← Product Code [OVD](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD) · K151481

# Sahara Stabilization System (K151481)

_K2m, Inc. · OVD · Dec 21, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD/K151481

## Device Facts

- **Applicant:** K2m, Inc.
- **Product Code:** [OVD](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD.md)
- **Decision Date:** Dec 21, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The SAHARA Stabilization System implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients. Lumbar implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non-operative treatment. The hyperlordotic lumbar implants (at angles > 15°) must be used with supplemental fixation (i.e., posterior pedicle screw and rod system) cleared for use in the lumbar spine, in addition to the bone screws provided. Otherwise, the Sahara Stabilization System implants may be used as a stand-alone device, which is intended to be used with the bone screws provided.

## Device Story

Expandable intervertebral body fusion device; provides support and stabilization to lumbar spine segments. Implants inserted into disc space; used with autogenous bone graft to promote fusion. Hyperlordotic variants (>15°) require supplemental posterior pedicle screw/rod fixation; standard variants may function as stand-alone devices using provided bone screws. Used by surgeons in spinal fusion procedures.

## Clinical Evidence

No clinical data. Evidence consists of mechanical bench testing (static compression, static torsion, static compression shear, dynamic compression, expulsion, and subsidence) and a cadaver implantation study.

## Technological Characteristics

Expandable intervertebral body fusion device. Materials: Titanium (ASTM F1472, F67) and Cobalt Chrome (ASTM F1537). Mechanical design for lumbar stabilization.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- K2M Chesapeake ([K092211](/device/K092211.md))
- K2M Chesapeake ([K120031](/device/K120031.md))
- K2M Chesapeake ([K142487](/device/K142487.md))
- K2M Santorini ([K111294](/device/K111294.md))
- Globus Rise ([K113447](/device/K113447.md))
- Wenzel Spine Vari-Lift-L ([K100820](/device/K100820.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 21, 2015

K2M, Incorporated Ms. Nancv Giezen Manager Regulatory Affairs 751 Miller Drive, Southeast Leesburg, Virginia 20175

Re: K151481

Trade/Device Name: SAHARA Stabilization System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: November 16, 2015 Received: November 17, 2015

Dear Ms. Giezen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K151481

Device Name Sahara Stabilization System

#### Indications for Use (Describe)

The SAHARA Stabilization System implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients. Lumbar implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non-operative treatment.

The hyperlordotic lumbar implants (at angles > 15°) must be used with supplemental fixation (i.e., posterior pedicle screw and rod system) cleared for use in the lumbar spine, in addition to the bone screws provided. Otherwise, the Sahara Stabilization System implants may be used as a stand-alone device, which is intended to be used with the bone screws provided.

| Type of Use (Select one or both, as applicable)                                                                                  |                                                                                                                                 |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary Sahara Stabilization System K2M, Inc.

#### Submitter

K2M, Inc. 751 Miller Drive SE Leesburg, VA 20175 Contact Person: Nancy Giezen Telephone: (571) 919-2000 Date Prepared: 12/16/2015

#### Classification

| Trade Name:       | Sahara Stabilization System |
|-------------------|-----------------------------|
| Common Name:      | Spinal Fixation System      |
| Regulatory Class: | Class II                    |

Classification Name(s): Intervertebral Body Fusion Device with Integrated Fusion (21 CFR 888.3080, Product Code OVD)

### Predicate Device(s)

Primary Predicate: K2M Chesapeake (K092211) Additional Predicates: K2M Chesapeake (K120031, 142487) K2M Santorini (K111294) Globus Rise (K113447) Wenzel Spine Vari-Lift-L (K100820)

#### Device Description

The subject of this 510(k) submission is the Sahara Stabilization System. The implants function as an expandable intervertebral body fusion devices to provide support and stabilization to the lumbar segments of the spine. The implants are manufactured from Titanium (ASTM F1472, F67) and Cobalt Chrome (ASTM F1537).

#### Indications for Use

The SAHARA Stabilization System implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients. Lumbar implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non-operative treatment.

{4}------------------------------------------------

The hyperlordotic lumbar implants (at angles > 15°) must be used with supplemental fixation (i.e., posterior pedicle screw and rod system) cleared for use in the lumbar spine, in addition to the bone screws provided. Otherwise, the Sahara Stabilization System implants may be used as a stand-alone device, which is intended to be used with the bone screws provided.

#### Technological Comparison to Predicate(s)

The Sahara Stabilization System was compared to predicate systems and were found to be substantially the same as these systems.

#### Non-clinical Performance Evaluation

Mechanical testing (including static compression, static torsion, static compression shear, dynamic compression, dynamic compression, expulsion, and subsidence) and a cadaver implantation study was performed in support of this submission and the proposed implants were determined to be substantially equivalent to predicate devices in design, function, intended use, materials, and size.

#### Conclusion

It is substantially equivalent to these other devices in design, function, material and intended use.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD/K151481](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD/K151481)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
