SCARLET AL-T
K192993 · Spineart · OVD · Jan 22, 2020 · Orthopedic
Device Facts
| Record ID | K192993 |
| Device Name | SCARLET AL-T |
| Applicant | Spineart |
| Product Code | OVD · Orthopedic |
| Decision Date | Jan 22, 2020 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3080 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Scarlet® AL-T system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These spinal implants are to be used with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage. When used with the integrated fixation by the mean of the bone screws provided, the Scarlet® AL-T is a stand-alone system and requires no additional supplemental fixation system. When used as a lumbar intervertebral fusion device (i.e. without the bone screws provided), the Scarlet® AL-T interbody device must be used with supplemental internal spinal fixation system that has been cleared by the FDA for use in the lumbosacral spine. The Scarlet® AL-T Hyperlordotic (≥ 20°) cages are to be used with BOTH the integrated fixation by the mean of the bone screws provided and ALSO additional supplemental fixation system that has been cleared by the FDA for use in the lumbosacral spine.
Device Story
Anterior lumbar interbody fusion device; provides mechanical support and maintains disc space during fusion. Implanted via anterior approach; box-shaped spacer with central cavity for bone graft. Features monolithic design with solid, lattice, and porous structures; rough superior/inferior surfaces for stability/bony integration. Integrated fixation uses three unicortical cancellous bone screws secured by two cam locks to prevent back-out. Hyperlordotic variants (≥ 20°) require additional supplemental fixation. Used in clinical settings by surgeons; provides structural support to facilitate spinal fusion; benefits patients by stabilizing vertebral segments and promoting bone growth.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing included static axial compression, static shear-compression, static torsion (ASTM F2077-18), subsidence (ASTM F2267-04), and expulsion (ASTM Draft F-04.25.02.02). Results demonstrated comparable mechanical properties to the predicate device.
Technological Characteristics
Medical grade titanium alloy (ASTM F136); additive manufacturing (SLM) per ASTM F3001. Monolithic design with lattice/porous structures. Integrated fixation via three bone screws and two cam locks. Gamma sterilization. Standalone or used with supplemental fixation.
Indications for Use
Indicated for skeletally mature patients with lumbar degenerative disc disease (DDD) at one or two contiguous levels (L2-S1) with discogenic back pain, including up to Grade 1 spondylolisthesis or retrolisthesis. Requires 6 months of prior non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
- Scarlet® Al-T (K181818)
- NuVasive BASE Interfixated Titanium System (K170592)
Related Devices
- K171567 — IRIX-A Lumbar Integrated Fusion System · X-Spine Systems, Inc. · Aug 28, 2017
- K180373 — CancelleX Porous Titanium Lumbar Interbody Device · Xenco Medical, LLC · Aug 24, 2018
- K201087 — CORBEL Spacers · Globus Medical, Inc. · Jul 31, 2020
- K221324 — ENZA-O Titanium Lateral Anterior Lumbar Interbody Fusion (ALIF) · Camber Spine Technologies · Mar 30, 2023
- K211837 — Altus Spine Interbody Standalone Fusion System · Altus Partners, LLC · Apr 19, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 22, 2020
Spineart Franck Pennesi Chief Technical Officer 3 Chemin du Pre Fleuri 1228 Plan Les Ouates, Geneva SWITZERLAND
Re: K192993
Trade/Device Name: Scarlet AL-T Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD, MAX Dated: October 22, 2019 Received: October 25, 2019
Dear Mr. Pennesi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Brent Showalter, PhD Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192993
Device Name Scarlet® AL-T
#### Indications for Use (Describe)
The Scarlet® AL-T system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These spinal implants are to be used with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.
When used with the integrated fixation by the mean of the bone screws provided, the Scarlet® AL-T is a stand-alone system and requires no additional supplemental fixation system.
When used as a lumbar intervertebral fusion device (i.e. without the bone screws provided), the Scarlet® AL-T interbody device must be used with supplemental internal spinal fixation system that has been cleared by the FDA for use in the lumbosacral spine.
The Scarlet® AL-T Hyperlordotic (≥ 20°) cages are to be used with BOTH the integrated fixation by the mean of the bone screws provided and ALSO additional supplemental fixation system that has been cleared by the FDA for use in the lumbosacral spine.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|----------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Spineart. The logo consists of the word "spineart" in a stylized font, with a butterfly above the word. The butterfly and the "art" portion of the word are purple, while the "spine" portion of the word is black.
