← Product Code [OWY](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OWY) · K222501

# Regeneten Bioinductive Implant (K222501)

_Smith and Nephew, Inc. · OWY · May 11, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWY/K222501

## Device Facts

- **Applicant:** Smith and Nephew, Inc.
- **Product Code:** [OWY](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OWY.md)
- **Decision Date:** May 11, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Regeneten Bioinductive Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

## Device Story

Resorbable type I bovine Achilles-derived collagen implant; provides collagen layer over injured tendons; creates barrier between tendon and surrounding tissue; large porosity and low tensile modulus facilitate recipient tissue ingrowth and natural remodeling; supports healing environment. Used in orthopedic procedures; arthroscopic configuration preloaded in cartridge; sterile, single-use. Benefits patient by promoting formation of tendon-like tissue and potentially changing course of tear progression.

## Clinical Evidence

No new clinical study conducted. Evidence consists of review of published clinical literature from investigator-initiated post-market studies to support performance claims. Previous animal study histological results from predicate submissions (K112423, K140300) are cited to support bioinduction and tissue formation claims.

## Technological Characteristics

Resorbable type I bovine Achilles-derived collagen; porous, soft, pliable, nonfriable structure. Sterile, single-use. No new consensus standards claimed.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- Rotation Medical Collagen Tendon Sheet ([K140300](/device/K140300.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Smith and Nephew Inc Jenna Horsley Senior Regulatory Affairs Manager 150 Minuteman Road Andover, Massachusetts 01810

## Re: K222501

Trade/Device Name: Regeneten Bioinductive Implant Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWY, ORQ Dated: April 11, 2023 Received: April 11, 2023

## Dear Jenna Horsley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yu-chieh Chiu -S

Digitally signed by Yu-chieh Chiu -S
Date: 2023.05.11 15:47:39-04'00'

Yu-Chieh Chiu Acting Assistant Director DHT6C: Division of Stereotaxic, Trauma and Restorative Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222501

Device Name Regeneten Bioinductive Implant

Indications for Use (Describe)

Regeneten Bioinductive Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5. 510(k) Summary

### 510(k) SUMMARY COMPLYING WITH 21 CFR 807.92

| Date Prepared:                           | April 11, 2022                                                                                          |
|------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Submitter Name:                          | Smith and Nephew Inc                                                                                    |
| Submitter Address:                       | 150 Minuteman Road<br>Andover, MA 01810                                                                 |
| FDA Establishment Owner/Operator Number: | 1020279                                                                                                 |
| FDA Establishment Registration Number:   | 3003604053                                                                                              |
| Submitter / Primary Contact:             | Jenna Horsley<br>Senior Regulatory Affairs Manager<br>(617) 218-7921<br>Jenna.Horsley@smith-nephew.com  |
| Alternate Contact:                       | Hamed Rahimi<br>Regulatory Affairs Specialist<br>(636) 628-6500<br>Email: hamed.rahimi@smith-nephew.com |
| Device Trade / Proprietary Name:         | Regeneten® Bioinductive Implant                                                                         |
| Device Common Name:                      | Surgical mesh                                                                                           |
| Device Classification Name:              | Mesh, Surgical, Collagen, Orthopaedics                                                                  |
| Regulation Medical Specialty:            | Office of General & Plastic Surgery                                                                     |
| Review Panel(s):                         | Office of General & Plastic Surgery                                                                     |
| Product Code(s):                         | OWY, ORQ                                                                                                |
| Regulation Numbers:                      | 21 CFR 878.3300                                                                                         |
| Submission Type:                         | Traditional 510(k)                                                                                      |
| Device Class:                            | Class II                                                                                                |
| Device Predicate:                        | Rotation Medical Collagen Tendon Sheet (K140300)                                                        |

{4}------------------------------------------------

### Device Description:

Regeneten Bioinductive Implant is a resorbable type I bovine Achilles-derived collagen implant that provides a layer of collagen over injured tendons. The device is designed to provide a layer between the tendon and the surrounding tissue during healing. The device features a large porosity design with a low tensile modulus to allow for recipient tissue ingrowth into the implant at the repair site for natural remodeling and the passive formation of collagen over injured tendons. The physical structure and placement of Regeneten Bioinductive Implant provides and supports an environment for healing. When hydrated, Regeneten Bioinductive Implant is an easy-to-use, soft, pliable, nonfriable, porous implant. Regeneten Bioinductive Implant is provided as a sterile, single-use device in a variety of sizes. The arthroscopic configuration of the Bioinductive implant is preloaded in a cartridge and packaged in a sterile, dual seal tray-in-tray configuration.

