Tornier Perform Humeral System – Stemless

K220418 · Tornier, Inc. · PKC · Sep 14, 2022 · Orthopedic

Device Facts

Record IDK220418
Device NameTornier Perform Humeral System – Stemless
ApplicantTornier, Inc.
Product CodePKC · Orthopedic
Decision DateSep 14, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3660
Device ClassClass 2
AttributesTherapeutic

Intended Use

The nucleus, humeral head coupler and humeral head are used in conjunction with a glenoid implant as a total replacement. The Tornier Perform® Humeral System - Stemless is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide mobility, stability, and to relieve pain. The Tornier Perform Humeral System -Stemless is indicated for use as a replacement of shoulder joints disabled by: - Non-inflammatory degenerative joint disease (i.e. osteoarthritis) and avascular necrosis - Post-traumatic arthritis Titanium humeral heads are intended for patients with suspected cobalt alloy material sensitivity. The wear properties of titanium and titanium alloys are inferior to that of cobalt alloy. A titanium humeral head is not recommended for patients without a suspected material sensitivity to cobalt alloy. All components are single use and intended for cementless use. The Tornier Perform Humeral System - Stemless is intended to be used with cemented polyethylene glenoid components, in an anatomic total shoulder arthroplasty.

Device Story

Metaphyseal humeral system for anatomic total shoulder arthroplasty; replaces shoulder joints disabled by degenerative disease or trauma. System comprises titanium nucleus components, humeral head couplers, and modular humeral heads; implanted with existing Tornier anatomic glenoid systems. Designed for cementless use in patients with intact/reconstructable rotator cuff. Titanium heads provided for patients with cobalt alloy sensitivity. Implanted by orthopedic surgeons in clinical settings. Provides joint mobility, stability, and pain relief. Substantial equivalence established via non-clinical performance testing including fatigue, finite element analysis, and wear evaluations.

Clinical Evidence

No clinical studies were performed. Substantial equivalence supported by non-clinical bench testing, including fatigue testing, displacement finite element analysis, static nucleus testing, static taper evaluation, wear and range of motion evaluations, MRI compatibility, and particulate testing.

Technological Characteristics

Metaphyseal humeral system; titanium nucleus components (additively manufactured, four-fin design); modular humeral heads. Cementless fixation. Intended for use with cemented polyethylene glenoid components. Evaluated for biocompatibility, sterilization, and shelf life per consensus standards.

Indications for Use

Indicated for patients with intact/reconstructable rotator cuff requiring shoulder joint replacement due to non-inflammatory degenerative joint disease (osteoarthritis), avascular necrosis, or post-traumatic arthritis. Titanium heads indicated for patients with suspected cobalt alloy sensitivity.

