The Reverse Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with a grossly deficient rotator cuff shoulder joint with severe arthropathy failed joint replacement and a grossly rotator cuff deficient shoulder joint. The patient's joint must be anatomically suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device. The glenoid baseplate is intended for cementless application with the addition of screws for fixation.
Device Story
Threaded Glenoid Baseplate; component of Medacta Shoulder System: Reverse; used for reverse shoulder arthroplasty. Implanted in glenoid bone via central threaded post and polyaxial locking screws; glenoidsphere attached via Morse taper and securing screw. Replaces glenoid side of glenohumeral joint. Used in clinical setting by orthopedic surgeons. Provides stable fixation for patients with rotator cuff deficiency; restores joint function; enables shoulder movement.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing: fatigue testing per ASTM F1378-12, micromotion assessment per ASTM F2028-14, and coating characterization. Pyrogenicity testing performed per USP <85> and <151>.
Technological Characteristics
Titanium Alloy (Ti6-Al 4-V) per ISO 5832-3:1996; Titanium Y367 coating per ASTM F1580. Threaded post design for cementless fixation with polyaxial locking screws. Morse taper connection for glenoidsphere. Sterilized via Gamma radiation.
Indications for Use
Indicated for patients with humeral fractures, severe arthropathy with grossly deficient rotator cuff, or failed joint replacement requiring primary or revision total shoulder replacement. Requires anatomically suited joint and functional deltoid muscle.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
December 13, 2017
Medacta International SA % Elizabeth Rose Manager, Regulatory Affairs Mapi USA, Inc. 2343 Alexandria Drive, Suite 100 Lexington, Kentucky 40504
Re: K171058
Trade/Device Name: Medacta Shoulder System: Threaded Glenoid Baseplate Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX Dated: November 10, 2017 Received: November 13, 2017
Dear Ms. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K171058
### Device Name
Medacta Shoulder System: Threaded Glenoid Baseplate
Indications for Use (Describe)
The Reverse Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with a grossly deficient rotator cuff shoulder joint with severe arthropathy failed joint replacement and a grossly rotator cuff deficient shoulder joint.
The patient's joint must be anatomically suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.
The glenoid baseplate is intended for cementless application with the addition of screws for fixation.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Medacta Shoulder System: Threaded Glenoid Baseplate Traditional 510(k)
## 3.0 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory Affairs Manager Date Prepared: April 07, 2017 Date Revised: December 11, 2017
#### II. Device
| Device Proprietary Name: | Medacta Shoulder System: Threaded Glenoid Baseplate |
|--------------------------|-----------------------------------------------------------------------|
| Common or Usual Name: | Shoulder Prosthesis, Reverse Configuration |
| Classification Name: | Shoulder joint metal/polymer semi-constrained cemented<br>prosthesis. |
| Primary Product Code: | PHX |
| Regulation Number: | 21 CFR 888.3660 |
| Device Classification | 2 |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices: Primary Predicate:
- Aequalis Reversed Shoulder Prosthesis, K132285, Tornier, SAS ●
- Encore Reverse Shoulder, K041066, Encore Medical, L.P. ●
Reference Devices
- Evolis Total Knee System, K081023, Medacta International, SA ●
- Medacta Shoulder System, K170452, Medacta International, SA
#### IV. Device Description
The purpose of this submission is to gain clearance for the new Medacta Shoulder System: Threaded Glenoid Baseplate (also referred to as Threaded Glenoid Baseplate) which is part of the Medacta Shoulder System: Reverse; a modular system intended to be used for Reverse Shoulder Arthroplasties (RSA). The Medacta Shoulder System: Threaded Glenoid Baseplate is
{4}------------------------------------------------
an alternative option to the pegged glenoid baseplate that is part of reference device Medacta Shoulder System (K170452). The Medacta Shoulder System: Threaded Glenoid Baseplate is made of titanium alloy.
The Threaded Glenoid Baseplate is intended to replace only the glenoid side of the glenohumeral joint. The Threaded Glenoid Baseplate is intended to be used in the reverse configuration only. The Threaded Glenoid Baseplate is designed to be fixed on the glenoid bone by means of a central threaded post and the help of Glenoid Polyaxial Locking Screws. The Glenoidsphere is intended to be assembled with a Glenoid Baseplate by means of a taper Morse connection and secured by a central securing screw. The Threaded Glenoid Baseplate directly couples with the Glenoid Polyaxial Locking Screws and Glenoidsphere as part of the Medacta Shoulder System (K170452).
The Medacta Shoulder System is similar to predicate devices Tornier's Aequalis Reversed Shoulder Prosthesis (K132285), Encore's Reverse Shoulder (K041066) and reference device Medacta's Shoulder System (K170452).
