Ventana C Spinal System
K222833 · Spinal Elements, Inc. · ODP · May 19, 2023 · Orthopedic
Device Facts
| Record ID | K222833 |
| Device Name | Ventana C Spinal System |
| Applicant | Spinal Elements, Inc. |
| Product Code | ODP · Orthopedic |
| Decision Date | May 19, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3080 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Ventana™ C devices are intended for spinal fusion at one or two contiguous levels in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Implanted via an open, anterior approach for the C2-C3 disc space to the C7-T1 disc space and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The device is intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems). Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with these devices.
Device Story
Ventana™ C is an intervertebral body fusion device for cervical spine surgery. Device is box-shaped with internal cavities for autograft or allogenic bone graft; includes a lid to secure graft material. Superior and inferior surfaces feature sharp teeth to prevent migration. Implanted via open, anterior approach by surgeons. Used in conjunction with cleared supplemental anterior plate fixation systems. Provides structural support to facilitate spinal fusion in patients with degenerative disc disease. Benefits include stabilization of disc space and promotion of bone fusion.
Clinical Evidence
Bench testing only. Performance evaluated per ASTM F2077 (static/dynamic compression and torsion) and ASTM F2267 (subsidence). Results demonstrate performance equivalent to legally marketed predicate devices.
Technological Characteristics
Intervertebral body fusion device; additively manufactured titanium alloy (Ti-6Al-4V ELI) per ASTM F3001. Box-shaped geometry with graft containment lid and surface teeth for fixation. Mechanical testing per ASTM F2077 and ASTM F2267.
Indications for Use
Indicated for skeletally mature patients with cervical degenerative disc disease (DDD) requiring spinal fusion at one or two contiguous levels (C2-T1). Requires prior 6-week non-operative treatment regimen. Contraindicated in patients without discogenic neck pain or radiographic evidence of degeneration.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
- Crystal® devices (K153352)
- Lucent® 3D devices (K203254)
- BEE Cage (K200429)
- Atlas Spine - HiJAK (K192570)
Related Devices
- K171075 — Calix-C Cervical Interbody Spacer · X-Spine Systems, Inc. · Aug 1, 2017
- K141314 — SCARLET AC-T SECURED ANTERIOR CERVICAL CAGE AND ASSOCIATED INSTRUMENTATION · Spineart · Sep 29, 2014
- K173832 — InTice-C Porous Ti Cervical Interbody System · X-Spine Systems, Inc. · May 17, 2018
- K150053 — STALIF C; STALIF C-Ti · Centinel Spine, Inc. · Jun 24, 2015
- K192132 — SeaSpine Beachside System · SeaSpine Orthopedics Corporation · Dec 3, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 19, 2023
Spinal Elements, Inc. Dr. Julie Lamothe Vice President of Regulatory Affairs & Quality Assurance 3115 Melrose Dr., Suite 200 Carlsbad, California 92010
Re: K222833
Trade/Device Name: Ventana™ C Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: April 17, 2023 Received: April 19, 2023
Dear Dr. Lamothe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22833
Device Name Ventana™ C
#### Indications for Use (Describe)
Ventana™ C devices are intended for spinal fusion at one or two contiguous levels in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Implanted via an open, anterior approach for the C2-C3 disc space to the C7-T1 disc space and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
The device is intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems).
Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with these devices.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K22833 510(k) Summary Ventana™ C
| Manufacturer Identification | |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by: | Spinal Elements, Inc.<br>3115 Melrose Dr., Suite 200<br>Carlsbad, CA 92010<br>P. 760-607-0121<br>F. 760-607-0125 |
| Contact Information: | Julie Lamothe<br>Spinal Elements, Inc.<br>3115 Melrose Dr., Suite 200<br>Carlsbad, CA 92010<br>760-607-1816<br>jlamothe@spinalelements.com |
| Date Prepared: | September 16, 2022 |
| Proprietary Name | Ventana™ C |
| Regulation Name | Intervertebral Body Fusion Device |
| Device Classification | 21 CFR 888.3080 (Appliance, Fixation Spinal<br>Intervertebral Body) |
| Proposed Regulatory Class | Class II |
| Device Product Code | ODP |
# Device Description
The Spinal Elements' Ventana C Interbody System is an intervertebral body fusion device for use in cervical spine surgery. The device is generally box-shaped with various holes throughout its geometry to allow for the placement of autograft or allogenic bone graft. It includes a lid which further facilitates installation of bone graft, and which is shut prior to implantation. The exterior of the device has "teeth" or other generally sharp engagement members on the superior and inferior surfaces to help prevent the device from migrating once it is implanted.
The devices submitted herein are additively manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F3001.
## Indication for Use
Ventana" C devices are intended for spinal fusion at one or two contiguous levels in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Implants are to be implanted via an open, anterior
{4}------------------------------------------------
approach for the C2-C3 disc space to the C7-T1 disc space and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
The device is intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems).
Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with these devices.
# Substantial Equivalence
The subject devices are substantially equivalent in indications for use, surgical technique, design features, manufacturing methods and raw materials to the following predicate devices:
Primary Predicate: Crystal® devices (K153352) Secondary Predicate: Lucent® 3D devices (K203254) Secondary Predicate: BEE Cage (K200429) Secondary Predicate: Atlas Spine - HiJAK (K192570)
# Technological Characteristics
The subject device was established as substantially equivalent to another predicate device cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate device through comparison in areas including design, materials, chemical composition, energy source, intended use, operating principle and function.
## Performance Data
Performance testing included:
- Static compression per ASTM F2077
- Dynamic compression per ASTM F2077
- Static Torsion per ASTM F2077 ●
- Dynamic Torsion per ASTM F2077 ●
- Subsidence per ASTM F2267
All data indicates that the device will perform equivalently to a legally marketed predicate device.
# CONCLUSIONS
Based on the indication for use, technological characteristics, and comparison to predicate devices, the subject device has been shown to be substantially equivalent to the aforementioned predicate devices cleared by FDA for commercial distribution in the United States.