif font and is located to the right of the logo. The overall design is simple and clean. TRADITIONAL 510(K): ARTHREX BIOSYNC BONE WEDGE ## 2.4 INDICATIONS FOR USE | DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 | |-----------------------------------------|-----------------------------------| | Food and Drug Administration | Expiration Date: January 31, 2017 | | | See PRA Statement below. | **Indications for Use** | 510(k) Number ( <i>if known</i> ) | K151256 | |-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Arthrex BioSync® Bone Wedge | | Indications for Use ( <i>Describe</i> ) | The Arthrex BioSync Bone Wedge is intended to be used for internal bone fixation for bone fractures, fusions, or osteotomies, in the ankle and foot, such as: | Cotton and Evans Wedges: - Opening wedge osteotomies of the bones of the foot including osteotomies for Hallux Valgus - Opening wedge of Medial Cuneiform or Cotton osteotomies - Lateral Column Lengthening (Evans Lengthening Osteotomy of Calcaneal Z Osteotomy) - Metatarsal/Cuneifrom arthrodesis Midfoot Wedges: - Opening wedge osteotomies of the bones of the foot including osteotomies for Hallux Valgus - Nonunion of arthrodesis of the Midfoot including Metatarsal/Cuneiform arthrodesis (TMT or Lapidus) This device is intended for use with ancillary fixation. The Arthrex BioSync Bone Wedge is not intended for use in the spine. | Type of Use ( <i>Select one or both, as applicable</i> ) | <span><input checked="true" type="checkbox"/></span> Prescription Use (Part 21 CFR 801 Subpart D) <span><input type="checkbox"/></span> Over-The-Counter Use (21 CFR 801 Subpart C) | |----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* {3}------------------------------------------------ ## 2.5 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS | Date Summary Prepared | August 5, 2015 | |------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer/<br>Distributor/<br>Sponsor | Arthrex, Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 USA | | 510(k) Contact | Laura Medlin<br>Regulatory Affairs<br>Arthrex, Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 USA<br>Telephone: 239/643.5553, ext. 72005<br>Fax: 239/598.5508<br>Email: Laura.Medlin@Arthrex.com | | Trade Name | Arthrex BioSync Bone Wedge | | Common Name | Plate, fixation, bone<br>Screw, fixation, bone | | Product Code,<br>Classification Name | PLF – Bone Wedge<br>HRS - Plate, Fixation, Bone<br>HWC - Screw, Fixation, Bone | | CFR | 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances<br>and accessories<br>21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener | | Predicate Device | K140531: Wright Medical Technology, Inc. BIOFOAM® Bone Wedge<br>K141635: Arthrex iBalance® TKA System | | Purpose of Submission | This traditional 510(k) premarket notification is submitted to obtain clearance for<br>the Arthrex BioSync Bone Wedge. | | Device Description | The Arthrex BioSync Bone Wedge is a family of pre-sized implantable titanium<br>porous metal wedges intended to be used for angular correction of small bones in<br>the ankle and foot. It is offered with varying widths and thicknesses to<br>accommodate a variety of small bone applications. | | Intended Use | The Arthrex BioSync Bone Wedge is intended to be used for internal bone fixation<br>for bone fractures, fusions, or osteotomies in the ankle and foot, such as:<br><br>Cotton and Evans Wedges:<br>Opening wedge osteotomies of the bones of the foot including osteotomies for Hallux ValgusOpening wedge of Medial Cuneiform or Cotton osteotomiesLateral Column Lengthening (Evans Lengthening Osteotomy of Calcaneal Z Osteotomy)Metatarsal/Cuneiform arthrodesis<br>Midfoot Wedges:<br>Opening wedge osteotomies of the bones of the foot including osteotomies for Hallux ValgusNonunion of arthrodesis of the Midfoot including Metatarsal/Cuneiform arthrodesis (TMT or Lapidus)<br>This device is intended for use with ancillary fixation. The Arthrex BioSync Bone<br>Wedge is not intended for use in the spine. | | Substantial<br>Equivalence Summary | The Arthrex BioSync Bone Wedge is substantially equivalent to the predicate<br>devices, in which the basic design features and intended uses are the same. Any<br>differences between the Arthrex BioSync Bone Wedge and the predicates are<br>considered minor and do not raise questions concerning safety and effectiveness.<br>The submitted mechanical testing data, inclusive of static compression, dynamic<br>compression and expulsion testing, demonstrates that the wedge is substantially | | | equivalent to that of the predicate devices. Based on the indications for use, technological characteristics, and the summary of data submitted, Arthrex, Inc. has determined that the <b>Arthrex BioSync Bone Wedge</b> is substantially equivalent to currently marketed predicate devices. | {4}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...