← Product Code [PHX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX) · K102951

# EXACTECH EQUINOX FIXED ANGLE REPLICATOR PLATE (0MM,1.5MM,4.5MM), HUMERAL HEAD-SHORT (44MM,47MM), GLENOSPHERE (38MM,42MM) (K102951)

_Exactech, Inc. · PHX · Nov 4, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX/K102951

## Device Facts

- **Applicant:** Exactech, Inc.
- **Product Code:** [PHX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX.md)
- **Decision Date:** Nov 4, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3660
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemiarthroplasty is determined by the surgeon to be the preferred method of treatment. The cemented primary humeral stem, long/revision stem, fracture stems and all Equinoxe glenoids are intended for cemented fixation. The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the discretion the surgeon. The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon. Humeral Heads are intended for use in cemented and press-fit applications.

## Device Story

Equinoxe Shoulder System comprises humeral stems, replicator plates, humeral heads, and glenospheres for hemi- and total-shoulder arthroplasty. Replicator plates connect to humeral stems via spherical bowl/torque-defining screw and to humeral heads via morse taper. System used in OR by orthopaedic surgeons to restore joint function in patients with degenerative disease or fractures. Replicator plates provide offset and angular adjustability for component positioning. Humeral heads and glenospheres provide articulating surfaces. Device benefits patients by restoring mobility and reducing pain associated with glenohumeral joint pathology. System supports cemented or press-fit fixation depending on component and surgeon preference.

## Clinical Evidence

Bench testing only. Fatigue testing conducted to verify safety and effectiveness of modified replicator plates and material changes for humeral heads and glenospheres. No clinical data provided.

## Technological Characteristics

Replicator plates: Wrought titanium alloy (Ti-6Al-4V E.L.I, ASTM F136). Humeral heads/Glenospheres: Cast Cobalt-Chrome alloy (Co-28Cr-6Mo, ASTM F75). Semi-constrained design. Fixation: Cemented or press-fit. Sterilization: Standard industry processes.

## Regulatory Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Predicate Devices

- Equinoxe Total Shoulder System ([K042021](/device/K042021.md))
- Equinoxe Reverse Total Shoulder System ([K063569](/device/K063569.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 2, 2016

Exactech, Incorporated Mr. Graham L. Cuthbert Regulatory Affairs Specialist II 2320 N.W. 66th Court Gainesville, Florida 32653

Re: K102951

N 102731
Trade/Device Name: Equinoxe Fixed-Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWT, KWS Dated: October 1, 2010 Received: October 5, 2010

Dear Mr. Cuthbert:

This letter corrects our substantially equivalent letter of November 4, 2010.

We have reviewed your Section 510(k) premarket notification of intent to market the device wo nave rollowed your wo determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the enactment date of the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

# Page 2 - Mr. Graham L. Cuthbert

(21 CFR Part 807); labeling (21 CFR-Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Lori A. Wiggins -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Exactech® Equinoxe Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - Indications for Use

# 510(k) Number (if known): K102951

Device Name: Equinoxe Fixed-Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres.

### INDICATIONS FOR USE:

Sign-Oft)
 Sign-Oft)
 Prative Devices

The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemiarthroplasty is determined by the surgeon to be the preferred method of treatment.

- The cemented primary humeral stem, long/revision stem, fracture stems and all 0 Equinoxe glenoids are intended for cemented fixation.
- The press-fit humeral stems are intended for press-fit applications but may be o used with bone cement at the discretion the surgeon.
- The reverse humeral components are intended to be used in cemented applications 0 or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon.
- 0 Humeral Heads are intended for use in cemented and press-fit applications.

Clinical indications for the PRIMARY (P), LONG/REVISION (L/R) and FRACTURE (F) humeral components are as follows:

|                                                                                                 |   |   | Indications                                                                                                               |
|-------------------------------------------------------------------------------------------------|---|---|---------------------------------------------------------------------------------------------------------------------------|
| <span style="writing-mode: vertical-rl; text-orientation: upright;">Stock Number K102951</span> | √ | √ | rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic<br>degenerative problems                            |
|                                                                                                 | √ | √ | congenital abnormalities in the skeletally mature                                                                         |
|                                                                                                 | √ | √ | primary and secondary necrosis of the humeral head.                                                                       |
|                                                                                                 | √ |   | humeral head fracture with displacement of the tuberosities                                                               |
|                                                                                                 | √ | √ | pathologies where arthrodesis or resectional arthroplasty of the<br>humeral head are not acceptable                       |
|                                                                                                 | √ | √ | revisions of humeral prostheses when other treatments or devices have<br>failed (where adequate fixation can be achieved) |
|                                                                                                 | - |   | displaced three-part and four-part upper humeral fractures                                                                |
|                                                                                                 | √ |   | spiral and other fractures of the mid-humerus (in combination with<br>glenohumeral degenerative diseases)                 |
|                                                                                                 | √ |   | revision of failed previous reconstructions when distal anchorage is<br>required                                          |
|                                                                                                 | √ | √ | to restore mobility from previous procedures (e.g. previous fusion)                                                       |

The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed

{3}------------------------------------------------

## Exactech® Equinoxe® Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - Indications for Use

glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.

