← Product Code [PHX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX) · K211359

# Tornier Perform Patient-Matched Primary Reversed Glenoid and BLUEPRINT Patient Specific Instrumentation (K211359)

_Tornier, Inc. · PHX · Nov 12, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX/K211359

## Device Facts

- **Applicant:** Tornier, Inc.
- **Product Code:** [PHX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX.md)
- **Decision Date:** Nov 12, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3660
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Software as a Medical Device

## Intended Use

Tornier Perform™ Patient-Matched Primary Reversed Glenoid: The implant is intended for replacement of the shoulder joint to reduce pain and improve shoulder mobility in comparison with preoperative status. BLUEPRINT™ Patient Specific Instrumentation: BLUEPRINT™ Glenoid Guides: The BLUEPRINT™ Glenoid Guides are intended to be used as surgical instruments to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans. BLUEPRINT™ 3D Planning Software: BLUEPRINT™ 3D Planning Software is an application that helps a surgeon plan their patients' shoulder prosthesis surgery. When possible, it generates the information required to produce patient specific components.

## Device Story

System comprises patient-matched reversed glenoid implant and BLUEPRINT™ patient-specific instrumentation (guides and 3D planning software). Surgeon uses software to process patient CT scans (DICOM format) to visualize, measure, and annotate shoulder anatomy; simulate range of motion; and plan implant positioning. Software generates planning report and design specifications for patient-specific glenoid components. Guides assist intraoperative positioning of glenoid components. Implant features baseplate with press-fit post, peripheral screws, and glenosphere for non-cemented fixation. Used by orthopedic surgeons in clinical settings for shoulder arthroplasty. Benefits include improved surgical planning and patient-specific implant fit.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical performance testing including baseplate pullout, fatigue, reverse glenoid loosening, range of motion, MRI compatibility, and porous structure evaluation. Biocompatibility, sterilization, and shelf-life assessed per consensus standards.

## Technological Characteristics

Implant: Baseplate with press-fit post, peripheral screws, glenosphere. Software: 3D planning application, DICOM-based, connected to Online Management System (OMS). Hardware: Patient-specific guides manufactured based on preoperative CT plan. Sterilization: Delivered sterile.

## Regulatory Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Predicate Devices

- AEQUALIS™ PERFORM Reversed Glenoid ([K161742](/device/K161742.md))
- BLUEPRINT™ Patient Specific Instrumentation ([K203315](/device/K203315.md))

## Reference Devices

- ZIMMER BIOMET Comprehensive Vault Reconstruction System ([K152754](/device/K152754.md))
- Exactech Equinoxe ([K110708](/device/K110708.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

November 12, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Tornier, Inc. Lisa Stahl Principal Specialist, Regulatory Affairs 10801 Nesbitt Ave South Bloomington, Minnesota 55437

## Re: K211359

Trade/Device Name: Tornier Perform™ Patient-Matched Primary Reversed Glenoid and BLUEPRINT™ Patient Specific Instrumentation Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, QHE Dated: October 8, 2021 Received: October 12, 2021

## Dear Lisa Stahl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Jiping Chen, PhD Acting Division Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known) K211359

#### Device Name

Tornier Perform™ Patient-Matched Primary Reversed Glenoid and BLUEPRINT™ Patient Specific Instrumentation

#### Indications for Use (Describe)

Tornier Perform™ Patient-Matched Primary Reversed Glenoid

The Tornier Perform™ Patient-Matched Primary Reversed Glenoid implant is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator with pain disabled by:

- · Rheumatoid arthritis
- · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- · Correction of functional deformity
- · Fractures of the humeral head
- · Traumatic arthritis
- · Revision of glenohumeral joint if sufficient native glenoid bone remains

All components are single use.

The Patient-Matched Glenoid implant is anchored to the bone with screws and is for non-cemented fixation. Note: A CT Scan is used to create the Tornier Perform Patient-Matched Primary Reversed Glenoid implant.

BLUEPRINT™ Patient Specific Instrumentation

#### BLUEPRINT ™ Glenoid Guides

The BLUEPRINT™ Glenoid Guides are intended to be used as surgical instruments to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans.

