CONDUIT™ SYNFIX™ Evolution Secured Spacer System

K250072 · Avalign Technologies, Inc. · OVD · Jul 10, 2025 · Orthopedic

Device Facts

Record IDK250072
Device NameCONDUIT™ SYNFIX™ Evolution Secured Spacer System
ApplicantAvalign Technologies, Inc.
Product CodeOVD · Orthopedic
Decision DateJul 10, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The CONDUIT™ SYNFIX™ Evolution Secured Spacer System is a stand-alone anterior interbody fusion device with a microscope roughened surface and micro and nano-scale features indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). The interior of the spacer component of the CONDUIT SYNFIX Evolution can be packed with autograft. If used with less than the four integrated bone screws, or for hyperlordotic implants (>20Deg), implants must be used with supplemental fixation systems cleared by the FDA for use in the lumbosacral spine. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Device Story

The CONDUIT™ SYNFIX™ Evolution Secured Spacer System is an anterior lumbar interbody fusion (ALIF) device. It consists of a 3D-printed titanium cage and an anterior plate, assembled at the point of use. The cage features a porous structure with micro and nano-scale surface roughness to facilitate fusion. During surgery, the device is implanted into the intervertebral space and secured to adjacent vertebral bodies using four integrated bone screws with a tapered-thread locking mechanism. The hollow cage interior is packed with autograft. The device provides structural support and stabilization to the spine. It is intended for use by surgeons in a clinical/OR setting. The system is MR conditional.

Clinical Evidence

No clinical data was necessary to demonstrate substantial equivalence; the submission relies on bench testing including static/dynamic axial compression, compression shear, subsidence, and expulsion testing per ASTM standards.

Technological Characteristics

Materials: Ti-6Al-4V ELI (ASTM F3001) for cages via selective laser melting; Ti-6Al-4V (ASTM F136) for plates; Ti-6Al-7Nb (ISO 5832) for screws. Features: Porous/solid structures, micro/nano-scale surface roughness. Connectivity: None. Sterilization: Not specified. Software: None.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade I spondylolisthesis. DDD defined as discogenic back pain with radiographic confirmation and six months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 10, 2025 Avalign Technologies, Inc. Jennifer Staunton Director Regulatory Affairs 8727 Clinton Park Drive Fort Wayne, Indiana 46825 Re: K250072 Trade/Device Name: CONDUIT™ SYNFIX™ Evolution Secured Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, OVD Dated: June 11, 2025 Received: June 11, 2025 Dear Jennifer Staunton: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250072 - Jennifer Staunton Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K250072 - Jennifer Staunton Page 3 Sincerely, Brent Showalter -S Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250072 Device Name CONDUIT™ SYNFIX™ Evolution Secured Spacer System Indications for Use (Describe) The CONDUIT™ SYNFIX™ Evolution Secured Spacer System is a stand-alone anterior interbody fusion device with a microscope roughened surface and micro and nano-scale features indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). The interior of the spacer component of the CONDUIT SYNFIX Evolution can be packed with autograft. If used with less than the four integrated bone screws, or for hyperlordotic implants (>20Deg), implants must be used with supplemental fixation systems cleared by the FDA for use in the lumbosacral spine. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} K250072 Page 1 of 3 510(k) Summary Submitter: Avalign Technologies, Inc. 8727 Clinton Park Drive Fort Wayne, IN 46825 USA Contact Person: Jennifer Staunton Director Regulatory Affairs Telephone: 219-718-1152 Date Prepared: January 9, 2025 Trade Name: CONDUIT™ SYNFIX™ Evolution Secured Spacer System Device Class: Class II Product Code: OVD, MAX Common Name: Intervertebral Fusion Device With Integrated Fixation, Lumbar Classification Name: Intervertebral Body Fusion Device Regulation Number: 21 CFR 888.3080 Classification Panel: Orthopaedic and Rehabilitation Devices Panel (87) Primary Predicate: Synthes SYNFIX™ Evolution Secured Spacer System (K150673) Additional Predicates: DePuy Synthes EIT Cellular Titanium® ALIF Cage (K201605) EIT Emerging Implant Technologies, GmbH Cellular Titanium® ALIF Cage (K170503) Stryker Monterey™ AL Interbody System (K201585) Device Description: The CONDUIT SYNFIX Evolution Secured Spacers are intervertebral body fusion devices intended for lumbar interbody fusion (ALIF). Four Screws are inserted through the anteriorly-located Plate into the adjacent vertebral bodies. The Screws lock securely to the Plate using a tapered-thread locking mechanism. The CONDUIT SYNFIX Evolution Secured Spacer System is available as non-assembled Cage and Plate components in various heights and geometries to suit individual pathology and anatomical conditions. The Cage and Plate components are intended to be assembled at the point of use prior to implantation. The CONDUIT SYNFIX Evolution Cages are made from Ti-6Al-4V {5} K250072 Page 2 of 3 ELI conforming to ASTM F3001 with an additive manufacturing process (Selective Laser Melting). The design contains solid structures and porous structures. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The 3D Printed Conduit Cellular Titanium Cages have a microscopic roughened surface with micro and nano-scale features. The micro and nano features are on all surfaces of the Cage, including the superior, inferior, and peripheral surfaces, as well as each member of the internal cell structure. **Indications for Use:** The CONDUIT™ SYNFIX™ Evolution Secured Spacer System is a stand-alone anterior interbody fusion device with a microscope roughened surface and micro and nano-scale features indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). The interior of the spacer component of the CONDUIT SYNFIX Evolution can be packed with autograft. If used with less than the four integrated bone screws, or for hyperlordotic implants (>20Deg), implants must be used with supplemental fixation systems cleared by the FDA for use in the lumbosacral spine. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. **Materials:** The components are manufactured from medical grade Titanium. The Cage from standard specification for Ti-6Al-4V ELI using full-melt powder bed fusion in conformance with ASTM F3001. The Plate is manufactured from Ti-6Al-4V per ASTM F136 and bone Screws are supplied by DePuy Synthes and manufactured from Ti-6Al-7Nb (ISO 5832). **Comparison to Predicate Device:** The substantial equivalence of the subject device to the predicates identified above is based upon the equivalence of intended use, design (fundamental scientific technology), performance, sterility, and biocompatibility. **Performance Data:** Mechanical testing, including static and dynamic axial compression per ASTM F2077-24, static and dynamic compression shear per ASTM F2077-24, subsidence per ASTM F2267-24, and expulsion was performed to provide data to support a substantial equivalence {6} K250072 Page 3 of 3 determination. The mechanical testing was performed to characterize the properties and functionality of the system, as well as to allow comparison with established acceptance criteria. Additionally, the subject device was evaluated for magnetically induced displacement force per ASTM F2052-21, magnetically induced torque per ASTM F2213-17, MR image artifact per ASTM F2119-24 and RF-induced heating per ASTM F2182-19e2 to support the MR Conditional labeling. **Clinical Test Summary:** No clinical data was necessary to demonstrate substantial equivalence, nor safety and effectiveness of this system. **Conclusion:** Based on the predicate comparison of intended use, indications, technological characteristics, and device performance, the CONDUIT SYNFIX Evolution Secured Spacer System has demonstrated substantial equivalence to the identified predicate device systems.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...