← Product Code [OVD](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD) · K160605

# MectaLIF Anterior Stand-Alone (K160605)

_Medacta International · OVD · Jul 21, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD/K160605

## Device Facts

- **Applicant:** Medacta International
- **Product Code:** [OVD](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD.md)
- **Decision Date:** Jul 21, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The MectaLIF Anterior Stand Alone system is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the interior of the spacer component of the MectaLIF Anterior Stand Alone can be packed with autograft or autologous bone graft. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The MectaLIF Anterior Stand-Alone system intended to be used with bone screws provided and requires no additional supplementary fixation. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).

## Device Story

Anterior interbody fusion device; used for spinal stabilization in DDD patients. System comprises PEEK disc spacer (optionally titanium-coated), titanium alloy plate, and titanium bone screws. Spacer interior packed with autograft/autologous bone graft. Plate secured to spacer via interlocking mechanism; assembly fixed to vertebral body with screws. Provides stand-alone fixation; no supplementary fixation required. Used by surgeons in clinical settings to facilitate fusion; benefits patient by stabilizing spinal segment and reducing discogenic pain.

## Clinical Evidence

Bench testing only. Performance testing conducted per ASTM F2077, ASTM F2267, ASTM F1877, and ISO 17853, including static/dynamic axial compression, static/dynamic compression-shear, static/dynamic torsion, subsidence, and expulsion testing.

## Technological Characteristics

Materials: PEEK (ASTM F2026), CPTi coating (ASTM F1580), Ti6Al4V ELI (ISO 5832-3/ASTM F136). Components: Disc spacer, plate (Flush, Long, L5-S1, Hybrid), bone screws. Tantalum markers (ISO 13782/ASTM F560). Stand-alone fixation via interlocking plate-spacer mechanism. Sterilization method not specified.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- MectaLIF Anterior ([K124034](/device/K124034.md))
- MectaLIF TiPEEK ([K133192](/device/K133192.md))
- Mecta-C TiPEEK ([K142744](/device/K142744.md))
- Vu aPOD ([K101310](/device/K101310.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 21, 2016

Medacta International % Ms. Roshana Ahmed. RAC Mapi USA, Incorporated 2343 Alexandria Drive, Suite 100 Lexington, Kentucky 40504

Re: K160605

Trade/Device Name: MectaLIF Anterior Stand-Alone Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: June 15, 2016 Received: June 15, 2016

Dear Ms. Ahmed:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K160605

Device Name MectaLIF Anterior Stand-Alone

#### Indications for Use (Describe)

The MectaLIF Anterior Stand Alone system is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the interior of the spacer component of the MectaLIF Anterior Stand Alone can be packed with autograft or autologous bone graft.

DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

The MectaLIF Anterior Stand-Alone system intended to be used with bone screws provided and requires no additional supplementary fixation. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Medacta International. The logo consists of the word "Medacta" in blue, with three blue triangles to the left of the word. There is a yellow line underneath the word "Medacta", and the word "International" is written in blue below the yellow line. To the right of the word "International" is a red square with a white cross in the center.

## 510(k) Summary

|                 | Applicant/Sponsor:    Medacta International SA |
|-----------------|------------------------------------------------|
|                 | Strada Regina                                  |
|                 | 6874 Castel San Pietro (CH)                    |
|                 | Switzerland                                    |
|                 | Phone (+41) 91 696 60 60                       |
|                 | Fax (+41) 91 696 60 66                         |
| Contact Person: | Roshana Ahmed, M.A., RAC                       |
|                 | Mapi USA, Inc.                                 |
|                 | 2343 Alexandria Drive, Suite 100               |
|                 | Lexington, KY 40504                            |
|                 | Sr. Manager, Regulatory Affairs                |
|                 | Phone: 657-248-7361                            |

Date Prepared: June 19, 2016

### DEVICE INFORMATION

| Trade/Proprietary Name:  | MectaLIF Anterior Stand-Alone                                 |
|--------------------------|---------------------------------------------------------------|
| Common or Usual<br>Name: | Intervertebral body fusion device                             |
| Classification Name:     | Intervertebral Fusion Device with Integrated Fixation, Lumbar |
| Product Code:            | OVD                                                           |
| Regulation Number:       | 888.3080                                                      |
| Device Class:            | 2                                                             |

### PREDICATE DEVICE INFORMATION

Primary Predicate:

MectaLIF Anterior, K124034, Medacta International SA

Additional Predicates:

MectaLIF TiPEEK, K133192, Medacta International SA Mecta-C TiPEEK, K142744, Medacta International SA Vu aPOD, K101310, Theken Spine

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Medacta International. The logo consists of two blue triangles on the left, followed by the name "Medacta" in blue. Underneath the name is a yellow line, and below that is the word "International" in gray. To the right of the word "International" is a red square with a white cross in the center.

### DEVICE DESCRIPTION

The MectaLIF Anterior Stand-Alone is a stand-alone anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. MectaLIF Anterior Stand-Alone is intended to be used with bone screws provided and requires no additional supplementary fixation.

MectaLIF Anterior Stand-Alone consists of disc spacers made of PEEK Implant Grade Polyetheretherketone (ASTM F 2026) body. The spacer is available without any coatings and with a commercially pure titanium (CPTi, ASTM F 1580) coating. Both coated and uncoated spacers contain tantalum markers (ISO 13782 / ASTM F 560), and include bone screws made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136), and a plate made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136).

The interior of the disc spacer can be packed with autograft or autologous bone graft. The plate comes in four different designs (Flush, Long, L5-S1, and Hybrid) and is secured to the disc spacer via an interlocking mechanism. The disc spacer and attached plate are secured to the vertebral body with the bone screws. The Flush, Long, and Hybrid plates are used with four bone screws while the L5-S1 plate is used with three bone screws.

The purpose of this submission is to alter the indications for use, and introduce new plate and screw designs.

### INDICATIONS FOR USE

The MectaLIF Anterior Stand Alone system is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The interior of the spacer component of the MectaLIF Anterior Stand Alone can be packed with autograft or autologous bone graft.

DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

The MectaLIF Anterior Stand-Alone system is a system intended to be used with bone screws provided and requires no additional supplementary fixation. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Medacta International. The logo consists of three blue triangles arranged in a row, followed by the word "Medacta" in blue. Below the triangles and the word is a yellow line, and below that is the word "International" in gray, followed by a red and white Swiss flag.

## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The MectaLIF Anterior Stand Alone is identical to the identified predicate devices in terms of cage size footprint, plate sizes, lordosis, screw plate interface, screw sizes, materials of construction, and sterilization method.

#### PERFORMANCE TESTING

The following tests were conducted on the MectaLIF Anterior Stand-Alone in accordance with ASTM F2077, ASTM F2267, ASTM F1877, and ISO 17853:

- 1. Static Axial Compression
- 2. Dynamic Axial Compression
- 3. Static Compression-Shear
- Dynamic Compression-Shear 4.
- న. Static Torsion
- 6. Dynamic Torsion
- 7. Subsidence
- 8. Expulsion

#### CONCLUSION:

Based on the information presented above and within the submission, the MectaLIF Anterior Stand Alone is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD/K160605](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD/K160605)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
