Patriot-SI Posterior Implant System

K232259 · Spinal Simplicity, LLC · OUR · Apr 19, 2024 · Orthopedic

Device Facts

Record IDK232259
Device NamePatriot-SI Posterior Implant System
ApplicantSpinal Simplicity, LLC
Product CodeOUR · Orthopedic
Decision DateApr 19, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Patriot-SI Posterior Implant System is intended for sacroiliac joint fusion including sacroiliac joint disruptions and degenerative sacroiliitis. When the Patriot Implant System (i.e., an "in-line" or "in-line" or "intraarticular" device) is implanted, it must be used with a Spinal Simplicity Liberty-SI Lateral System device (i.e., a "transfixing" device) implanted across the same sacroiliac joint to create a hybrid SI joint fusion construct.

Device Story

Minimally invasive sacroiliac (SI) joint fusion implant; designed for posterior approach into SI joint space. Device features external threads for fixation into sacrum and ilium. Must be used with Spinal Simplicity Liberty-SI Lateral System (transfixing device) to create hybrid fusion construct. Implanted by surgeons using dedicated surgical instruments. Bone graft materials may be used. Provides stabilization and fusion of SI joint; intended to benefit patients with joint disruptions or degenerative sacroiliitis.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including ASTM F3574 (static shear, static/dynamic torsion, static pushout, driving torque) and ASTM F2077 (static/dynamic compression-shear) testing, plus cadaver testing.

Technological Characteristics

Titanium alloy (Ti-6Al-4V-ELI per ASTM F3001); optional hydroxyapatite (HA) coating (ASTM F1185). Threaded metallic bone fixation fastener. Sterile, single-use. No software or electronic components.

Indications for Use

Indicated for sacroiliac joint fusion in patients with sacroiliac joint disruptions and degenerative sacroiliitis. Must be used in conjunction with the Spinal Simplicity Liberty-SI Lateral System to create a hybrid SI joint fusion construct.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 19, 2024 Spinal Simplicity LLC Adam Rogers Vice President of Regulatory and Engineering 6363 College Blvd, Ste 320 Overland Park, Kansas 66211 Re: K232259 Trade/Device Name: Patriot-SI Posterior Implant System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR Dated: March 11, 2024 Received: March 11, 2024 Dear Adam Rogers: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Colin O'neill -S Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K232259 Device Name Patriot-SI Posterior Implant System #### Indications for Use (Describe) The Patriot-SI Posterior Implant System is intended for sacroiliac joint fusion including sacroiliac joint disruptions and degenerative sacroiliitis. When the Patriot Implant System (i.e., an "in-line" or "in-line" or "intraarticular" device) is implanted, it must be used with a Spinal Simplicity Liberty-SI Lateral System device (i.e., a "transfixing" device) implanted across the same sacroiliac joint to create a hybrid SI joint fusion construct. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | 510(k) #: K232259 | 510(k) Summary | Page 1 of 2<br>Prepared on: 2024-03-13 | |-------------------|----------------|----------------------------------------| |-------------------|----------------|----------------------------------------| Contact Details | Applicant Name | Spinal Simplicity LLC | 21 CFR 807.92(a)(1) | |-----------------------------|----------------------------------------------------------------|---------------------| | Applicant Address | 6363 College Blvd Ste 320 Overland Park KS 66211 United States | | | Applicant Contact Telephone | 913-451-4414 | | | Applicant Contact | Mr. Adam Rogers | | | Applicant Contact Email | arogers@spinalsimplicity.com | | Device Name | Device Trade Name | Patriot-SI Posterior Implant System | 21 CFR 807.92(a)(2) | |---------------------|----------------------------------------------------|---------------------| | Common Name | Smooth or threaded metallic bone fixation fastener | | | Classification Name | Sacroiliac Joint Fixation | | | Regulation Number | 888.3040 | | | Product Code | OUR | | Legally Marketed Predicate Devices | | | 21 CFR 807.92(a)(3) | |-------------|----------------------------------------------------------|---------------------| | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K190580 | NADIA® SI Fusion System | OUR | Device Description Summary | | | 21 CFR 807.92(a)(4) | |--|--|---------------------| |--|--|---------------------| The Patriot-Sl Posterior Implant System is a minimally invasive sacrolliac joint fusion implant that is intended for implantation on a trajectory in line with the joint space (i.e., the device is an in-line or intra-articular implant). The Patriot Implant System device is required to be implanted with the Spinal Simplicity Liberty-Sl Lateral System device implanted across the same sacroiliac joint to create a hybrid SI joint fusion construct for the purpose of stabilizing and fusing the sacroiliac joint. It is available in one size and may be implanted using the designated surgical instruments through a posterior approach into the Sl joint space, fixating on both the sacrum and ilium via external threads. Bone graft materials may be used with the Patriot-Sl Posterior Implant System. The Patriot-Sl Posterior Implant System device is manufactured titanium alloy, Ti-6AI-4V-ELI per ASTM F3001 and is offered with or without a hydroxyapatite (HA) coating per ASTM F1185. The Patriot-SI Posterior Implant System device is provided sterile and individually packed. ## Intended Use/Indications for Use The Patriot-Sl Posterior Implant System is intended for sacroiliac joint fusion including sacroiliac joint disruptions and degenerative sacroiliitis. When the Patriot-Sl Posterior Ine., an "in-line" or "intra-articular" device) is implanted, it must be used with a Spinal Simplicity Liberty-Sl Lateral System device (i.e., a "transfixing" device) implanted across the same sacroiliac joint to create a hybrid SI joint fusion construct. 21 CFR 807.92(a)(5) 21 CFR 807.92(a)(5) ## Indications for Use Comparison The Patriot-SI Posterior Implant System has the same indications for use as the predicate device. {4}------------------------------------------------ K232259 ### Technological Comparison 21 CFR 807.92(a)(6) The Patriot-SI Posterior Implant System is as safe and effective as the identified predicate device. The Patriot-Sl Posterior Implant System has the same technological characteristics and surgical approach as its predicate device. Performance testing demonstrates that the Patriot-SI Posterior Implant System is substantially equivalent to the predicate device. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non clinical mechanical tests were performed to support substantial equivalence of the Patriot Implant System. Cadaver testing was also performed to support the Patriot-Sl Posterior Implant System. The following tests were performed: ASTM F3574 Static Shear Testing ASTM F3574 Static and Dynamic Torsion Testing ASTM F3574 Static Pushout Testing ASTM F3574 Driving Torque Testing ASTM F2077 Static and Dynamic Compression-Shear Testing Clinical performance testing was not performed for this submission. Spinal Simplicity concludes that the performance testing provided in this 510(k) application demonstrates that the Patriot SJ Posterior lmplant System is capable of performing as intended and is as safe and effective as the legally marketed predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...