INHANCE™ Reverse Shoulder System

K253624 · Depuy Ireland UC · PHX · Mar 11, 2026 · Orthopedic

Device Facts

Record IDK253624
Device NameINHANCE™ Reverse Shoulder System
ApplicantDepuy Ireland UC
Product CodePHX · Orthopedic
Decision DateMar 11, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3660
Device ClassClass 2
AttributesTherapeutic

Intended Use

Anatomic Total Shoulder or Hemi-Shoulder The INHANCE SHOULDER SYSTEM with the humeral stemless anchor is intended for use in anatomic total shoulder replacement procedures to address the following: - Osteoarthritis - Post-traumatic arthrosis - Focal avascular necrosis of the humeral head - Previous surgeries of the shoulder that do not compromise the fixation The INHANCE SHOULDER SYSTEM with a humeral stem is intended for use in anatomic total or hemi-shoulder replacement procedures to address the following: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. - Rheumatoid arthritis. - Revision where other devices or treatments have failed. - Correction of functional deformity. - Fractures of the humeral head (with Short Humeral Stems). - Fractures of the humeral head and proximal humerus, where other methods of treatments are deemed inadequate (with Standard or Long Stems). - Difficult clinical management problems where other methods of treatment may not be suitable or may be inadequate. Fixation Methods The humeral stem is intended for cemented or cementless use. The humeral stemless anchor is intended for cementless use. The anatomic glenoid and anatomic hybrid glenoid are intended for cemented use only. The baseplate components of the convertible glenoids are intended for cementless application with the addition of screw fixation. When a convertible glenoid liner is used, the baseplate is indicated for use with a central screw and three peripheral locking screws that are 25mm or greater in length. The peripheral non-locking screws, peripheral posts and central posts are not indicated for use with the convertible glenoid. Reverse Total Shoulder The INHANCE SHOULDER SYSTEM Reverse Total Shoulder with a humeral stem is indicated for primary, fracture or revision total reverse shoulder replacement procedures to address the following. The system is indicated for use in patients whose shoulder joint has a gross rotator cuff deficiency. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The system is also indicated for conversion from an anatomic to reverse shoulder prosthesis without the removal of a well-fixed INHANCE humeral stem. - A severely painful, disabling, arthritic joint - Fractures of the humeral head (with Short Humeral Stems) - Fractures of the humeral head and proximal humerus (with Standard or Long Stems) - Revision of previously failed shoulder joint replacement Fixation Methods The humeral stem is intended for cemented or cementless use. The glenoid baseplate components are intended for cementless application with the addition of screw fixation. When a Peripheral Post is used, the baseplate is indicated for use with a Central Screw and two Peripheral Locking Screws that are 25mm or greater in length. The peripheral Non-Locking screws and Central Post are not indicated for use with the Peripheral Post.

Device Story

Reverse shoulder prosthesis system; replaces damaged shoulder joint components to restore function. System includes metal humeral stem, shell, glenosphere, baseplate, and cross-linked VE UHMWPE liner. Subject device adds alternate angle humeral liners (10-degree inclination) to existing system. Implants surgically placed by orthopedic surgeons in clinical/OR setting. Provides mechanical articulation for patients with rotator cuff deficiency; improves joint mobility and reduces pain. Output is physical joint reconstruction; clinical decision-making based on patient anatomy and pathology.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing including construct fatigue, wear characteristics, geometric comparison, range of motion (ASTM F1378), MR safety (ASTM F2052, F2119, F2182, F2213), biocompatibility (ISO 10993-1), sterilization validation (ISO 11137), and packaging integrity (ISO 11607).

Technological Characteristics

Materials: Titanium alloy (stem, shell, baseplate, screws, augment), Cobalt-chromium (glenosphere), Cross-linked VE UHMWPE (liner). Principle: Semi-constrained cemented/cementless prosthesis. Sterilization: ISO 11137-1/2. Connectivity: N/A. Form factor: Modular components with 32mm, 36mm, 40mm diameter options and 10-degree inclination liners.

Indications for Use

Indicated for patients requiring anatomic total/hemi-shoulder or reverse total shoulder replacement. Anatomic indications: osteoarthritis, post-traumatic arthrosis, avascular necrosis, rheumatoid arthritis, functional deformity, fractures, or revision. Reverse indications: gross rotator cuff deficiency, severely painful/disabling arthritic joint, fractures, or revision. Requires functional deltoid muscle and suitable anatomy.