#### Intended Use:

The subject device and predicate device have the same intended use:

The Regeneten Bioinductive Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

#### Indications for Use:

The subject device has the same indication for use as the predicate device:

The Regeneten Bioinductive Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

## Summary of Technological Characteristics Compared to Predicate Device(s):

The subject device (Regeneten Bioinductive Implant) and the predicate device (Rotation Medical Collagen Tendon Sheet-DDI, K140300), are identical devices, having the same physical performance characteristics, packaging, and material composition. Past safety, bench testing, and animal study of the predicate device are directly applicable to the subject device. Previous verification testing completed is directly applicable to the subject device, including staple pull-out testing, hydrothermal transition temperature, endotoxin testing, and mechanical integrity.

The subject device has similar technological characteristics as the predicate device with the addition of performance claims of bioinduction, formations of tendon-like tissue, and change the course of tear progression (see the performance claims table in the device description section). The addition of this performance claim did not require any change in the device design, material composition, and packaging and this technological difference is supported by, valid, reproducible performance testing.

{5}------------------------------------------------

## Biocompatibility:

The subject device (Regeneten Bioinductive Implant) and the predicate device (Rotation Medical Collagen Tendon Sheet-DDI, K140300) are identical devices. Therefore, past biocompatibility tests included in 510(k) K112423 are applicable to the subject device. Routine monitoring of endotoxin is being performed for every production batch as part of the finished product release process.

## Performance Data (Bench Testing):

The Regeneten Bioinductive Implant is identical to the predicate device (Rotation Medical Collagen Tendon Sheet-DDI) and performance specifications are identical. Past safety, verification, and mechanical characterization tests of the predicate device are directly applicable to the subject device, including staple pull-out testing, hydrothermal transition temperature, endotoxin testing, and mechanical integrity.

### Performance Data (Animal Testing):

Results of the original animal study submitted as part of 510(k) K112423 and K140300 is applicable to the subject device since the subject device and the predicate device have identical indication for use, design, material composition, packaging, and sterilization. Animal study histological results presented in this submission support the performance claims of bioinduction and formation of tendon-like tissue of the predicate device.

## Performance Data (Clinical Study):

No clinical study was conducted for the predicate device and a clinical study was not required to substantiate new performance claims for the subject device. Published clinical literature from investigator-initiated post-market clinical studies was reviewed in this submission in support of new performance claims for the subject device.

#### Compliance to Standards:

This submission does not include any performance or preclinical testing and therefore, is not claiming compliance to any new FDA recognized consensus standards.

#### Substantial Equivalence and Conclusion:

No design modifications were necessary to support the additional performance claims and the subject device (Regeneten Bioinductive Implant) is identical to the predicate device (Rotation Medical Collagen Tendon Sheet-DDI, K140300) in terms of material composition, design, packaging, sterilization, and biocompatibility. The only difference is the addition of performance claims of bioinduction, formation of tendon-like tissue, and change the course of tear progression to the subject device. This difference is substantiated by reviewing the results of the animal study in section 19 and presenting clinical study results from the postmarket clinical studies in section 20. Based on the comparative analysis, Regeneten Bioinductive Implant is substantially equivalent to its predicate, Collagen Tendon Sheet-DDI; they are the same device, only with the addition of performance claims of bioinduction,

{6}------------------------------------------------

formation of tendon-like tissue, and change the course of tear progression have has been added to the subject device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWY/K222501](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWY/K222501)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