#### BLUEPRINT™ 3D Planning Software

BLUEPRINT™ 3D Planning Software is a medical device for surgeons.

BLUEPRINT™ 3D Planning Software is intended to be used as a pre-surgical planner for shoulder replacement surgerv. BLUEPRINT™ 3D Planning Software requires CT scan images showing the anatomical shoulder structure in a DICOM format.

BLUEPRINT™ 3D Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. BLUEPRINT™ 3D Planning Software allows surgeons to design patient specific components (Patient-Specific instruments and Tornier Perform™ Patient-Matched Primary Reversed Glenoid*) based on the pre-surgical plan. BLUEPRINT™ 3D Planning Software leads to the generation of a planning report.

BLUEPRINT™ 3D Planning Software is to be used for adult men and women patients only whose bone maturity is

reached and should not be used for diagnostic purpose.

Note: Measures and patient specific guide design are provided depending on the case profiles.

*Only if Patient-Specific instruments or Tornier Perform™ Patient-Matched Primary Reversed Glenoid are available in your geography.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "TORNIER" in a large, sans-serif font, colored in blue. To the right of the word is a hexagonal shape, also in blue, containing a stylized letter "T" within it. The logo appears to be a corporate logo or brand identity.

Date Prepared: October 8, 2021

# Administrative Information

|          | Tornier Perform Patient-Matched Primary Reversed Glenoid (Implant) |
|----------|--------------------------------------------------------------------|
| Name:    | Tornier, Inc.                                                      |
| Address: | 10801 Nesbitt Avenue South                                         |
|          | Bloomington, MN 55437                                              |
|          | United States of America                                           |

#### BLUEPRINT Patient Specific Instrumentation

| Name:           | Tornier SAS                                               |
|-----------------|-----------------------------------------------------------|
| Address:        | 161 rue Lavoisier<br>38330 Montbonnot Saint Martin-France |
| Contact Person: | Lisa Stahl                                                |
| Title:          | Principal Specialist, Regulatory Affairs                  |
| Phone (mobile): | 612-849-9970                                              |
| Fax:            | 952-426-7601                                              |

# Device Information

## Tornier, Inc.

| Name of Device:   | Tornier Perform™ Patient-Matched Primary Reversed Glenoid |
|-------------------|-----------------------------------------------------------|
| Common Name (s):  | Shoulder Prosthesis                                       |
| Regulatory Class: | II                                                        |
| Regulation:       | 21 CFR 888.3660                                           |
| Product Codes:    | PHX, KWS                                                  |

# Tornier SAS

| Name of Device:   | BLUEPRINT™ Patient Specific Instrumentation             |
|-------------------|---------------------------------------------------------|
| Common Name (s):  | Patient Specific Instrumentation + 3D Planning Software |
| Regulatory Class: | II                                                      |
| Regulation:       | 21 CFR 888.3660                                         |
| Product Codes:    | KWS, QHE                                                |

# Predicate Device Information

| Tornier, Inc.      |                                                                                                |
|--------------------|------------------------------------------------------------------------------------------------|
| Primary Predicate: | AEQUALIS™ PERFORM Reversed Glenoid, K161742                                                    |
| Reference Device:  | ZIMMER BIOMET Comprehensive Vault Reconstruction System, K152754<br>Exactech Equinoxe, K110708 |
| Tornier SAS        |                                                                                                |

| Primary Predicate: | BLUEPRINT™ Patient Specific Instrumentation, K20331. |
|--------------------|------------------------------------------------------|
|--------------------|------------------------------------------------------|

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "TORNIER" in a large, bold, blue font. To the right of the word is a blue hexagon with a stylized "T" inside. The "T" is also blue and has a smaller hexagon inside of it.

# Device Description

### Tornier Perform™ Patient-Matched Primary Reversed Glenoid

The Tornier Perform™ Patient-Matched Primary Reversed Glenoid implant (Patient-Matched Glenoid) is intended to replace the native glenoid surface of the scapulohumeral joint as part of a reverse shoulder prosthesis.