Regulatory Classification

Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,” (vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ September 14, 2022 Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. To the right of the square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue. Tornier, Inc. Renee Stoffel Principal Specialist, Regulatory Affairs 10801 Nesbitt Ave South Bloomington, Minnesota 55437 ## Re: K220418 Trade/Device Name: Tornier Perform Humeral System - Stemless Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: PKC Dated: August 19, 2022 Received: August 19, 2022 Dear Renee Stoffel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Victoria Lilling, M.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K220418 Device Name Tornier Perform® Humeral System - Stemless #### Indications for Use (Describe) The nucleus, humeral head coupler and humeral head are used in conjunction with a glenoid implant as a total replacement. The Tornier Perform® Humeral System - Stemless is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide mobility, stability, and to relieve pain. The Tornier Perform Humeral System -Stemless is indicated for use as a replacement of shoulder joints disabled by: - · Non-inflammatory degenerative joint disease (i.e. osteoarthritis) and avascular necrosis - · Post-traumatic arthritis Titanium humeral heads are intended for patients with suspected cobalt alloy material sensitivity. The wear properties of titanium and titanium alloys are inferior to that of cobalt alloy. A titanium humeral head is not recommended for patients without a suspected material sensitivity to cobalt alloy. All components are single use and intended for cementless use. The Tornier Perform Humeral System - Stemless is intended to be used with cemented polyethylene glenoid components, in an anatomic total shoulder arthroplasty. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------|--------------------------------------------------------------| | <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "TORNIER" in a serif font, with each letter in blue. To the right of the word is a blue hexagon containing a stylized letter "T" inside. The letter "T" is also blue and has a smaller hexagon shape in the middle. Date Prepared: February 11, 2022 #### Administrative Information | Name: | Tornier, Inc. | |-----------------|---------------------------------------------------------------------------------| | Address: | 10801 Nesbitt Avenue South<br>Bloomington, MN 55437<br>United States of America | | Contact Person: | Renee Stoffel | | Title: | Principal Specialist, Regulatory Affairs | | Phone: | 952-683-7471 | | Fax: | 952-426-7601 | #### Device Information | Name of Device: | Tornier Perform® Humeral System - Stemless | |-------------------|----------------------------------------------------------------------------------------| | Common Name (s): | Shoulder Prosthesis | | Regulatory Class: | II | | Regulation: | 21 CFR 888.3660, Shoulder joint metal/polymer semi-constrained cemented<br>prosthesis. | | Product Codes: | PKC | ## Predicate Device Information | Primary Predicate: | Simpliciti Shoulder System, K143552 | |-----------------------|------------------------------------------------------------------------------------------------| | Additional Predicate: | Perform Humeral System – Stem, K201315 | | Reference Device: | Aequalis Perform Reversed Glenoid, Aequalis Perform + Reversed Glenoid,<br>K161742 and K183696 | ## Device Description The Tornier Perform Humeral System - Stemless is metaphyseal humeral system intended for anatomic total shoulder arthroplasty. The Perform Humeral System - Stemless is implanted with existing Tornier anatomic glenoid systems. The Tornier Perform Humeral System – Stemless includes new titanium nucleus components and previously-cleared modular humeral heads (K201315). The system also includes reusable instruments used to implant the shoulder prosthesis. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "TORNIER" in a large, sans-serif font, colored in blue. To the right of the word is a logo consisting of a stylized letter "T" inside a hexagon. The logo is also colored in blue and has a modern, geometric design. # Indications for Use The nucleus, humeral head coupler and humeral head are used in conjunction with a glenoid implant as a total replacement. The Tornier Perform® Humeral System - Stemless is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide mobility, and to relieve pain. The Tornier Perform Humeral System - Stemless is indicated for use as a replacement of shoulder joints disabled by: - Non-inflammatory degenerative joint disease (i.e. osteoarthritis) and avascular necrosis - Post-traumatic arthritis ● Titanium humeral heads are intents with suspected cobalt alloy material sensitivity. The wear properties of titanium alloys are inferior to that of cobalt alloy. A titanium humeral head is not recommended for patients without a suspected material sensitivity to cobalt alloy. All components are single use and intended for cementless use. The Tornier Perform Humeral System - Stemless is intended to be used with cemented polyethylene glenoid components, in an anatomic total shoulder arthroplasty. # Comparison to Predicate Device The Tornier Perform Humeral System - Stemless and the primary predicate Simpliciti Shoulder System have the same intended use, same principle of operation, and similar technological features. Differences for the subject Tornier Perform Humeral System - Stemless system include nucleus components that are additively manufactured with a four-fin design and connection to modular humeral heads (previouslycleared in K201315). # Non-clinical Performance Testing Non-clinical testing was performed to demonstrate substantial equivalence to the predicate device. - Fatigue testing ● - Displacement finite element analysis - Comparative static nucleus testing - Static taper evaluation ● - Wear and range of motion evaluations - MRI compatibility evaluation ● - o Particulate testing Biocompatibility, sterilization, cleaning, endotoxin, packaging, shelf life, and distribution for the Tornier Perform Humeral System - Stemless components were also assessed in accordance with recognized consensus standards. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "TORNIER" in a large, sans-serif font, colored in blue. To the right of the word is a logo consisting of a stylized letter "T" inside a hexagon. The logo is also colored in blue and appears to be a modern design. # Clinical Testing No clinical studies were performed. ## Conclusions The Tornier Perform® Humeral System - Stemless does not raise different questions of safety or effectiveness. Differences in technological characteristics have been addressed with performance testing. The results of performance testing for the Tornier Perform® Humeral System - Stemless support substantial equivalence to the primary predicate Simpliciti Shoulder System (K143552, cleared March 4, 2015).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...