The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemi-arthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.

X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

Please do not write below this line - use another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sonitu for mxn

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

and/or

510(k) Number K102951

{4}------------------------------------------------

K102951

1/5

### Exactech® Equinoxe® Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - Substantial Equivalence

| I. Company:           | Exactech, Inc.<br>2320 N.W. 66 <sup>th</sup> Court<br>Gainesville, FL 32653                                        |
|-----------------------|--------------------------------------------------------------------------------------------------------------------|
|                       | Phone: (352) 377-1140<br>Fax: (352) 378-2617                                                                       |
| Contact Person:       | Graham L. Cuthbert, Regulatory Affairs Specialist II                                                               |
| Date:                 | October 1 <sup>st</sup> , 2010                                                                                     |
| II. Proprietary Name: | Exactech Equinoxe Fixed Angle Replicator Plate<br>Exactech Equinoxe Humeral Heads<br>Exactech Equinoxe Glenosphere |
| Common Name:          | Total Shoulder Prosthesis<br>Reverse Total Shoulder Prosthesis                                                     |

NOV - 4 2010### Classification Name:

- Shoulder joint metal/polymer non-constrained cemented prosthesis (21 CFR . 888.3650, Class II, Product Code KWT)
- Prosthesis, Shoulder, Semi-constrained, metal/polymer cemented (21 CFR . 888.3660.

Class II, Product Code KWS)

Legally Marketed Devices to Which Substantial Equivalence Is Claimed: III.

| 510(k) Number | Trade of Proprietary Model Name        | Manufacturer  |
|---------------|----------------------------------------|---------------|
| K042021       | Equinoxe Total Shoulder System         | Exactech, Inc |
| K063569       | Equinoxe Reverse Total Shoulder System | Exactech, Inc |

#### IV. Device Description:

The Equinoxe Shoulder System is comprised of primary, fracture and reverse product lines for use in hemi- and total-shoulder joint replacement procedures. The Equinoxe Primary and Fracture system utilize both cemented and press-fit, semi-constrained glenohumeral prostheses for use in hemi-shoulder and totalshoulder joint replacement procedures, and a cemented semi-constrained glenohumeral fracture prosthesis for use in fractures of the proximal humerus.

The Equinoxe Reverse Shoulder System includes a reverse semi-constrained prosthesis for use in total-shoulder joint replacement procedures in cases with an irreparable or nonfunctional rotator cuff. The Equinoxe Reverse Shoulder System is designed to function with the Equinoxe primary press-fit, primary cemented, and long/revision humeral stems.

{5}------------------------------------------------

K102951 2/5

### Exactech® Equinoxe® Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - 510(k) Summary of Safety and Effectiveness

The proposed fixed angle replicator plates are manufactured from wrought titanium alloy (Ti-6Al-4V E.L.I) per ASTM F136 and connect to the humeral stem via a spherical bowl and are locked in place using the previously cleared primary torque-defining screw. The replicator plate attaches to the humeral head using a morse taper. The mating features on the proposed replicator plates (specifically the morse taper geometry and the spherical bowl geometry) are identical to the predicate part, only the offset and angular adjustability have been altered.

The proposed humeral heads are manufactured from Cast Cobalt-Chrome Alloy (Co-28Cr-6Mo) per ASTM F75 and provide an articulating surface for hemi and total joint replacement. They feature female taper geometry for mating with proposed and predicate replicator plate designs. No design changes have been made to this device. It is a material change only.

The proposed glenospheres are manufactured from Cast Cobalt-Chrome Alloy (Co-28Cr-6Mo) per ASTM F75and connect to the plate via an apical hole screw and slip fit with oval shaped plate. No design changes have been made to this device. It is a material change only.

#### Intended Use of the Device V.

The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemiarthroplasty is determined by the surgeon to be the preferred method of treatment.

- The cemented primary humeral stem, long/revision stem, fracture stems . and all Equinoxe glenoids are intended for cemented fixation.
- The press-fit humeral stems are intended for press-fit applications but may . be used with bone cement at the discretion the surgeon.
- The reverse humeral components are intended to be used in cemented ● applications or in revision cases when the humeral component is wellfixed/stable, as deemed by the orthopaedic surgeon.
- Humeral Heads are intended for use in cemented and press-fit . applications.