The glenoid implant is composed of a baseplate with a press-fit post, peripheral anchoring screws, and a glenosphere. Ancillary instruments are also provided for the implantation of the prosthesis.

#### BLUEPRINT™ Patient Specific Instrumentation

BLUEPRINT™ Patient Specific Instrumentation is composed of two components: BLUEPRINT™ Glenoid Guides (hardware) and BLUEPRINT™ 3D Planning Software (software).

BLUEPRINT™ Patient Specific Instrumentation which includes the BLUEPRINT™ Glenoid Guides and BLUEPRINT™ 3D Planning Software is the responsibility of Tornier is the legal manufacturer for the hardware and the software.

#### BLUEPRINT ™ Glenoid Guides

The BLUEPRINT ™ Glenoid Guides are patient-specially designed to facilitate the implantation of WRIGHT-TORNIER glenoid prostheses.

The BLUEPRINT™ Glenoid Guides are designed and manufactured based on a pre-operative plan generated only by the software BLUEPRINT™ 3D Planning Software.

#### BLUEPRINT™ 3D Planning Software

BLUEPRINT™ 3D Planning Software is a software connected to an Online Management System (OMS). The user interface software is installed on a computer is intended to be used by orthopedic surgeons, as a preoperative planning software for shoulder arthroplasty surgery (anatomic and reversed).

It is intended to help to plan an operation by allowing surgeons to:

- . Plan for shoulder arthroplasty cases
- . Position and select glenoid and humeral implants,
- Simulate the prosthetic range of motion,
- Interact with implants and different computed measurements
- . Generate information required to design a patient-specific glenoid component when appropriate.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Tornier. The word "TORNIER" is written in large, blue, sans-serif font. To the right of the word is a blue hexagon containing a stylized letter "T" inside of another stylized letter "T".

# Intended Use

### Tornier Perform™ Patient-Matched Primary Reversed Glenoid

The implant is intended for replacement of the shoulder joint to reduce pain and improve shoulder mobility in comparison with preoperative status.

### BLUEPRINT™ Patient Specific Instrumentation

### BLUEPRINT ™ Glenoid Guides

The BLUEPRINT™ Glenoid Guides are intended to be used as surgical instruments to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patientspecific preoperative CT scans.

### BLUEPRINT™ 3D Planning Software

BLUEPRINT™ 3D Planning Software is an application that helps a surgeon plan their patients' shoulder prosthesis surgery. When possible, it generates the information required to produce patient specific components.

# Indications for Use

### Tornier Perform™ Patient-Matched Primary Reversed Glenoid

The Tornier Perform™ Patient-Matched Primary Reversed Glenoid implant is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and nonrepairable rotator cuff-tear with pain disabled by:

- Rheumatoid arthritis ●
- . Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- Correction of functional deformity .
- Fractures of the humeral head ●
- Traumatic arthritis ●
- . Revision of glenohumeral joint if sufficient native glenoid bone remains

All components are single use.

The Patient-Matched Glenoid implant is anchored to the bone with screws and is for non-cemented fixation.

Note: A CT Scan is used to create the Tornier Perform Patient-Matched Primary Reversed Glenoid implant.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "TORNIER" in a large, blue, serif font. To the right of the word is a blue hexagon containing a stylized letter "T" inside of a smaller hexagon. The logo appears to be for a company or organization called "Tornier".

#### BLUEPRINT™ Patient Specific Instrumentation

#### BLUEPRINT ™ Glenoid Guides

The BLUEPRINT™ Glenoid Guides are intended to be used as surgical instruments to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans.

#### BLUEPRINT™ 3D Planning Software

BLUEPRINT™ 3D Planning Software is a medical device for surgeons.

BLUEPRINT™ 3D Planning Software is intended to be used as a pre-surgical planner for shoulder replacement surgery.

BLUEPRINT™ 3D Planning Software requires CT scan images showing the anatomical shoulder structure in a DICOM format.

BLUEPRINT™ 3D Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data.

