← Product Code [OVE](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE) · K141599

# DIVERGENCE ANTERIOR CERVICAL FUSION SYSTEM (K141599)

_Medtronic Sofamor Danek USA, Inc. · OVE · Jan 21, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE/K141599

## Device Facts

- **Applicant:** Medtronic Sofamor Danek USA, Inc.
- **Product Code:** [OVE](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE.md)
- **Decision Date:** Jan 21, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The DIVERGENCE™ Anterior Cervical Fusion System consists of a stand-alone interbody device indicated for use in anterior cervical interbody fusion procedures in sketally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The DIVERGENCE™ stand-alone cervical interbody device must be used with internal screw fixation. The DIVERGENCE™ stand-alone cervical interbody device is also required to be used with autogenous bone graft and is to be implanted via an open, anterior approach. This cervical device is to be used in patients who have had six weeks of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.

## Device Story

DIVERGENCE™ Anterior Cervical Fusion System is a stand-alone interbody fusion device for single-level cervical spine procedures. System comprises PEEK interbody cage and titanium alloy bone screws. Cage features built-in rotary locking mechanism and radiopaque titanium markers. Screws are self-drilling/self-tapping. Implanted via open, anterior approach by surgeons. Device contains autogenous bone graft to facilitate fusion between vertebral bodies. Provides internal fixation to stabilize cervical segment. Benefits patients by promoting spinal fusion in cases of symptomatic nerve root or spinal cord compression.

## Clinical Evidence

Bench testing only. Testing included static and dynamic compression, compression shear, and torsion per ASTM F2077-11, and subsidence testing per ASTM F2267-04(2011). All results met pre-determined acceptance criteria.

## Technological Characteristics

Interbody cage: PEEK OPTIMA™ LT-1 (ASTM F2026). Screws: Ti-6Al-4V ELI (ASTM F136). Features: 0/6 degree lordosis, 5-12mm heights, 15-20mm widths, 12-16mm depths. Screws: 3.5/4.0mm diameters, 9-17mm lengths. Includes titanium rotary locking mechanism. Sterile, single-use, standalone interbody fusion device.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- PEEK PREVAIL® Cervical Interbody Device ([K113252](/device/K113252.md))
- CORNERSTONE® PSR Cervical Fusion System ([K111264](/device/K111264.md); [K100214](/device/K100214.md))
- AFFINITY® Anterior Cervical Cage System ([P000028](/device/P000028.md))
- ZEPHIR® Anterior Cervical System ([K030327](/device/K030327.md))
- ATLANTIS® Anterior Cervical Plate System ([K130640](/device/K130640.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three faces overlapping to represent the department's focus on people. The logo is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 21, 2015

Medtronic Sofamor Danek USA, Incorporated Ms. Shweta Sharma Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132

Re: K141599

Trade/Device Name: DIVERGENCE™ Anterior Cervical Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: December 23, 2014 Received: December 24, 2014

Dear Ms. Sharma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Shweta Sharma

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Vincent J. Devlin -S

 for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K141599

### Device Name

DIVERGENCE™ Anterior Cervical Fusion System (For Stand-Alone Interbody Device Only)

### Indications for Use (Describe)

The DIVERGENCE™ Anterior Cervical Fusion System consists of a stand-alone interbody device indicated for use in anterior cervical interbody fusion procedures in sketally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The DIVERGENCE™ stand-alone cervical interbody device must be used with internal screw fixation. The DIVERGENCE™ stand-alone cervical interbody device is also required to be used with autogenous bone graft and is to be implanted via an open, anterior approach. This cervical device is to be used in patients who have had six weeks of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.

Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# DIVERGENCE™ Anterior Cervical Fusion System 510(k) Summary

# January 2015

| I.   | Company                | Medtronic Sofamor Danek USA, Inc.<br>1800 Pyramid Place<br>Memphis, TN 38132<br>(901) 396-3133                                                                                                                                                                                                      |
|------|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| II.  | Contact                | Shweta Sharma<br>Senior Regulatory Affairs Specialist                                                                                                                                                                                                                                               |
| III. | Proprietary Trade Name | DIVERGENCETM Anterior Cervical<br>Fusion System                                                                                                                                                                                                                                                     |
| IV.  | Common Name            | Cervical Intervertebral Fusion Device<br>with Bone Graft                                                                                                                                                                                                                                            |
| V.   | Classification Name    | 21 CFR 888.3080 - Intervertebral Body<br>Fusion Device                                                                                                                                                                                                                                              |
|      | Classification         | Class II                                                                                                                                                                                                                                                                                            |
|      | Product Codes          | OVE                                                                                                                                                                                                                                                                                                 |
|      | Primary Predicate      | PEEK PREVAIL® Cervical Interbody<br>Device (K113252, SE 01/17/2012)                                                                                                                                                                                                                                 |
|      | Secondary Predicates   | •CORNERSTONE® PSR Cervical<br>Fusion System (K111264; K100214)<br>•AFFINITY® Anterior Cervical Cage<br>System (P000028, down-classified to<br>Class II special controls, Date of Final<br>Order 06/12/2007)<br>•ZEPHIR® Anterior Cervical System<br>(K030327)<br>•ATLANTIS® Anterior Cervical Plate |

