← Product Code [OVE](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE) · K162760

# Spine Wave Anterior Cervical Spine System (K162760)

_Spine Wave, Inc. · OVE · Feb 16, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE/K162760

## Device Facts

- **Applicant:** Spine Wave, Inc.
- **Product Code:** [OVE](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE.md)
- **Decision Date:** Feb 16, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Spine Wave Anterior Cervical Spine System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have at least six (6) weeks of non-operative treatment. The Spine Wave Anterior Cervical Spine System is to be used with autogenous bone graft and placed via an open, anterior approach; supplemental fixation (i.e., posterior cervical screw fixation) is required to properly utilize this system.

## Device Story

System consists of non-sterile, single-use cervical interbody cages; components include cage body, anchor plate, and graft cap. Surgeon uses dedicated inserter to place cage into disc space via open anterior approach; deploys anchor plate into adjacent vertebral bodies; packs cage with autogenous bone graft; snaps graft cap into place. Device provides intervertebral body fusion; requires supplemental posterior cervical screw fixation. Used in clinical settings by surgeons to treat DDD; benefits patients by stabilizing cervical spine segment to facilitate fusion and alleviate radicular symptoms.

## Clinical Evidence

Bench testing only. Mechanical performance demonstrated via static and dynamic axial compression, compression-shear, torsion, subsidence, and expulsion testing per ASTM standards (F2077, F2267) and simulated use testing.

## Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V, ASTM F136), commercially pure titanium (ASTM F67), and PEEK-OPTIMA with 6% BaSO4. Multi-component cage design (body, anchor plate, graft cap). Mechanical testing per ASTM F2077, F2267, and draft standard F-04.25.02.02. Non-sterile, single-use.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- ROI-C Cervical Cage System ([K150765](/device/K150765.md))
- Cervical Interbody Spacer System ([K091531](/device/K091531.md))
- Gen II Expandable Interbody System ([K152620](/device/K152620.md))
- Leva® Interbody Device ([K153222](/device/K153222.md))
- CapSure® PS Spine System ([K081228](/device/K081228.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 16, 2017

Spine Wave, Inc. Amnon Talmor Senior Regulatory Affairs Manager 3 Enterprise Drive, Suite 210 Shelton, Connecticut 06484

Re: K162760

Trade/Device Name: Spine Wave Anterior Cervical Spine System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: January 9, 2017 Received: January 10, 2017

Dear Mr. Talmor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/7 description: The image shows the name "Mark N. Melkerson -S" in a bold, sans-serif font. The text is black against a white background. The letters are evenly spaced and clearly legible.

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K162760

Device Name

Spine Wave Anterior Cervical Spine System

Indications for Use (Describe)

The Spine Wave Anterior Cervical Spine System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have at least six (6) weeks of non-operative treatment. The Spine Wave Anterior Cervical Spine System is to be used with autogenous bone graft and placed via an open, anterior approach; supplemental fixation (i.e., posterior cervical screw fixation) is required to properly utilize this system.

X | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the words "Spine Wave" in a bold, sans-serif font. The words are arranged horizontally, with "Spine" on the left and "Wave" on the right. A blue wave design is located under the words, adding a visual element to the logo. The overall design is simple and modern.

# 510(k) Summary Spine Wave Anterior Cervical Spine System

## 1. Submitter Information

| Submitter:     | Spine Wave, Inc.       |
|----------------|------------------------|
| Address:       | Three Enterprise Drive |
|                | Suite 210              |
|                | Shelton, CT 06484      |
| Contact:       | Amnon Talmor           |
| Date Prepared: | February 15, 2017      |

## 2. Device Information

| Trade Name:          | Spine Wave Anterior Cervical Spine System                          |
|----------------------|--------------------------------------------------------------------|
| Common Name:         | Intervertebral Fusion Device with Integrated Fixation,<br>Cervical |
| Classification:      | Class II (special controls) per 21 CFR 888.3080                    |
| Classification Name: | Intervertebral Body Fusion Device                                  |
| Product Code:        | OVE                                                                |

#### 3. Predicate Device Information

The Spine Wave Anterior Cervical Spine System described in this submission is substantially equivalent to the following:

| Primary Predicate Device   | Manufacturer  | 510(k) No. |
|----------------------------|---------------|------------|
| ROI-C Cervical Cage System | LDR Spine USA | K150765    |

| Additional Predicate Devices       | Manufacturer | 510(k) No. |
|------------------------------------|--------------|------------|
| Cervical Interbody Spacer System   | Choice Spine | K091531    |
| Gen II Expandable Interbody System | Spine Wave   | K152620    |
| Leva® Interbody Device             | Spine Wave   | K153222    |
| CapSure® PS Spine System           | Spine Wave   | K081228    |

# 4. Device Description

The Spine Wave Anterior Cervical Spine (ACS) System consists of a selection of non-sterile, single use cervical interbody cages indicated for skeletally mature

{4}------------------------------------------------

patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from C2-T1.

The Spine Wave ACS cages are multi-component devices comprised of a cage body, anchor plate and graft cap. Spine Wave supplies the cage body and anchor plate as a paired combination, and the surgeon assembles the graft cap in situ. The Spine Wave ACS cages are offered in a variety of sizes and shapes to accommodate different patient anatomies and manufactured from titanium alloy (Ti-6Al-4V per ASTM F136), commercially pure titanium (ASTM F67) and PEEK-OPTIMA with 6% BaSO4.

Using the Spine Wave ACS Inserter, the surgeon places the cage body with undeployed anchor plate into the disc space. After placement, the surgeon uses the cage inserter to deploy the superior and inferior portions of the anchor plate into the adjacent vertebral bodies and pack the open end of the cage with autograft. Once grafting is complete, the surgeon snaps the graft cap into the open end of the cage.

#### 5. Indications for Use

The Spine Wave Anterior Cervical Spine System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disk level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have at least six (6) weeks of non-operative treatment. The Spine Wave Anterior Cervical Spine System is to be used with autogenous bone graft and placed via an open, anterior approach; supplemental fixation (i.e., posterior cervical screw fixation) is required to properly utilize this system.

#### 6. Comparison of Technological Characteristics

The substantial equivalence of the Spine Wave Anterior Cervical Fusion Cage System to the listed predicate is shown by similarity in intended use, indications for use, materials and performance.

#### 7. Non-Clinical Performance Data

The following mechanical tests were performed to demonstrate the substantial equivalence of the system to its predicates:

- Static and dynamic axial compression (per ASTM F2077) ●
- Static and dynamic compression-shear (per ASTM F2077)
- Static and dynamic torsion ●
- Subsidence (per ASTM F2267) ●
- Static expulsion (per ASTM Draft Standard F-04.25.02.02) ●
- o Simulated use testing

{5}------------------------------------------------

## 8. Conclusion

Based on the indications for use, technical characteristics, performance testing, and comparison to the predicates, the Spine Wave Anterior Cervical Spine System is substantially equivalent to the predicate devices identified in this submission and does not present any new issues of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE/K162760](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE/K162760)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
