← Product Code [PEK](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PEK) · K140236

# AILERON INTERSPINOUS FIXATION SYSTEM (K140236)

_Life Spine, Inc. · PEK · Oct 20, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PEK/K140236

## Device Facts

- **Applicant:** Life Spine, Inc.
- **Product Code:** [PEK](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PEK.md)
- **Decision Date:** Oct 20, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing. The Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes in skeletally mature patients for the purpose of achieving, in conjunction with autogenous bone graft, single level supplemental fusion in the following conditions: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), (2) trauma (i.e., fracture or dislocation), (3) spinal tumor, (4) spondylolisthesis. The Aileron Interspinous Fixation System is not intended for stand-alone use.

## Device Story

Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device for the non-cervical spine. It consists of titanium alloy implants designed for plate fixation to spinous processes. Used by surgeons in clinical settings to stabilize and maintain spinal alignment during fusion procedures. It acts as a load-sharing device to support healing in conjunction with autogenous bone graft. It is not intended for stand-alone use or to replace normal body structures. The system is available in various sizes to accommodate patient anatomy.

## Clinical Evidence

Bench testing only. Performance data included static and dynamic compression and torsion testing per modified ASTM F1717, plus spike pull-off and static axial grip testing.

## Technological Characteristics

Posterior, non-pedicle supplemental fixation device. Materials: 6AL-4V-ELI titanium per ASTM F136. Single-use, non-sterile components. Mechanical fixation via plate attachment to spinous processes.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Aileron Interspinous Fixation System ([K113157](/device/K113157.md))
- Life Spine Interspinous Fixation System ([K100407](/device/K100407.md))
- Lanx Spinal Fixation System ([K090252](/device/K090252.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines above them. The profiles are arranged in a way that suggests unity and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 20, 2014

Life Spine, Incorporated Mr. Randy Lewis General Manager 2401 West Hassell Road, Suite 1535 Hoffman Estates, Illinois 60169

Re: K140236

Trade/Device Name: Aileron Interspinous Fixation System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: PEK Dated: September 24, 2014 Received: September 25, 2014

Dear Mr. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Ronald#Alean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K140236

Device Name Aileron Interspinous Fixation System

#### Indications for Use (Describe)

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing.

The Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes in skeletally mature patients for the purpose of achieving, in conjunction with autogenous bone graft, single level supplemental fusion in the following conditions: (1) degenerative discase (is defined as back pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), (2) trauma (i.e., fracture or dislocation), (3) spinal tumor, (4) spondylolisthesis. The Aileron Interspinous Fixation System is not intended for stand-alone use.

Type of Use (Select one or both, as applicable)

| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary Aileron Interspinous Fixation System

| Submitted By:     | Life Spine, Inc.<br>2401 W. Hassell Road, Suite 1535<br>Hoffman Estates, IL 60169<br>Telephone: 847-884-6117<br>Fax: 847-884-6118                                                      |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact:   | Randy Lewis<br>Life Spine<br>2401 W. Hassell Road, Suite 1535<br>Hoffman Estates, IL 60169<br>Telephone: 847-884-6117<br>Fax: 847-884-6118                                             |
| Date Prepared:    | September 24th, 2014                                                                                                                                                                   |
| Trade Name:       | Aileron Interspinous Fixation System                                                                                                                                                   |
| Common Name:      | Spinal interlaminal fixation orthosis                                                                                                                                                  |
| Classification:   | PEK, 888.3050, Class II, Spinal interlaminal fixation orthosis                                                                                                                         |
| Predicate Device: | Primary Predicate: Aileron Interspinous Fixation System (K113157)<br>Additional Predicates: Life Spine Interspinous Fixation System (K100407)<br>Lanx Spinal Fixation System (K090252) |

#### Device Description:

The Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1). Implants are manufactured from titanium alloy per ASTM F136 and are available in a range of sizes to suit the individual pathology and anatomical conditions of the patient. The purpose of this submission is to clear device modifications to the Aileron Interspinous Fixation System.

{4}------------------------------------------------

#### Intended Use of the Device:

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing.

The Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes in skeletally mature patients for the purpose of achieving, in conjunction with autogenous bone graft, single level supplemental fusion in the following conditions: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), (2) trauma (i.e., fracture or dislocation), (3) spinal tumor, (4) spondylolisthesis. The Aileron Interspinous Fixation System is not intended for stand-alone use.

## Technological Characteristics:

The Aileron Interspinous Fixation System is substantially equivalent to the predicate systems in terms of design, materials, and indications for use.

## Material:

The Aileron Interspinous Fixation System is 6AL-4V-ELI titanium manufactured according to ASTM F136. The device is comprised of a variety of non-sterile titanium, single use components.

#### Performance Data:

Static and dynamic compression and torsion testing in accordance with a modified ASTM F1717 setup, in addition to Spike Pull-off and Static Axial Grip testing, was presented to demonstrate substantial equivalency.

#### Conclusion:

The information presented demonstrates the substantial equivalency of the Aileron Interspinous Fixation System.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PEK/K140236](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PEK/K140236)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