BLUEPRINT™ 3D Planning Software allows surgeons to design patient specific components (Patient-Specific instruments and Tornier Perform™ Patient-Matched Primary Reversed Glenoid*) based on the pre-surgical plan.

BLUEPRINT™ 3D Planning Software leads to the generation of a planning report.

BLUEPRINT™ 3D Planning Software is to be used for adult men and women patients only whose bone maturity is reached and should not be used for diagnostic purpose.

Note: Measures and patient specific guide design are provided depending on the case profiles.

*Only if Patient-Specific instruments or Tornier Perform™ Patient-Matched Primary Reversed Glenoid are available in your geography.

# Comparison to Predicate Device

#### Tornier Perform™ Patient-Matched Primary Reversed Glenoid

The TORNIER PERFORM™ Patient-Matched Primary Reversed Glenoid and the predicate AEQUALISTM PERFORM Reversed Glenoid have the same intended use, same principle of operation, and similar technological features. The subject device patient-specific augmentation.

#### BLUEPRINT™ Patient Specific Instrumentation

The subject device BLUEPRINT™ Patient Specific Instrumentation and the predicate device BLUEPRINT™ Patient Specific Instrumentation (K203315) have the same intended use, similar principal of operation and similar general technological features.

Differences for subject BLUEPRINTTM Patient Specific Instrumentation include:

- . For the hardware: New guides dedicated to Tornier Perform™ Patient-Matched Primary Reversed Glenoid with nonfunctional dimensions changes, delivered sterile.
- For the software, planning for Tornier Perform™ Patient-Matched Primary Reversed Glenoid ● additional measurement and screw planning.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "TORNIER" in large, blue, sans-serif font, followed by a blue hexagon containing a stylized letter "T" with a smaller hexagon inside it. The word "TORNIER" is the main focus, with the hexagon and "T" symbol serving as a logo or emblem. The overall design is clean and professional, suggesting a corporate or institutional identity.

# Non-clinical Performance Testing

#### Tornier Perform™ Patient-Matched Primary Reversed Glenoid

Non-clinical testing was performed to demonstrate substantial equivalence to the predicate device.

- Baseplate Pullout ●
- . Fatigue
- . Reverse Glenoid Loosening
- Range of Motion
- MRI compatibility evaluation .
- . Porous Structure

Biocompatibility, sterilization, cleaning, endotoxin, particulate, packaging, shelf life, and distribution for the Tornier Perform "M Patient-Matched Primary Reversed Glenoid components were also assessed in accordance with recognized consensus standards.

# BLUEPRINT™ Patient Specific Instrumentation

Technological differences between the subject and predicate hardware devices are supported by the dimensional test performed on the predicate device hardware which remain applicable to the subject hardware device, as the changes to the subject device do not impact functional dimensions or material, and cadaveric test performed on the subject device.

Technological differences between the subject and predicate software devices are supported with verification and validation evaluations. The operating principle of the subject device is that of the predicate device.

The differences in design specifications do not raise new questions of safety and effectiveness over the predicate device as demonstrated in validation testing.

# Clinical Testing

No clinical studies were performed.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the word "TORNIER" in large, blue, sans-serif font. To the right of the word is a blue hexagon containing a stylized letter "T". The letter "T" is also blue and is made up of two smaller hexagons.

# Conclusions

## Tornier Perform™ Patient-Matched Primary Reversed Glenoid

The Tornier Perform™ Patient-Matched Primary Reversed Glenoid does not raise new questions of safety or effectiveness. Differences in technological characteristics have been addressed with performance testing. The results of performance testing for the Tornier Perform™ Patient-Matched Primary Reversed Glenoid support substantial equivalence to the primary predicate AEQUALIS™ PERFORM™ Reversed Glenoid (K161742, cleared November 15, 2016).

### BLUEPRINT™ Patient Specific Instrumentation

The BLUEPRINT™ Patient Specific Instrumentation does not raise new questions of safety or effectiveness. Differences in technological characteristics have been addressed with verification and validation testing. The results support substantial equivalence to the predicate BLUEPRINT™ Patient Specific Instrumentation (K203315, cleared April 15, 2021).

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX/K211359](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX/K211359)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
