The TOMCAT™ Cervical Spinal System is indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients. The interbody is used with bone screws provided and requires no additional supplementary fixation. The interbody is inserted between the vertebral bodies into the disc space at one or two contiguous levels from the C2/C3 disc space to the C7/T1 disc space for the treatment of cervical degenerative disc disease. Cervical degenerative disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history or radiographic studies. The device system is designed for use with autograft bone and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, to facilitate fusion. The device is implanted by an anterior approach. The TOMCAT™ implant must be used with the screws included in the TOMCAT™ system. This device is to be used in patients who have had six weeks of non-operative treatment.
Device Story
TOMCAT™ Cervical Spinal System is an anterior cervical interbody fusion device; provides stabilization of cervical vertebral column (C2-T1) as adjunct to fusion. System includes interbody spacer (zero or hybrid profile) and titanium bone screws; requires no additional supplementary fixation. Implanted via anterior approach; screws placed into vertebral body/end plate. Used with autograft or allogenic bone graft to facilitate fusion. Intended for use in clinical settings by surgeons. Benefits include alternative to traditional cervical plate and interbody spacer ACDF procedures.
Clinical Evidence
Bench testing only. Sterilization validation performed per ISO 11137; pyrogenicity testing performed per USP <85> and ANSI/AAMI ST72.
Technological Characteristics
Interbody spacer made of PEEK (ASTM F2026), titanium alloy (ASTM F136), tantalum markers (ASTM F560), and nitinol clips (ASTM F2063). Includes titanium bone screws. Zero and hybrid profile designs. Sterilization via validated methods (ISO 11137).
Indications for Use
Indicated for skeletally mature patients with cervical degenerative disc disease (neck pain of discogenic origin confirmed by history/radiography) requiring stand-alone anterior cervical interbody fusion at one or two contiguous levels (C2/C3 to C7/T1) after six weeks of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K152515 — TOMCAT Cervical Spinal System · Choicespine, LP · Dec 18, 2015
K152282 — Cam Lock Plating System · Life Spine, Inc. · Nov 20, 2015
K142060 — TRANSOM ANTERIOR CERVICAL PLATE · Neurostructures, Inc. · Nov 18, 2014
K181562 — Curiteva Anterior Cervical Plate System · Curiteva, LLC · Aug 2, 2018
K250764 — SpineLinc Anterior Cervical Implant System · Lincotek Medical · Dec 5, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 21, 2017
Choice Spine, LP Kim Finch Manager of Regulatory Affairs 400 Erin Drive Knoxville, Tennessee 37919
Re: K170953
Trade/Device Name: TOMCAT™ Cervical Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE Dated: March 29, 2017 Received: March 31, 2017
Dear Ms. Finch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Vincent J. Devlin -S
- for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
> K170953 Page 1 of 1
Device Name
TOMCATTM Cervical Spinal System
Indications for Use (Describe)
The TOMCAT™ Cervical Spinal System is indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients. The interbody is used with bone screws provided and requires no additional supplementary fixation. The interbody is inserted between the vertebral bodies into the disc space at one or two contiguous levels from the C2/C3 disc space to the C7/T1 disc space for the treatment of cervical degenerative disc disease. Cervical degenerative disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history or radiographic studies. The device system is designed for use with autograft bone and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, to facilitate fusion. The device is implanted by an anterior approach. The TOMCAT™ implant must be used with the screws included in the TOMCAT™ system. This device is to be used in patients who have had six weeks of non-operative treatment.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| [X] Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 7. 510(k) Summary
#### I. Submitter
Choice Spine, LP.
400 Erin Drive
Knoxville, TN 37919
Phone: 865-246-3333
Fax: 865-246-3334
Contact Person: Kim Finch, Manager of Regulatory Affairs
Contact Email: kfinch@choicespine.com
Date Prepared: March 30, 2017
# II. Device
Name of Device: TOMCAT™ Cervical Spinal System
Classification Name: Intervertebral Body Fusion Device
Product Class: Class II
Regulatory Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Product Code: OVE
Panel Code: 87- Orthopedics Devices
#### III. Predicate Device
The TOMCAT™ Cervical Spinal System K152515 is the primary predicate device provided for this submission.
# IV. Device Description
The TOMCAT™ Cervical Spinal System is an anterior cervical spinal fixation system for an effective means of stabilizing the cervical vertebral column (C2-T1) as an adjunct to fusion of vertebral bodies. The TOMCAT™ System will provide an alternative to the more common cervical plate and cervical interbody spacer Anterior Cervical Discectomy & Fusion (ACDF) surgical procedure. The TOMCAT™ Cervical Spinal System is a radiolucent and radiopaque intervertebral body fusion device. The interbody is made from PEEK per ASTM F2026 with titanium alloy (Ti-6Al-4V ELI) per ASTM F136, tantalum radiopaque markers per ASTM F560, and nitinol clips per ASTM F2063. This device accepts titanium (Ti-6Al-4V ELI) bone screws that are available in two diameters and multiple lengths.
The system will be composed of a cervical interbody spacer with a zero profile and a hybrid profile design. The hybrid device is implanted anteriorly by inserting two screws,
{4}------------------------------------------------
one screw into the anterior face of vertebral body and the other diagonally though the end plate. The zero profile device implants are implanted anteriorly and stabilized by two diagonally placed screws.
## V. Indications for Use
The TOMCAT™ Cervical Spinal System is indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients. The interbody is used with bone screws provided and requires no additional supplementary fixation. The interbody is inserted between the vertebral bodies into the disc space at one level from the C2/C3 disc space to the C7/T1 disc space for the treatment of cervical degenerative disc disease. Cervical degenerative disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history or radiographic studies. The device system is designed for use with autograft bone and/or allogenic bone graft composed of cancellous and for corticocancellous bone graft, to facilitate fusion. The device is implanted by an anterior approach. The TOMCAT™ implant must be used with the screws included in the TOMCAT™ system. This device is to be used in patients who have had six weeks of non-operative treatment.
# VI. Comparison of Technological Characteristics with the Predicate Device
The device design remains unchanged from the original 510(k) submission (K152515).
### VII. Performance Data
Sterilization validation performed in accordance with ISO 11137 Pyrogenicity testing performed in accordance with USP <85> Bacterial Endotoxin Test (BET) and ANSI/AAMI ST72.
#### VII. Conclusions
The purpose of this submission is to gain clearance for an additional method of sterilization for the TOMCAT™ Interbodies of the TOMCAT™ Cervical Spinal System. The indications for use and design characteristics of the system devices have not changed. Therefore, it can be concluded that the subject device is substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.