← Product Code [OUR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR) · K152237

# The Entasis Dual-Lead Sacroiliac Implant (K152237)

_Corelink, LLC · OUR · Feb 4, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K152237

## Device Facts

- **Applicant:** Corelink, LLC
- **Product Code:** [OUR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR.md)
- **Decision Date:** Feb 4, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Entasis™ Dual-Lead Sacroiliac Implant system is intended for sacroiliac joint fusion for conditions including degenerative sacroilitis and sacroiliac joint disruptions.

## Device Story

Entasis™ Dual-Lead Sacroiliac Implant system consists of titanium alloy bone fixation screws. Device used by surgeons in clinical settings to achieve sacroiliac joint fusion. Screws implanted across sacroiliac joint to provide mechanical stabilization and facilitate fusion. Output is physical fixation of joint. Benefits include stabilization of joint to address degenerative sacroilitis or disruptions.

## Clinical Evidence

Bench testing only. Testing included static torsion, static pullout, static driving torque, static three-point bending, and dynamic three-point bending per ASTM F543 and ASTM F2193 standards.

## Technological Characteristics

Dual-lead bone fixation screws; Material: Titanium (Ti-6Al-4V ELI) per ASTM F136; Dimensions: 30-70mm length, 7-11.5mm diameter; Principle: Mechanical bone fixation fastener.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- 6.5mm Cannulated Screw ([K021932](/device/K021932.md))
- Silex ([K140079](/device/K140079.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center is an abstract image of a bird-like figure, composed of three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 4, 2016

CoreLink, LLC % Kenneth Maxwell II Regulatory and Quality Specialist Empirical Testing Corporation 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K152237

Trade/Device Name: The Entasis™ Dual-Lead Sacroiliac Implant Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR Dated: December 31, 2015 Received: January 4, 2015

Dear Mr. Maxwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Lori A. Wiggins -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120      |
|-----------------------------------------|---------------------------------------|
| Food and Drug Administration            | Expiration Date: January 31, 2017     |
| Indications for Use                     | See PRA Statement on last page.       |
| 510(k) Number (if known)                | K152237                               |
| Device Name                             | Entasis™ Dual-Lead Sacroiliac Implant |
| Indications for Use (Describe)          |                                       |

The Entasis™ Dual-Lead Sacroiliac Implant system is intended for sacroiliac joint fusion for
conditions including degenerative sacroilitis and sacroiliac joint disruptions.

| Type of Use (Select one or both, as applicable)                                                                                                                                                       |  |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div><label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label></div> |  |

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) |  |
|------------------------------------------------------------------------------|--|
|------------------------------------------------------------------------------|--|

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 5.510(K) SUMMARY

| Submitter's Name:          | CoreLink, LLC                                       |
|----------------------------|-----------------------------------------------------|
| Submitter's Address:       | 7606 Forsyth Blvd                                   |
|                            | St. Louis, MO 63105                                 |
| Submitter's Telephone:     | (888) 349-7808                                      |
| Contact Person:            | Kenneth C. Maxwell II                               |
|                            | Empirical Testing Corp.                             |
|                            | 719.291.6874                                        |
| Date Summary was Prepared: | 04 February 2016                                    |
| Trade or Proprietary Name: | Entasis™ Dual-Lead Sacroiliac Implant               |
| Common or Usual Name:      | Smooth or threaded metallic bone fixation fastener  |
| Classification:            | Class II per 21 CFR §888.3040 Device Classification |
| Product Code:              | OUR                                                 |
| Classification Panel:      | Division of Orthopedic Devices                      |

### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION

The Entasis™ Dual-Lead Sacroiliac Implant system is composed of dual-lead sacroiliac screws manufactured from titanium (Ti-6AI-4V ELI) per ASTM F136. The screws are available in lengths of 30-70mm and diameters of 7-11.5mm.

#### INDICATIONS FOR USE

The Entasis™ Dual-Lead Sacroiliac Implant system is intended for sacroiliac joint fusion for conditions including degenerative sacroilitis and sacroiliac joint disruptions.

The indications for use for the CoreLink, LLC Entasis™ Dual-Lead Sacroiliac Implant is similar to that of the predicates listed in Table 5-1 Predicate Devices.

#### TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have similar technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. The following characteristics are similar between the subject and predicate devices:

- Indications for Use
- Materials of Manufacture ●
- Principles of Operation

#### Table 5-1: Predicate Devices

| 510k Number | Trade or Proprietary or Model<br>Name | Manufacturer | Predicate<br>Type |
|-------------|---------------------------------------|--------------|-------------------|
| K021932     | 6.5mm Cannulated Screw                | Synthes      | Primary           |
| K140079     | Silex                                 | X-spine      | Additional        |

{5}------------------------------------------------

#### Performance Data

The Entasis™ Dual-Lead Sacroiliac Implant has been tested in the following test modes:

- Static torsion per ASTM F543 ●
- Static pullout per ASTM F543 ●
- Static driving torque per ASTM F543 ●
- Static three-point bending per ASTM F2193 ●
- Dynamic three-point bending per ASTM F2193 ●

The results of this non-clinical testing show that the strength of the Entasis™ Dual-Lead Sacroiliac Implant is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

#### CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Entasis™ Dual-Lead Sacroiliac Implant is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K152237](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K152237)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
