Browse hierarchy: [Orthopedic (OR)](/submissions/OR) → [Subpart D — Prosthetic Devices](/submissions/OR/subpart-d%E2%80%94prosthetic-devices) → [21 CFR 888.3660](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/888.3660) → PKC — Prosthesis, Total Anatomic Shoulder, Uncemented Metaphyseal Humeral Stem With No Diaphyseal Incursion, Semi-Constrained

# PKC · Prosthesis, Total Anatomic Shoulder, Uncemented Metaphyseal Humeral Stem With No Diaphyseal Incursion, Semi-Constrained

_Orthopedic · 21 CFR 888.3660 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC

## Overview

- **Product Code:** PKC
- **Device Name:** Prosthesis, Total Anatomic Shoulder, Uncemented Metaphyseal Humeral Stem With No Diaphyseal Incursion, Semi-Constrained
- **Regulation:** [21 CFR 888.3660](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/888.3660)
- **Device Class:** 2
- **Review Panel:** [Orthopedic](/submissions/OR)
- **Implant:** yes

## Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Classification Rationale

Class II. The special controls for this device are:

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Recent Cleared Devices (19 of 19)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K254154](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K254154.md) | FX V135 EASYTECH® Shoulder System | FX Shoulder Solutions, Inc. | May 1, 2026 | SESE |
| [K252129](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K252129.md) | AETOS Shoulder System Stemless Humeral Prosthesis | Smith & Nephew, Inc. | Oct 3, 2025 | SESE |
| [K243248](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K243248.md) | INHANCE INTACT™ | Depuy Ireland UC | Dec 23, 2024 | SESE |
| [K241873](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K241873.md) | OsseoFit Stemless Shoulder System | Zimmer, Inc. | Dec 11, 2024 | SESE |
| [K240716](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K240716.md) | AETOS Shoulder System Stemless Humeral Prosthesis | Smith & Nephew, Inc. | Nov 20, 2024 | SESE |
| [K221758](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K221758.md) | SMR Stemless Anatomic | Lima Corporate S.P.A. | Mar 17, 2023 | SESE |
| [K220418](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K220418.md) | Tornier Perform Humeral System – Stemless | Tornier, Inc. | Sep 14, 2022 | SESE |
| [K213615](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K213615.md) | Shoulder Innovations Total Shoulder System | Shoulder Innovations, Inc. | Mar 11, 2022 | SESE |
| [K212683](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K212683.md) | GLOBAL ICON Stemless Shoulder System | Depuy(Ireland) | Jan 19, 2022 | SESE |
| [K203108](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K203108.md) | Ignite Stemless Anatomic Shoulder System | Ignite Orthopedics, LLC | Jul 16, 2021 | SESE |
| [K201391](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K201391.md) | Easytech® Anatomical Shoulder System | Fx Shoulder USA, Inc. | Feb 16, 2021 | SESE |
| [K203100](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K203100.md) | Arthrex Eclipse Titanium Humeral Head | Arthrex, Inc. | Nov 20, 2020 | SESE |
| [K201542](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K201542.md) | Arthrex Eclipse Shoulder Prosthesis System | Arthrex, Inc. | Jul 8, 2020 | SESE |
| [K193226](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K193226.md) | AltiVate Anatomic Canal-Sparing (CS) Shoulder | Encore Medical L.P. | Jun 3, 2020 | SESE |
| [K192097](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K192097.md) | Exactech® Equinoxe® Stemless Humeral Components | Exactech, Inc. | Apr 29, 2020 | SESE |
| [K182516](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K182516.md) | Comprehensive Nano Stemless Shoulder | Biomet Manufacturing Corp | Apr 22, 2019 | SESE |
| [K173388](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K173388.md) | Exactech Equinoxe Stemless Shoulder | Exactech, Inc. | Mar 9, 2018 | SESE |
| [K171858](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K171858.md) | Sidus Stem-Free Shoulder | Zimmer GmbH | Dec 18, 2017 | SESE |
| [K143552](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC/K143552.md) | Simpliciti Shoulder System | Tornier, Inc. | Mar 4, 2015 | SESE |

## Top Applicants

- Arthrex, Inc. — 2 clearances
- Exactech, Inc. — 2 clearances
- Smith & Nephew, Inc. — 2 clearances
- Tornier, Inc. — 2 clearances
- Biomet Manufacturing Corp — 1 clearance

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PKC)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).