Regulatory Classification

Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,” (vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 11, 2026 Depuy Ireland UC % Yayoi Fujimaki Regulatory Affairs Technical Manager DePuy Orthopaedics, Inc. 700 Orthopaedic Dr., Warsaw, Indiana 46582 Re: K253624 Trade/Device Name: INHANCE™ Reverse Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, KWT, MBF, PAO Dated: February 12, 2026 Received: February 12, 2026 Dear Yayoi Fujimaki: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1} K253624 - Yayoi Fujimaki Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn {2} K253624 - Yayoi Fujimaki Page 3 (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, FARZANA SHARMIN -S Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K253624 Device Name INHANCE™ Reverse Shoulder System Indications for Use (Describe) Anatomic Total Shoulder or Hemi-Shoulder The INHANCE SHOULDER SYSTEM with the humeral stemless anchor is intended for use in anatomic total shoulder replacement procedures to address the following: - Osteoarthritis - Post-traumatic arthrosis - Focal avascular necrosis of the humeral head - Previous surgeries of the shoulder that do not compromise the fixation The INHANCE SHOULDER SYSTEM with a humeral stem is intended for use in anatomic total or hemi-shoulder replacement procedures to address the following: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. - Rheumatoid arthritis. - Revision where other devices or treatments have failed. - Correction of functional deformity. - Fractures of the humeral head (with Short Humeral Stems). - Fractures of the humeral head and proximal humerus, where other methods of treatments are deemed inadequate (with Standard or Long Stems). - Difficult clinical management problems where other methods of treatment may not be suitable or may be inadequate. ## Fixation Methods The humeral stem is intended for cemented or cementless use. The humeral stemless anchor is intended for cementless use. The anatomic glenoid and anatomic hybrid glenoid are intended for cemented use only. The baseplate components of the convertible glenoids are intended for cementless application with the addition of screw fixation. When a convertible glenoid liner is used, the baseplate is indicated for use with a central screw and three peripheral locking screws that are 25mm or greater in length. The peripheral non-locking screws, peripheral posts and central posts are not indicated for use with the convertible glenoid. ## Reverse Total Shoulder The INHANCE SHOULDER SYSTEM Reverse Total Shoulder with a humeral stem is indicated for primary, fracture or revision total reverse shoulder replacement procedures to address the following. The system is indicated for use in patients whose shoulder joint has a gross rotator cuff deficiency. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The system is also indicated for conversion from an anatomic to reverse shoulder prosthesis without the removal of a well-fixed INHANCE humeral stem. - A severely painful, disabling, arthritic joint - Fractures of the humeral head (with Short Humeral Stems) - Fractures of the humeral head and proximal humerus (with Standard or Long Stems) - Revision of previously failed shoulder joint replacement ## Fixation Methods FORM FDA 3881 (6/20) PEC Publishing Services (301) 443-6740 {4} The humeral stem is intended for cemented or cementless use. The glenoid baseplate components are intended for cementless application with the addition of screw fixation. When a Peripheral Post is used, the baseplate is indicated for use with a Central Screw and two Peripheral Locking Screws that are 25mm or greater in length. The peripheral Non-Locking screws and Central Post are not indicated for use with the Peripheral Post. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 2 of 2 {5} 510(k) #: K253624 510(k) Summary Prepared on: 2026-03-11 # Contact Details 21 CFR 807.92(a)(1) | Applicant Name | DePuy Ireland UC | | --- | --- | | Applicant Address | Loughbeg, Ringaskiddy Co. Cork Munster Ireland | | Applicant Contact Telephone | 617 837 1918 | | Applicant Contact | Ms. Yayoi Fujimaki | | Applicant Contact Email | yfujima1@its.jnj.com | | Correspondent Name | DePuy Orthopaedics, inc. | | Correspondent Address | 700 Orthopaedic Drive N/A Warsaw IN 46582 United States | | Correspondent Contact Telephone | 617 837 1918 | | Correspondent Contact | Ms. Yayoi Fujimaki | | Correspondent Contact Email | yfujima1@its.jnj.com | # Device Name 21 CFR 807.92(a)(2) | Device Trade Name | INHANCE™ Reverse Shoulder System | | --- | --- | | Common Name | Shoulder joint metal/polymer semi-constrained cemented prosthesis | | Classification Name | Shoulder Prosthesis, Reverse Configuration | | Regulation Number | 888.3660 | | Product Code(s) | PHX, KWS, KWT, MBF, PAO | # Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | --- | --- | --- | | K212737 | INHANCE™ Reverse Shoulder System | PHX | | K192855 | DELTA XTEND™ Reverse Shoulder System | PHX | | K122698 | AEQUALIS™ ASCEND™ FLEX Reverse Shoulder System | PHX | # Device Description Summary 21 CFR 807.92(a)(4) The INHANCE Reverse Total Shoulder System consists of individually packaged implants: a metal