K120031 · K2m, Inc. · OVD · Mar 16, 2012 · Orthopedic
Device Facts
Record ID
K120031
Device Name
CHESAPEAKE SPINAL SYSTEM
Applicant
K2m, Inc.
Product Code
OVD · Orthopedic
Decision Date
Mar 16, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as a cervical intervertebral body fusion device, the Chesapeake implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients. Cervical IBF implants are intended for use at one level in the cervical spine, from C2 to T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is intended to be used in patients who have had six weeks of non-operative treatment. When used as a lumbar intervertebral body fusion device, the Chesapeake implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients. Lumbar I implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non-operative treatment. The Chesapeake spinal System may be used as a stand-alone device or in conjunction with supplemental fixation. When used as a stand-alone device the Chesapeake spacers are intended to be used with the bone screws provided. When using a 2 screw implant, 2 screws must be used. When using a 3 screw implant 3 screws must be used.
Device Story
Chesapeake Spinal System consists of PEEK spacers and titanium bone screws for intervertebral body fusion; used as stand-alone or with supplemental fixation. Spacers are hollow tubes for bone graft packing; feature machined teeth on bone-contacting surfaces for endplate engagement; include embedded tantalum markers for radiographic visualization. Available in various footprints to accommodate anatomical variations. Implanted by surgeons in clinical settings to provide spinal support and stabilization. Output is physical structural support for vertebral bodies; clinical decision-making based on patient history and radiographic confirmation of disc degeneration. Benefits include stabilization of lumbar/cervical segments to facilitate fusion.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: PEEK-OPTIMA LT1 (ASTM F2026), Tantalum (ASTM F560), Titanium Alloy (ASTM F1472), CP Titanium (ASTM F67). Form factor: hollow spacers with machined teeth and integrated bone screws. Standalone or supplemental fixation. No software or electronic components.
Indications for Use
Indicated for spinal fusion in skeletally mature patients. Cervical: one level (C2-T1) for discogenic neck pain after 6 weeks of failed non-operative treatment. Lumbar: one or two contiguous levels (L2-S1) for degenerative disc disease with up to Grade 1 spondylolisthesis after 6 months of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
K2M Aleutian Interbody Spacer Systems (K051454, K082698)
{0}------------------------------------------------
12003
MAR 1 6 2012
#### 510(k) Summary Chesapeake Spinal System K2M, Inc.
This safety and effectiveness summary for the Aleutian IBF System is provided as required per Section 513(i)(3) of the Food, Drug and Cosmetic Act.
# 1. Submitter :
- (
K2M, Inc. 751 Miller Drive SE Suite F 1 Leesburg, VA 20175
#### Contact Person : Richard W. Woods K2M, Inc. 751 Miller Drive SE, Suite FI Leesburg, VA 20175 Telephone: 703-777-3155
| Date Prepared: | July 16, 2009 |
|----------------|---------------|
|----------------|---------------|
| 2. Tradename: | Chesapeake Spinal System |
|----------------------|-----------------------------------|
| Common Name: | Intervertebral Body Fusion Device |
| Classification Name: | OVD |
| Regulation Number: | 888.3080 |
## 3. Description of the device:
The Chesapeake Spinal System consists of PEEK spacers and titanium bone screws for intervertebral body fusion, without the need for supplementary fixation. The spacers are hollow tube structures that can be packed with bone graft and allow for passage of screws for fixation to the vertebral body. The bone contacting surfaces of the implants have machined teeth which are designed to engage with the vertebral body endplates. Tantalum markers are embedded within the spacers identify the location and orientation of the implants to radiographically. Implants with footprints ranging from 24x30mm-28x36mm are available to accommodate anatomical variations. The purpose of this submission is to add 30x40mm footprints to the system.
#### Materials:
| PEEK-OPTIMA LT1 | F2026 |
|-----------------|-------|
| TANTALUM | F560. |
| TITANIUM ALLOY | F1472 |
| CP TITANIUM | F67 |
Function: The system functions as an intervertebral body fusion device to provide support and stabilization of the lumbar segments of the spine.
#### 4. Intended Use:
When used as a cervical intervertebral body fusion device, the Chesapeake implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients. Cervical IBF implants are intended for use at one level in the cervical spine, from C2 to T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is intended to be used in patients who have had six weeks of non-operative treatment.
When used as a lumbar intervertebral body fusion device, the Chesapeake implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients. Lumbar I implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non-operative treatment.
sheet 1 of 2
{1}------------------------------------------------
The Chesapeake spinal System may be used as a stand-alone device or in conjunction with supplemental fixation. When used as a stand-alone device the Chesapeake spacers are intended to be used with the bone screws provided. When using a 2 screw implant, 2 screws must be used. When using a 3 screw implant 3 screws must be used.
#### 5. Predicate or legally marketed devices which are substantially equivalent:
Documentation was provided which demonstrated that the subject Chesapeake Spinal System components are substantially equivalent to the K2M Aleutian Interbody Spacer Systems (K051454, K082698), Surgicraft Stalif System (K073109), Synthes Synfix (K062083, K072253), Spinal Elements (K083475), and Biomet Solitaire (K081501).
#### 6. Comparison of the technological characteristics of the device to predicate and legally marketed devices:
There are no significant differences between the Chesapeake Spinal System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.
sheet 2. of 2
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird with three wing-like shapes extending upwards and to the right.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR 1 6 2012
K2M, Inc. % Ms. Nancy Giezen Manager, Regulatory Affairs 751 Miller Drive SE, Suite F1 Leesburg, Virginia 20175
Re: K120031
Trade/Device Name: Chesapeake Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: February 16, 2012 Received: February 17, 2012
#### Dear Ms. Giezen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
# Page 2 - Ms. Nancy Giezen
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
# 510(k) Number (if known):
#### Device Name : Chesapeake Spinal System
#### Indications For Use :
When used as a cervical intervertebral body fusion device, the Chesapeake implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients. Cervical IBF implants are intended for use at one level in the cervical spine, from C2 to T1, for the treatment of cervical disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is intended to be used in patients who have had six weeks of non-operative treatment.
When used as a lumbar intervertebral body fusion device, the Chesapeake implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients. Lumbar I implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non-operative treatment.
The Chesapeake spinal System may be used as a stand-alone device or in conjunction with supplemental fixation. When used as a stand-alone device the Chesapeake spacers are intended to be used with the bone screws provided. When using a 2 screw implant, 2 screws must be used. When using a 3 screw implant 3 screws must be used.
Prescription use X
OR ·
Over-the-counter use ( PER 21 CFR 801.109)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED )
Concurrence of CDRH, Office of Device Evaluation ( ODE )
Aout th to. RPJ
Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number_K12003 ﺎ