{4}------------------------------------------------

## System (K130640)

#### Product Description VI.

The DIVERGENCE™ Anterior Cervical Fusion System consists of a stand-alone interbody fusion device with internal screw fixation. The DIVERGENCE™ Anterior Cervical Fusion System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. This system is indicated for single-level use only in the C2-T1 anterior spine. The system is comprised of a PEEK interbody cage and screws.

The DIVERGENCE™ anterior cervical cages are provided in 0 and 6 degrees of lordosis, 5-12mm heights, 15-20mm widths and 12-16mm depths. This device is intended to be radiolucent, and the interior space of the product is to be used with autogenous bone graft. The DIVERGENCE™ stand-alone cervical interbody device is manufactured from medical grade polyetheretherketone (PEEK) and contains radiopaque markers made from medical grade titanium alloy. The PEEK interbody cage also comes preassembled with a titanium alloy, built-in rotary locking mechanism.

The bone screws used with this device are provided in self-drilling and self-tapping options and are manufactured from medical grade titanium alloy. The bone screws are provided in 3.5mm and 4.0mm diameters and 9-17mm lengths.

The PEEK material (PEEK OPTIMA™ LT-1) used conforms to ASTM F2026 and the titanium alloy material (Ti-6Al-4V ELI) used conforms to ASTM F136 standards. These devices are offered in sterile form.

### Indications for Use VII.

The DIVERGENCE™ Anterior Cervical Fusion System consists of a stand-alone interbody device indicated for use in anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc

{5}------------------------------------------------

to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The DIVERGENCE™ stand-alone cervical interbody device must be used with internal screw fixation. The DIVERGENCE™ stand-alone cervical interbody device is also required to be used with autogenous bone graft and is to be implanted via an open, anterior approach. This cervical device is to be used in patients who have had six weeks of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.

#### Summary of Technological Characteristics VIII.

The subject DIVERGENCE™ Anterior Cervical Fusion System interbody cage component has the same fundamental scientific technology as the predicate PEEK PREVAIL® Cervical Interbody Device. CORNERSTONE® PSR Cervical Fusion System and AFFINITY® Anterior Cervical Cage System. The subject interbody cage is manufactured from the same PEEK and titanium alloy material as the predicate devices. The predicate and subject devices are also both interbody devices designed to contain graft material and facilitate a fusion between two vertebral bodies.

The subject DIVERGENCE™ Anterior Cervical Fusion System bone screws have the same fundamental scientific technology as the predicate ZEPHIR® Anterior Cervical System and ATLANTIS® Anterior Cervical Plate System screws. The subject devices are manufactured from the same titanium alloy material as the predicates. In both predicate and subject devices, fixation is provided by bone screws inserted into the vertebral body of the cervical spine.

### Identification of Legally Marketed Predicate Devices Used to Claim Substantial IX. Equivalence

The subject interbody cage is substantially equivalent to the predicates:

{6}------------------------------------------------

- PEEK PREVAIL® Cervical Interbody Device (K113252, SE 01/17/2012; . K094042, SE 06/30/2010)
- CORNERSTONE® PSR Cervical Fusion System (K111264, SE 10/12/2011; . K100214, SE 06/25/2010)
- AFFINITY® Anterior Cervical Cage System (P000028, Approval Date . 06/13/2002, down-classified to Class II special controls, Date of Final Order 06/12/2007)

The subject bone screw is substantially equivalent to the predicates:

- ZEPHIR® Anterior Cervical System (K030327, SE 02/26/2003) .
- ATLANTIS® Anterior Cervical Plate System (K130640, SE 06/04/2013) .

#### X. Brief Discussion of the Non-Clinical Tests Submitted

The subject DIVERGENCE™ Anterior Cervical Fusion System devices were tested in accordance with ASTM F2077-11 "Test Methods For Intervertebral Body Fusion Devices" for static and dynamic compression, static and dynamic compression shear, and static and dynamic torsion testing, and ASTM F2267-04(2011) "Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression" for subsidence testing.

The subject devices met the pre-determined acceptance criteria for all the tests. The test results are provided to demonstrate that the subject devices are substantially equivalent to the predicate devices.

#### XI. Conclusions Drawn from the Non-Clinical Tests

Based on the non-clinical test results and additional supporting documentation provided in this pre-market notification, the subject devices demonstrated substantial equivalence to the listed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE/K141599](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVE/K141599)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
