← Product Code [OUR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR) · K222448

# UNITY Sacroiliac Joint Fixation System (K222448)

_Dio Medical Corporation · OUR · Oct 13, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K222448

## Device Facts

- **Applicant:** Dio Medical Corporation
- **Product Code:** [OUR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR.md)
- **Decision Date:** Oct 13, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The UNITY Sacroiliac Joint Fixation System is indicated for use in skeletally mature patients for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

## Device Story

UNITY Sacroiliac Joint Fixation System consists of metallic screws designed to facilitate sacroiliac joint fusion. System includes cannulated screws in various diameters and lengths, categorized into three types: standard thread (with/without slots), lag (with/without slots), and washer (with slots). Device is used by surgeons in clinical settings to stabilize the sacroiliac joint. Screws are implanted to provide mechanical fixation, promoting fusion in patients with joint disruptions or degenerative sacroiliitis. Output is the physical stabilization of the joint, aiding patient recovery and pain reduction.

## Clinical Evidence

No clinical data. Substantial equivalence established via comparison of technological characteristics, design features, and mechanical performance. Device does not create a new worst-case for performance; additional testing deemed unnecessary.

## Technological Characteristics

Metallic bone fixation fasteners (screws). Cannulated design. Available in standard thread, lag, and washer configurations with optional slots. Various diameters and lengths to accommodate patient anatomy. Class II device (21 CFR 888.3040).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Huvexel Co. Ltd. - UNITY Sacroiliac Joint Fixation System ([K173201](/device/K173201.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 13, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the agency's acronym in a blue square, followed by the full name "U.S. Food & Drug Administration" in blue text.

Dio Medical Corporation Milan George VP of R&D 2100 Campus Lane Suite 100 East Norriton, Pennsylvania 19403

Re: K222448

Trade/Device Name: UNITY Sacroiliac Joint Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR Dated: August 12, 2022 Received: August 15, 2022

Dear Milan George:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2023<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|

**Indications for Use**

| 510(k) Number (if known)                        | K222448                                                                                                                                                                                                              |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                                     | UNITY Sacroiliac Joint Fixation System                                                                                                                                                                               |
| Indications for Use (Describe)                  | The UNITY Sacroiliac Joint Fixation System is indicated for use in skeletally mature patients for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.       |
| Type of Use (Select one or both, as applicable) | <span> <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D)             </span> <span>                      Over-The-Counter Use (21 CFR 801 Subpart C)             </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to
review instructions, search existing data sources, gather and maintain the data needed and complete and review the
collection of information. Send comments regarding this burden estimate or any other aspect of this information
collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA)  

 Staff PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a
collection of information unless it displays a currently valid OMB number."*

| FORM FDA 3881 (6/20) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 |
|----------------------|-------------|----------------------------------------|
|----------------------|-------------|----------------------------------------|

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

## 510(k) SUMMARY

## Dio Medical UNITY Sacroiliac Joint Fixation System

| Sponsor:                         | Manufacturer:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Dio Medical Corp.<br>2100 Campus Lane, Suite 100<br>East Norriton, PA 19403 |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
|                                  | Official Contact:<br>Email:<br>Phone:                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Milan George<br>mgeorge@dio-us.com<br>1-877-394-5407 ext.102                |
| Date Prepared:                   | August 12, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                             |
| Device Name:<br>Common Name:     | UNITY Sacroiliac Joint Fixation System<br>Sacroiliac Joint Fixation Screw                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                             |
| Classification<br>Name:          | Smooth or threaded metallic bone fixation fastener                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                             |
| Classification<br>Number:        | 21 CFR 888.3040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                             |
| Product Code/<br>Classification: | OUR, class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                             |
| Description:                     | The UNITY Sacroiliac Joint Fixation System consists of screws<br>designed to enhance sacroiliac joint fusion. The UNITY Sacroiliac<br>Joint Fixation System is offered in various diameters, lengths, and<br>three screw types in cannulated form to accommodate patient<br>anatomy. The three design types of the subject device are:<br>1. Standard Thread Screw (with and without slots)<br>2. Lag Screw (with and without slots) and<br>3. Washer Screw (with slots)                            |                                                                             |
| Indications<br>for Use:          | The UNITY Sacroiliac Joint Fixation System is indicated for use in<br>skeletally mature patients for sacroiliac joint fusion for conditions<br>including sacroiliac joint disruptions and degenerative sacroiliitis.                                                                                                                                                                                                                                                                                |                                                                             |
| Predicate<br>Device:             | Huvexel Co. Ltd. - UNITY Sacroiliac Joint Fixation System<br>(K173201)                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                             |
| Substantial<br>Equivalence:      | The UNITY Sacroiliac Joint Fixation System has the same intended<br>use, indications for use, and principles of operation as the primary<br>predicate device.  It has similar technological characteristics as its<br>predicate device.  There are no technological differences between<br>the subject device and its predicate device resulting in no new<br>issues of safety or effectiveness.  Thus, the UNITY Sacroiliac Joint<br>Fixation System is substantially equivalent to the predicate. |                                                                             |
| Performance<br>Data:             | The UNITY Sacroiliac Joint Fixation does not create a new worst<br>case for device performance; additional testing is not needed.                                                                                                                                                                                                                                                                                                                                                                   |                                                                             |
| Conclusion:                      | Based upon a comparison of technological characteristics,<br>intended use, design features, and mechanical performance, the<br>Dio Medical UNITY Sacroiliac Joint Fixation System does not raise<br>any new safety or efficacy concerns and has demonstrated                                                                                                                                                                                                                                        |                                                                             |

{4}------------------------------------------------

substantial equivalence to the identified predicates.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K222448](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K222448)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