humeral stem (titanium alloy), a shell (titanium alloy), a liner (Cross-Linked, VE UHMWPE) in combination with a glenosphere (cobalt-chromium), a baseplate (titanium alloy), peripheral screws (titanium alloy), a baseplate augment (titanium alloy), and either a central screw (titanium alloy) or a central post (titanium alloy). The purpose of this submission is to add alternate angle humeral liners to the INHANCE reverse shoulder system. The subject device is manufactured from a Cross-linked, VE UHMWPE. Size options include diameter 32mm, 36mm and 40mm with thickness +0mm or +4mm. Standard and Retentive options are available. {6} Intended Use/Indications for Use 21 CFR 807.92(a)(5) Anatomic Total Shoulder or Hemi-Shoulder The INHANCE SHOULDER SYSTEM with the humeral stemless anchor is intended for use in anatomic total shoulder replacement procedures to address the following: - Osteoarthritis - Post-traumatic arthrosis - Focal avascular necrosis of the humeral head - Previous surgeries of the shoulder that do not compromise the fixation The INHANCE SHOULDER SYSTEM with a humeral stem is intended for use in anatomic total or hemi-shoulder replacement procedures to address the following: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. - Rheumatoid arthritis. - Revision where other devices or treatments have failed. - Correction of functional deformity. - Fractures of the humeral head (with Short Humeral Stems). - Fractures of the humeral head and proximal humerus, where other methods of treatments are deemed inadequate (with Standard or Long Stems). - Difficult clinical management problems where other methods of treatment may not be suitable or may be inadequate. ## Fixation Methods The humeral stem is intended for cemented or cementless use. The humeral stemless anchor is intended for cementless use. The anatomic glenoid and anatomic hybrid glenoid are intended for cemented use only. The baseplate components of the convertible glenoids are intended for cementless application with the addition of screw fixation. When a convertible glenoid liner is used, the baseplate is indicated for use with a central screw and three peripheral locking screws that are 25mm or greater in length. The peripheral non-locking screws, peripheral posts and central posts are not indicated for use with the convertible glenoid. ## Reverse Total Shoulder The INHANCE SHOULDER SYSTEM Reverse Total Shoulder with a humeral stem is indicated for primary, fracture or revision total reverse shoulder replacement procedures to address the following. The system is indicated for use in patients whose shoulder joint has a gross rotator cuff deficiency. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The system is also indicated for conversion from an anatomic to reverse shoulder prosthesis without the removal of a well-fixed INHANCE humeral stem. - A severely painful, disabling, arthritic joint - Fractures of the humeral head (with Short Humeral Stems) - Fractures of the humeral head and proximal humerus (with Standard or Long Stems) - Revision of previously failed shoulder joint replacement ## Fixation Methods The humeral stem is intended for cemented or cementless use. The glenoid baseplate components are intended for cementless application with the addition of screw fixation. When a Peripheral Post is used, the baseplate is indicated for use with a Central Screw and two Peripheral Locking Screws that are 25mm or greater in length. The peripheral Non-Locking screws and Central Post are not indicated for use with the Peripheral Post. ## Indications for Use Comparison 21 CFR 807.92(a)(5) The indications for use are the same as the predicate device. ## Technological Comparison 21 CFR 807.92(a)(6) The overall design of the subject humeral liner is similar to the predicate INHANCE humeral liner (K212737) with the key differentiation being the addition of a 10-degree inclination to the subject humeral liner. While the small technological difference is identified, the humeral neck-shaft angles of the INHANCE Reverse Shoulder System fall within the legally marketed predicate reverse shoulder system devices. The intended use and principles of operation remain unchanged. Device material, sterilization method, and sterile packaging barrier are the same as the predicate device. {7} Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non-clinical studies and analyses were performed. Applicable technical standards, guidance or well-established test methods were referenced for the evaluations. - Construct fatigue - Wear characteristics - Geometric Comparison - Range of Motion (ASTM F1378) - MR (magnetic resonance) safety (ASTM F2052, ASTM F2119, ASTM F2182, ASTM F2213) - Biocompatibility (ISO 10993-1) - Sterilization validation (ISO 11137-1, ISO 11137-2) - Packaging integrity and stability (ISO 11607-1, ISO 11607-2) Clinical Study was not applicable. Based on the comparison of the technological characteristics and the engineering analyses, it can be concluded that the subject device is substantially equivalent to the predicate. The difference in the technological characteristics did not raise any different questions in safety and performance.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...