← Product Code [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP) · K212038

# K&J IVA (ACIF, DLIF, PLIF, TLIF, ALIF) PEEK Cage (K212038)

_K&J Consulting Corporation · ODP · Aug 27, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K212038

## Device Facts

- **Applicant:** K&J Consulting Corporation
- **Product Code:** [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP.md)
- **Decision Date:** Aug 27, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The K&J IVA (ACIF) PEEK Cage is intended to be used for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-TI disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage. The K&J IVA (PLIF, TLIF, DLIF and ALIF) PEEK Cage is indicated for intervertebral body fusion of the lumbar spine, L2 to S1, in skeletally mature patients who have had six months of nonoperative treatment. The device is intended for use at one level or two continuous levels for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

## Device Story

Intervertebral body fusion device; PEEK and Tantalum construction; implanted surgically to facilitate spinal fusion; used in conjunction with supplemental fixation and autograft; intended for skeletally mature patients with degenerative disc disease; cervical or lumbar spine applications; device provides structural support to disc space during fusion process.

## Clinical Evidence

No clinical data; bench testing not required as the subject device is identical to the previously cleared predicate device (K162220).

## Technological Characteristics

Materials: PEEK and Tantalum. Form factor: Intervertebral cage for ACIF, PLIF, TLIF, DLIF, and ALIF procedures. Design and manufacturing are identical to the predicate device.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- DIO Medical IVA (ACIF, DLIF, PLIF, TLIF, ALIF) PEEK Cage ([K162220](/device/K162220.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 27, 2021

K&J Consulting Corporation % Barry Sands President RQMIS Inc. 110 Haverhill Road, Suite 524 Amesbury, Massachusetts 01913

Re: K212038

Trade/Device Name: K&J IVA (ACIF, DLIF, PLIF, TLIF, ALIF) PEEK Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, MAX Dated: June 29, 2021 Received: June 30, 2021

Dear Barry Sands:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K212038

## Device Name

K&J IVA (ACIF, PLIF. TLIF. DLIF and ALIF) PEEK Cage

### Indications for Use (Describe)

The K&J IVA (ACIF) PEEK Cage is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage.

The K&J IVA (PLIF, TLIF, DLIF and ALIF) PEEK Cage is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who had six months of non-operative treatment. The device is intended for use at one level or two continuous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

| Type of Use (Select one or both, as applicable)                                                 |                                             |
|-------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

# DIO Medical Co., Ltd's IVA PEEK Cage

### l. Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

K&J Consulting Corporation 2260 Glenview Dr, Lansdale PA 19446 വട Phone: (267)-438-9293 Email: kj.pd2021@gmail.com

#### ll. Contact Person Name and Address:

Barry Sands RQMIS, Inc. 110 Haverhill Road, Suite 524 Amesbury, MA. 01913 Phone: (978) 358-7307 Facsimile: (978) 477-0206 Email: barrysands@rqmis.com

#### lll. Subject Device:

| Trade Name:          | K&J IVA (ACIF, DLIF, PLIF, TLIF, ALIF) PEEK Cage                                                                              |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Common Names:        | Intervertebral body fusion device                                                                                             |
| Classification:      | Class II                                                                                                                      |
| Classification Name: | Intervertebral Body Fusion Device, Cervical (Product Code ODP)<br>Intervertebral Body Fusion Device, Lumbar (Product Code MAX |
| Regulation Number:   | 21 C.F.R. §888.3080                                                                                                           |

#### IV. Predicate Devices:

DIO Medical IVA (ACIF, DLIF, PLIF, TLIF, ALIF) PEEK Cage (K162220)

{4}------------------------------------------------

#### V. Device Description:

# Intended Use / Indications for Use

Indications for Use:

The K&J IVA (ACIF) PEEK Cage is intended to be used for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-TI disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage

The K&J IVA (PLIF, TLIF, DLIF and ALIF) PEEK Cage is indicated for intervertebral body fusion of the lumbar spine, L2 to S1, in skeletally mature patients who have had six months of nonoperative treatment. The device is intended for use at one level or two continuous levels for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

# Technological Characteristics

The K&J IVA (ACIF, PLIF, TLIF, DLIF and ALIF) PEEK Cage consists of PEEK+Tantalum which is identical to its predicate devices. The design, material composition and manufacturing are same as the predicate.

## VI. Comparison of Technological Characteristics with the Predicate Device: (Substantial Equivalence)

The K&J IVA (ACIF, PLIF, TLIF, DLIF and ALIF) PEEK Cage is identical to the predicate device and is as safe and effective as the DIO Medical IVA (ACIF, PLIF, TLIF, DLIF and ALIF) PEEK Cage. The Subject device has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. There are no technological differences between the Subject device and its predicate devices resulting in no new issues of safety or effectiveness. Thus, the K&J IVA (ACIF, PLIF, TLIF, DLIF and ALIF) PEEK Cage is identical/substantially equivalent.

{5}------------------------------------------------

#### VII. Performance Data

The subject and predicate devices are identical and therefore, no performance testing is required. Submission is only transferring name of a system that has already been cleared under K162220. No testing is required.

#### VIII. Conclusion:

The K&J IVA (ACIF, PLIF, TLIF, DLIF and ALIF) PEEK Cage has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. Thus, the subject device is identical/substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K212038](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K212038)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).