The SQUALE devices are indicated for use in cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The SQUALE devices are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, and are to be implanted via an open, anterior approach. The SQUALE devices are intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine. This cervical device is to be used in patients who have had six weeks of non-operative treatment.
Device Story
SQUALE is a cervical interbody fusion device; consists of hollow PEKK cages; implanted via open anterior approach; used to support and correct vertebral bodies during fusion. Device filled with autogenous or allogenic bone graft; requires supplemental fixation. Used by surgeons in clinical settings for patients with discogenic neck pain. Provides structural support to facilitate spinal fusion; benefits patient by stabilizing vertebral segment.
Clinical Evidence
Bench testing only. Performance testing conducted per ASTM F2077 and ASTM F2267, including static/dynamic axial compression, compression shear, torsion, subsidence, and expulsion testing. Results demonstrate substantial equivalence to predicate devices.
Technological Characteristics
Cervical cages constructed of Polyetherketoneketone (PEKK) compliant with ASTM F2820. Features hollow geometry for bone graft containment. Dimensions vary in width and height. Standalone implant; non-powered; non-resorbable.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one level (C2-C3 to C7-T1) who have failed six weeks of non-operative treatment. Requires use with autogenous or allogenic bone graft and supplemental cervical fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K121649 — CONSTRUX MINI PEEK TI SPACER SYSTEM · Orthofix · Nov 29, 2012
K172320 — Neurostructures Cavetto® Cervical Cage System · Neurostructures, Inc. · Feb 26, 2018
K212358 — Romero Cervical Cage · Spineup, Inc. · Jan 10, 2022
K162402 — 4CIS® Marlin ACIF Cage System · Solco Biomedical Co., Ltd. · Feb 16, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 28, 2023
Implanet, S.A. % Kelliann Payne Partner Hogan Lovells US LLP 1735 Market Street, 23rd Floor Philadelphia, Pennsylvania 19103
Re: K230026
Trade/Device Name: SQUALE Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: March 9, 2023 Received: March 9, 2023
Dear Kelliann Payne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Katherine D. Kavlock -S for
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K230026
Device Name
#### SQUALE
Indications for Use (Describe)
The SQUALE devices are indicated for use in cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The SQUALE devices are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, and are to be implanted via an open, anterior approach.
The SQUALE devices are intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine. This cervical device is to be used in patients who have had six weeks of non-operative treatment.
Type of Use (Select one or both, as applicable)
図 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K230026 510(k) SUMMARY Implanet, S.A.'s SQUALE
Implanet. S.A. Technopole Bordeaux Montesquieu Allée François Magendie 33650 Martillac Phone: + 33 557 995 555
Contact Person: Régis Le Couëdic
Date Prepared: March 12, 2023
Name of Device: SQUALE
Common or Usual Name: Cervical Intervertebral Body Fusion Device
Regulatory Class: 21 CFR 888.3080
Product Code: ODP
## Predicate Devices
- Primary Predicate: Stryker's AVS AS PEEK Spacer (K142251) ●
- Additional Predicate: SpineUp, Inc Romero Cervical Cage (K212358)
- Additional Predicate: RTI Surgical, Fortilink™-C with TETRAfuse™ 3D Technology ● (K163673)
## Intended Use / Indications for Use
The SQUALE devices are indicated for use in cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The SQUALE devices are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, and are to be implanted via an open, anterior approach.
The SQUALE devices are intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine. This cervical device is to be used in patients who have had six weeks of non-operative treatment.
## Technological Characteristics
The SQUALE device consists of Polyetherketoneketone (PEKK) (compliant with ASTM F2820 standard) cervical cages of various widths and heights, which can be inserted between two cervical or cervico-thoracic vertebral bodies to give support and correction during cervical interbody fusion surgeries. The hollow geometry of the implants allows them to be filled with autogenous and/or allogenic bone graft.
{4}------------------------------------------------
#### Performance Data
Performance testing was conducted per ASTM F2077 and ASTM F2267. Specifically, Implanet performed static and dynamic axial compression testing, static and dynamic compression shear testing, static and dynamic torsional testing, subsidence testing, and expulsion testing. The results of these studies were determined to be substantially equivalent to legally marketed devices.
#### Substantial Equivalence
The SQUALE is as safe and effective as the primary predicate device, Stryker AS AVS PEEK spacer. The SQUALE has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the SQUALE and its predicate device raise no new issues of safety or effectiveness and are covered by the proposed secondary predicate device. Performance data demonstrate that the SQUALE is as safe and effective as the predicate device. Thus, the SQUALE is substantially equivalent.
#### Conclusions
The SQUALE is intended to treat degenerative disc disease, and bench testing demonstrates that it is as safe, as effective, and performs as well as the primary predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.