NuVasive Modulus ALIF System

K230894 · Nu Vasive, Incorporated · OVD · Jun 16, 2023 · Orthopedic

Device Facts

Record IDK230894
Device NameNuVasive Modulus ALIF System
ApplicantNu Vasive, Incorporated
Product CodeOVD · Orthopedic
Decision DateJun 16, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The NuVasive Modulus ALIF System is indicated for spinal fusion procedures in skeletally mature patients. These devices are intended to be used with bone screws, anchoring blades, or a combination of the two. When used with Modulus ALIF bone screws, the Modulus ALIF System 10° 20° lordotic cages may be used as a standalone system. When used with Modulus ALIF anchoring blades, the Modulus ALIF Interfixated System must be used with supplemental internal spinal fixation systems (e.q., posterior pedicle screw and rod system) that are cleared for use in the lumbar spine. When used with either the Modulus ALIF bone screws or the Modulus ALIF anchoring blades, the Modulus ALIF Interfixated system 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems that are cleared for use in the lumbar spine. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive Modulus ALIF System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and imaging studies (e.g., radiographs, CT scan, MRI scan). The Modulus ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolistheses and spinal stenosis at one or two adjacent levels, the Modulus ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by FDA for use in the lumbar spine in addition to the integrated screws/anchoring blades.

Device Story

Intervertebral body fusion device; titanium alloy (Ti-6Al-4V ELI) construction; powder bed fusion manufacturing. Features hollow core for bone graft packing; internal lattice structure; microporous textured surfaces for vertebral gripping. Used in lumbar spine (L2-S1) to facilitate fusion; implanted by surgeons during spinal fusion procedures. Employs integrated fixation via bone screws or new anchoring blades. 10°-20° lordotic cages may be used standalone with screws; 25°-30° cages and all blade-fixated configurations require supplemental posterior pedicle screw/rod fixation. Provides structural support and stability to intervertebral space; promotes fusion through graft containment; reduces risk of device migration via textured surfaces.

Clinical Evidence

Bench testing only. Performed mechanical performance testing including static/dynamic axial compression and compression-shear (ASTM F2077), fixation push-out, gravimetric/particulate analysis, adjacent-level impingement trajectory engineering analysis, subsidence, Taber abrasion, and stereological/metallurgical analysis. No clinical data provided.

Technological Characteristics

Material: Titanium alloy (Ti-6Al-4V ELI) per ASTM F3001 Class C. Manufacturing: Powder bed fusion. Design: Hollow core, internal lattice, microporous textured surfaces. Fixation: Integrated bone screws or anchoring blades. Connectivity: None. Sterilization: Provided sterile and non-sterile.