Clinical indications for the PRIMARY (P), LONG/REVISION (L/R) and FRACTURE (F) humeral components are as follows:

| P | L/R | F | Indications                                                                                                               |
|---|-----|---|---------------------------------------------------------------------------------------------------------------------------|
| √ | √   |   | rheumatoid arthritis, osteoarthritis, osteonecrosis or post-<br>traumatic degenerative problems                           |
| √ | √   |   | congenital abnormalities in the skeletally mature                                                                         |
| √ | √   |   | primary and secondary necrosis of the humeral head.                                                                       |
| √ |     | √ | humeral head fracture with displacement of the tuberosities                                                               |
| √ | √   |   | pathologies where arthrodesis or resectional arthro plasty of the<br>humeral head are not acceptable                      |
| √ | √   |   | revisions of humeral prostheses when other treatments or devices have<br>failed (where adequate fixation can be achieved) |
| √ |     | √ | displaced three-part and four-part upper humeral fractures                                                                |

{6}------------------------------------------------

K102951 3/5

# Exactech® Equinoxe® Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - 510(k) Summary of Safety and Effectiveness

| P | L/R | F | Indications                                                                                               |
|---|-----|---|-----------------------------------------------------------------------------------------------------------|
|   | √   |   | spiral and other fractures of the mid-humerus (in combination with<br>glenohumeral degenerative diseases) |
|   | √   |   | revision of failed previous reconstructions when distal anchorage is<br>required                          |
| √ | √   |   | to restore mobility from previous procedures (e.g. previous fusion)                                       |

The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.

The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemiarthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.

#### Rationale for Substantial Equivalence VI.

Equinoxe Fixed Angle Replicator Plate:

- Intended Use / Indications for Use. The Equinoxe Fixed Angle Replicator ● Plates and predicate devices are intended for use in total shoulder joint replacement and have identical indications for use.
- The Equinoxe Fixed Angle Replicator Plates and predicate � Materials. devices are composed of equivalent biocompatible materials conforming to recognized industry standards for permanent implants.
- Design Features. The Equinoxe Fixed Angle Replicator Plates and predicate ● devices have the same design features.
- The Equinoxe Fixed Angle Replicator Plates and predicate ● Dimensions. devices have identical geometry for mating with the previously cleared humeral stems and humeral heads. The only difference is a new offset option and a reduction in the through hole diameter to prevent angulation of the plate. This has no effect on the mechanical strength of the design.
- Packaging and Sterilization. The Equinoxe Fixed Angle Replicator Plates . and predicate devices are packaged and sterilized using the same materials and processes.
- Device Shelf Life. The Equinoxe Fixed Angle Replicator Plates and predicate . devices have the same shelf life.

{7}------------------------------------------------

K102951 4/5

Exactech® Equinoxe® Total Shoulder – Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - 510(k) Summary of Safety and Effectiveness

Cast Humeral Heads:

- Intended Use / Indications for Use. The Equinoxe Cast Humeral Heads and . predicate devices are intended for use in total shoulder joint replacement and have identical indications for use.
- Materials. The Equinoxe Cast Humeral Heads and predicate devices are . composed of Co-28Cr-6Mo Alloys which are biocompatible materials conforming to recognized industry standards for permanent implants.
- Design Features. The Equinoxe Cast Humeral Heads and predicate devices ● have the same design features.
- Dimensions. The Equinoxe Cast Humeral Heads and predicate devices have . identical geometry.
- Packaging and Sterilization. The Equinoxe Cast Humeral Heads and ● predicate devices are packaged and sterilized using the same materials and processes.
- Device Shelf Life. The Equinoxe Cast Humeral Heads and predicate devices ● have the same shelf life.

Cast Glenospheres:

- Intended Use / Indications for Use. The Equinoxe Cast Glenospheres and . predicate devices are intended for use in reverse total shoulder joint replacement and have identical indications for use.
- The Equinoxe Cast Glenospheres and predicate devices are Materials. . composed of Co-28Cr-6Mo Alloys which are biocompatible materials conforming to recognized industry standards for permanent implants.
- Design Features. The Equinoxe Cast Glenospheres and predicate devices . have the same design features.
- Dimensions. The Equinoxe Cast Glenospheres and predicate devices have . identical geometry.
- Packaging and Sterilization. The Equinoxe Cast Glenospheres and predicate ● devices are packaged and sterilized using the same materials and processes.
- Device Shelf Life. The Equinoxe Cast Glenospheres and predicate devices ● have the same shelf life.

VII. Summary of Non-Clinical Performance Data

> Fatigue testing was conducted to verify the safety and effectiveness of the proposed Exactech Equinoxe Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres. These tests were done to mitigate the risks that were affected by the proposed changes. Risks that were not affected by the new design were not re-verified.

### Substantial Equivalence Conclusion

Results from these tests and analyses provided within this 510(k) demonstrate that the Exactech Equinoxe Fixed Angle Replicator Plates, Cast Humeral Heads, and

{8}------------------------------------------------

K102951 5/5

Exactech® Equinoxe® Total Shoulder – Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) -- 510(k) Summary of Safety and Effectiveness

·

Cast Glenospheres are substantially equivalent to the identified predicate devices and will be safe for clinical use.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX/K102951](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PHX/K102951)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