# K192993 510(k) SUMMARY
| 510k | TRADITIONAL |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Basis for submission | New devices - Product Line Extension |
| Submitted by | SPINEART |
| | 3 Chemin du Pré Fleuri |
| | 1228 PLAN LES OUATES<br>GENEVA SWITZERLAND |
| Contacts | Franck PENNESI Chief Technical Officer |
| | Phone: +41 22 570 1200 Fax: +41 22 594 8306 |
| | Mail: fpennesi@spineart.com<br>Regulatory contact: Dr Isabelle DRUBAIX (Idée Consulting) idrubaix@nordnet.fr |
| Date Prepared | December 19, 2019 |
| Common Name | Intervertebral body fusion device |
| Trade Name | Scarlet® AL-T |
| Classification Name | Intervertebral Fusion Device with Integrated Fixation Lumbar |
| Class | II |
| Product Code | OVD |
| CFR section | 888.3080 |
| Device panel | ORTHOPEDIC |
| Legally marketed<br>predicate devices | Primary predicate: Scarlet® Al-T manufactured by Spineart (K181818) |
| | Additional predicate: NuVasive BASE Interfixated Titanium System manufactured by NuVasive,<br>Incorporated (K170592) |
| Indications for use | The Scarlet® AL-T system is indicated for intervertebral body fusion procedures in skeletally mature<br>patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels<br>from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by<br>patient history and radiographic studies. These DDD patients may also have up to Grade 1<br>spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with<br>autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone<br>graft to facilitate fusion. Patients should have at least six (6) months of non-operative treatment<br>prior to treatment with an intervertebral cage.<br>When used with the integrated fixation by the mean of the bone screws provided, the Scarlet® AL-<br>T is a stand-alone system and requires no additional supplemental fixation system.<br>When used as a lumbar intervertebral fusion device (i.e. without the bone screws provided), the<br>Scarlet® AL-T interbody device must be used with supplemental internal spinal fixation system that<br>has been cleared by the FDA for use in the lumbosacral spine.<br>The Scarlet® AL-T Hyperlordotic (≥ 20°) cages are to be used with BOTH the integrated fixation by<br>the mean of the bone screws provided and ALSO additional supplemental fixation system that has<br>been cleared by the FDA for use in the lumbosacral spine. |
| Description<br>of<br>the<br>device | The Scarlet® AL-T spinal system is an anterior lumbar interbody fusion device with integrated<br>fixation intended to provide mechanical support to the lumbar spine and maintain adequate disc<br>space until fusion occurs. The Scarlet® AL-T system comprises a range of intervertebral spacers<br>implanted via an anterior approach, and having various sizes, heights, footprints and lordosis so as<br>to adapt individual pathology and different patient's anatomical conditions. The interbody device<br>is a box-shaped spacer with a large central cavity that can receive bone graft intended to promote<br>intervertebral fusion. The Scarlet® AL-T spacers are all made from medical grade titanium alloy<br>conforming to ASTM F136 standard and are produced by additive manufacturing (SLM) according<br>to ASTM F3001. Subsequently the spacer is machined (thread tapping) and polished. |
| | The Scarlet® AL-T interbody spacer has a monolithic design that incorporates solid, lattice and |
| | porous structures along with superior and inferior rough surfaces intended to increase implant |
| | stability into the intervertebral space and bony integration throughout the implant. |
| | When used with its integrated fixation, the spacer is crossed by three (3) bone screws protruding |
| | into the vertebral endplates. The bone screws are secured by the mean of two (2) cam locks that |
| | prevent backing out. |
| | The Scarlet® AL-T spinal implants are delivered sterile (gamma sterilization) and supplied with |
| | dedicated surgical instruments (reusable - provided non-sterile). Bacterial endotoxin testing as |
| | specified in USP standard is used for pyrogenicity testing to achieve the Endotoxin limit of 20 EU / |
| | device. |
| | The subject product line extension consists of an extended range of titanium alloy spacers intended |
| | to be used with the three unicortical cancellous bone screws previously approved (K181818). |
| | Scarlet® AL-T product line extension comprises a range of hyperlordotic intervertebral spacers |
| | implanted via an anterior approach, and having various sizes, heights, footprints and lordosis so as |
| | to adapt individual pathology and different patient's anatomical conditions. The Scarlet® AL-T |
| | Hyperlordotic (≥ 20°) cages are to be used with BOTH the integrated fixation by the mean of the |
| Technological<br>characteristics<br>compared to<br>the | bone screws provided and ALSO additional supplemental fixation system that has been cleared by |
| predicate devices | the FDA for use in the lumbosacral spine. |
| | Except for the lordosis, the Scarlet® AL-T hyperlordotic implants present the same design features, |
| | anti-backout mechanism, integrated fixation and bone screws as the cleared Scarlet® AL-T devices |
| | (K181818). The Scarlet® AL-T hyperlordotic implants are manufactured using the same |
| | manufacturing technology, i.e. additive manufacturing (SLM) as predicate device Scarlet® AL-T |
| | (K181818). Except for the dedicated trial devices and implant holder, the Scarlet® AL-T |
| | hyperlordotic implants share the same surgical instrumentation as predicate device Scarlet® AL-T |
| | (K181818). |
| Discussion of Testing | The following non-clinical tests were conducted on the Scarlet® AL-T hyperlordotic cages: Static |
| | Axial Compression according to ASTM F2077-18, Static Shear-compression according to ASTM…