#### V. Indications for Use
The Reverse Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with a grossly deficient rotator cuff shoulder joint with severe arthropathy or a previously failed joint replacement and a grossly rotator cuff deficient shoulder joint. The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.
The glenoid baseplate is intended for cementless application with the addition of screws for fixation.
#### VI. Comparison of Technological Characteristics
The Medacta Shoulder System: Threaded Glenoid Baseplate and the predicate devices share the following characteristics:
- Indications for Use
- Materials ●
- Design ●
- Sterile ●
- Coating ●
- Threaded ●
- Method of Fixation
- Lengths
The Medacta Shoulder System: Threaded Glenoid Baseplate is technologically different from the predicate devices as follows:
{5}------------------------------------------------
#### Diameters .
The Medacta Shoulder System: Threaded Glenoid Baseplate is manufactured from Titanium Alloy (Ti6-Al 4-V) according to ISO 5832-3:1996 Implants For Surgery - Metallic Materials -Part 3: Wrought Titanium 6-Aluminum 4-Vanadium Alloy with a Titanium Y367 coating.
These materials have an extensive amount of biocompatibility history that has been provided in the listed predicate devices' 510(k) submissions; Tornier's Aequalis Reversed Shoulder Prosthesis (K132285). Encore's Reverse Shoulder (K041066) and reference device Medacta's Shoulder System (K170452). The Titanium Y367 coating used on the Medacta Shoulder System: Threaded Glenoid Baseplate is provided by Eurocoating S.p.A. The Titanium Y367 coating was previously reviewed in reference device Medacta's Evolis Total Knee System (K081023).
Additional biocompatibility testing was deemed unnecessary because the materials are the same or similar to the predicate devices and follow standards for manufacturing.
A comparison of the subject and predicate devices are provided in the table below.
| Parameters | Medacta Shoulder<br>System<br>(Subject Device) | Tornier Aequalis<br>Reversed Shoulder<br>Prosthesis<br>K132285<br>(Predicate Device) | Encore Reverse<br>Shoulder<br>K041066<br>(Predicate Device) |
|-----------------------|------------------------------------------------|--------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Material | Titanium Alloy | Titanium Alloy | Titanium Alloy |
| Coating | Titanium<br>(ASTM F1580) | Titanium<br>(ASTM F1580) | Hydroxyapatite<br>(ASTM F1185)<br>3DMatrix Coating |
| Diameter | Ø24.5 mm and Ø27 mm | Ø25 mm and Ø29 mm | Ø26 mm |
| Post Length | 25 mm, 30 mm and 35 mm | 20 mm, 25 mm, 30 mm, 35 mm, 40 mm,<br>45 mm and 50 mm | 30 mm |
| Method of<br>Fixation | Uncemented | Uncemented | Uncemented |
| Post Design | Threaded | Threaded | Threaded |
| Device usage | Single Use | Single Use | Single Use |
| Biocompatibility | Implant with<br>permanent >30 day | Implant with<br>permanent >30 day | Implant with<br>permanent >30 day |
| Sterilization | Gamma | Gamma | Gamma |
Technological comparison
{6}------------------------------------------------
## Discussion
As seen above, the technological differences between the subject and predicate devices do not raise new questions of safety and effectiveness. The Medacta Shoulder System: Threaded Glenoid Baseplate is the same or similar to the predicate devices in terms of intended use, materials of construction, design, coating, threaded, method of fixation, lengths and sterility. Based on the comparison of technological characteristics and performance data provided within this submission, the data supports the substantial equivalence of the Medacta Shoulder System: Threaded Glenoid Baseplate to the identified predicate devices.
#### Performance Data VII.
The following mechanical tests are being provided in support of a substantial equivalence determination. Based on the risk analysis, testing was conducted to written protocols with acceptance criteria that were based on standards.
Non-Clinical Studies
- Performance Tests ●
- Fatigue Testing: ASTM F1378-12: Standard Specification For Shoulder O Prostheses
- Micromotions Assessment In Reverse Configuration: ASTM F2028-14: Standards o Test Methods For Dynamic Evaluation of Glenoid Loosening Or Disassociation Prostheses
- Coating Tests ●
- o Glenoid Baseplate: Characterization Report Titanium Y367 Coating on Threaded Glenoid Baseplate Medacta
- Pyrogenicity
- o Medacta uses both the Bacterial Endotoxin Test (LAL test) according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>) and the Pyrogen Test according to USP chapter <151> for pyrogenicity determination; and
- Medacta has no intentions of labeling the subject devices as non-pyrogenic or o pyrogen free.
Clinical Studies
- No clinical studies were conducted. ●
#### Conclusion VIII.
The information provided above supports that the Medacta Shoulder System: Threaded Glenoid Baseplate is as safe and effective as the predicate devices. Therefore, it is concluded that the Medacta Shoulder System: Threaded Glenoid Baseplate is substantially equivalent to the predicate devices.