Indications for Use

Indicated for skeletally mature patients requiring spinal fusion in the lumbar spine (L2-S1) following discectomy for symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, or spinal stenosis at 1-2 levels. Also indicated as an adjunct for multilevel degenerative scoliosis and sagittal deformity. Patients must have failed at least 6 months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 16, 2023 NuVasive, Incorporated Alexander R. Stevens Regulatory Affairs Specialist 7475 Lusk Blvd San Diego, California 92121 Re: K230894 Trade/Device Name: NuVasive Modulus ALIF System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD, MAX Dated: March 30, 2023 Received: March 31, 2023 Dear Alexander R. Stevens: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Brent Showalter -S Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Submission Number (if known) K230894 Device Name NuVasive Modulus ALIF System ### Indications for Use (Describe) The NuVasive Modulus ALIF System is indicated for spinal fusion procedures in skeletally mature patients. These devices are intended to be used with bone screws, anchoring blades, or a combination of the two. When used with Modulus ALIF bone screws, the Modulus ALIF System 10° 20° lordotic cages may be used as a standalone system. When used with Modulus ALIF anchoring blades, the Modulus ALIF Interfixated System must be used with supplemental internal spinal fixation systems (e.q., posterior pedicle screw and rod system) that are cleared for use in the lumbar spine. When used with either the Modulus ALIF bone screws or the Modulus ALIF anchoring blades, the Modulus ALIF Interfixated system 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems that are cleared for use in the lumbar spine. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive Modulus ALIF System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and imaging studies (e.g., radiographs, CT scan, MRI scan). The Modulus ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolistheses and spinal stenosis at one or two adjacent levels, the Modulus ALIF System must be used with a supplemental internal spinal fixation system (e.q., pedicle screw system) cleared by FDA for use in the lumbar spine in addition to the integrated screws/anchoring blades. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape on the left, with the word "NUVASIVE" in gray letters to the right of the leaf. The leaf shape is split into two halves, with the top half being a darker purple color and the bottom half being a lighter gray color. ### 510(k) Summary In accordance with Title 21 CFR §807.92, the following summary of information is provided: #### Submitted by: A. Alexander R. Stevens Specialist, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (347) 620-3110 Date Prepared: June 13, 2023 #### B. Device Name | Trade or Proprietary Name: | NuVasive Modulus ALIF System | |----------------------------|-----------------------------------| | Common or Usual Name: | Intervertebral Body Fusion Device | | Classification Name: | Intervertebral Body Fusion Device | | Device Class: | Class II | |-----------------|-------------------| | Classification: | 21 CFR § 888.3080 | | Product Code: | OVD, MAX | #### C. Predicate Devices The subject device of this premarket submission are substantially equivalent to the following predicate devices: ### Primary - . NuVasive Modulus ALIF System (Previously known as 3DP Interfixated ALIF System) - K193593 ### Additional - NuVasive Modulus ALIF System K210271 ● - NuVasive Thoracolumbar Interbody Systems K203714 ● - Globus Medical Hedron IA Integrated Lumbar Spacers K191391 ● #### Device Description D. The subject NuVasive Modulus ALIF System interbody implants are manufactured from titanium alloy (Ti-6Al-4V ELI) powder conforming to ASTM F3001 Class C. The solid and porous structures are simultaneously built using a powder bed fusion method. The hollow core, or graft aperture, allows for packing of graft to aid in the promotion of a solid fusion. Similarly, the macroporous internal lattice structure provides additional space for graft packing. The microporous, textured surface on the superior and inferior ends of the device serves to grip the adjacent vertebrae to resist migration of the device. The implants are available in a variety sizes and lordotic angles to suit the individual pathology and anatomical conditions of the patient. When used with the Modulus ALIF bone screws, the Modulus {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf-like shape on the left, with the word "NUVASIVE" in gray letters to the right of the shape. The leaf shape is split into two halves, with the top half in purple and the bottom half in gray. ALIF Interfixated System 10°- 20° lordotic cages may be used as a standalone system. When used with the Modulus ALIF anchoring blades, the Modulus ALIF Interfixated System must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared for use in the lumbar spine. When used with either the Modulus ALIF bone screws or the Modulus ALIF anchoring blades, the Modulus ALIF Interfixated system 25°- 30° lordotic cages must be used with supplemental internal spinal fixation systems that are cleared for use in the lumbar spine. The purpose of this submission is to introduce the Modulus ALIF anchoring blades (blades), which are a new in-line fixation option, provided sterile and non-sterile, for the NuVasive Modulus ALIF System. ### E. Intended Use ### NuVasive Modulus ALIF System The NuVasive Modulus ALIF System is indicated for spinal fusion procedures in skeletally mature patients. These devices are intended to be used with bone screws, anchoring blades, or a combination of the two. When used with Modulus ALIF bone screws, the Modulus ALIF System 10° - 20° lordotic cages may be used as a standalone system. When used with Modulus ALIF anchoring blades, the Modulus ALIF Interfixated System must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared for use in the lumbar spine. When used with either the Modulus ALIF bone screws or the Modulus ALIF anchoring blades, the Modulus ALIF Interfixated system 25°-30° lordotic cages must be used with supplemental internal spinal fixation systems that are cleared for use in the lumbar spine. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive Modulus ALIF System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and imaging studies (e.g., radiographs, CT scan, MRI scan). The Modulus ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolistheses and spinal stenosis at one or two adjacent levels, the Modulus ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by FDA for use in the lumbar spine in addition to the integrated screws/anchoring blades. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape on the left and the word "NUVASIVE" on the right. The leaf shape is divided into two parts, with the top part being purple and the bottom part being gray. The word "NUVASIVE" is written in gray, sans-serif letters. #### F. Technological Characteristics As established in this eSTAR submission, the subject NuVasive Modulus ALIF System is substantially equivalent to the predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to have substantially equivalent technological characteristics to the predicate devices through comparison of design, intended use, material composition, function, and range of sizes. #### Performance Data G. The purpose of this submission is to introduce the Modulus ALIF anchoring blades. Mechanical performance testing was conducted to assess the safety and compatibility of the subject device. The following testing was performed: - F2077 Static & Dynamic Axial Compression ● - F2077 Static & Dynamic Compression-Shear ● - Fixation Push-Out - Gravimetric and Particulate Analysis ● - Adjacent-Level Impingement Trajectory (Engineering Analysis) ● - Subsidence ● - Taber Abrasion ● - Stereological Evaluation and Metallurgical Analysis ● The results of these studies demonstrate that the subject NuVasive Modulus ALIF System is substantially equivalent to the predicate devices. #### H. Conclusions Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject NuVasive Modulus ALIF System has been shown to be substantially equivalent to the legally marketed predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...