← Product Code [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP) · K172926

# SeaSpine Vu cPOD, Zuma-C, Complete Cervical (K172926)

_SeaSpine Orthopedics Corporation · ODP · Dec 21, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K172926

## Device Facts

- **Applicant:** SeaSpine Orthopedics Corporation
- **Product Code:** [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP.md)
- **Decision Date:** Dec 21, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Vu c·POD: The Vu coPOD Intervertebral Body Fusion Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical degenerative disc disease (DDD) at one level from C2-C3 to C7-T1. Cervical degenerative disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Vu c•POD implants are to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and /or corticocancellous bone and implanted via an open, anterior approach. The cervical device is to be used in patients who have had six weeks of non-operative treatment. The Vu coPOD Intervertebral Body Fusion Device is intended for use with supplemental internal fixation systems. Complete Cervical: The Complete Cervical Intervertebral Body Fusion Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Complete Cervical implants are to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and /or corticocancellous bone and implanted via an anterior approach. The cervical device is to be used in patients who have had six (6) weeks of non-operative treatment. The cervical device is to be used with two titanium alloy screws which accompany the implant. Zuma-C: Zuma-C is a stand-alone anterior cervical interbody fusion device intended for use as an adjunct to fusion at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as discogenic neck pain with degeneration of the disc confirmed by history and radiographic studies). Patients should have received at least six weeks of non-operative treatment prior to treatment with the device. Zuma-C is to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and /or corticocancellous bone and implanted via an open, anterior approach. Zuma-C is intended to be used with the bone screw fixation provided and requires no additional fixation.

## Device Story

Intervertebral fusion devices (Vu c-POD, Complete Cervical, Zuma-C) act as disc spacers to hold bone graft and promote fusion in cervical spine; implanted via anterior approach by surgeons. Devices are box-shaped with central canal for autograft/allogeneic bone graft. Zuma-C functions as stand-alone system with integrated bone screws; Vu c-POD requires supplemental internal fixation; Complete Cervical uses two titanium alloy screws. Benefit: stabilization of cervical segment to treat discogenic neck pain. No software or AI components.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and mechanical performance comparisons.

## Technological Characteristics

Materials: PEEK (ASTM F2026), Tantalum markers (ASTM F560), Titanium alloy (ASTM F136). Form factor: Box-shaped interbody spacer with central graft canal. Sterilization: Steam sterilization. No software or electronic components.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Cambria System ([K171046](/device/K171046.md))
- Vu c-POD ([K111675](/device/K111675.md), [K101363](/device/K101363.md), [K050058](/device/K050058.md), [K032064](/device/K032064.md))
- Complete Cervical ([K102323](/device/K102323.md))
- Zuma-C ([K092521](/device/K092521.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" underneath.

December 21, 2017

SeaSpine Orthopedics Corporation Gina Flores Specialist, Regulatory Affairs 5770 Armada Drive Carlsbad, California 92008

Re: K172926

Trade/Device Name: SeaSpine Vu cPOD, Zuma-C, Complete Cervical Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, OVE Dated: September 22, 2017 Received: September 25, 2017

Dear Gina Flores:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Katherine D. Kavlock -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K172926

Device Name SeaSpine Vu c.POD

#### Indications for Use (Describe)

The Vu c POD Intervertebral Body Fusion Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical degenerative disc disease (DDD) at one level from C2-C3 to C7-T1. Cervical degenerative disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

The Vu c.POD implants are to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/ or corticocancellous bone and implanted via an open, anterior approach. The cervical device is to be used in patients who have had six weeks of non-operative treatment. The Vu c•POD Intervertebral Body Fusion Device is intended for use with supplemental internal fixation systems.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K172926

Device Name SeaSpine Zuma-C

#### Indications for Use (Describe)

Zuma-C is a stand-alone anterior cervical interbody fusion device intended for use as an adjunct to fusion at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as discogenic neck pain with degeneration of the disc confirmed by history and radiographic studies). Patients should have received at least six weeks of non-operative treatment prior to treatment with the device. Zuma-C is to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and implanted via an open, anterior approach. Zuma-C is intended to be used with the bone screw fixation provided and requires no additional fixation.

| Type of Use (Select one or both, as applicable)                                                                                  |                                                                                                               |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K172926

Device Name SeaSpine Complete Cervical

#### Indications for Use (Describe)

The Complete Cervical Intervertebral Body Fusion Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

The Complete Cervical implants are to be used with autograft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and implanted via an anterior approach. The cervical device is to be used in patients who have had six (6) weeks of non-operative treatment. The cervical device is to be used with two titanium alloy screws which accompany the implant.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

# 510(k) Summary

# Contact Details

| Applicant Name:      | SeaSpine Orthopedics Corporation                                                     |  |  |
|----------------------|--------------------------------------------------------------------------------------|--|--|
| Address:             | 5770 Armada Drive, Carlsbad CA                                                       |  |  |
| Phone number:        | (760) 216-5136                                                                       |  |  |
| Fax number:          | (760) 683-6874                                                                       |  |  |
| Contact person:      | Gina Flores, Sr. Regulatory Specialist                                               |  |  |
| Email address:       | gina.flores@seaspine.com                                                             |  |  |
| Date Prepared:       | September 22, 2017                                                                   |  |  |
| Device Name          |                                                                                      |  |  |
| Trade Name:          | SeaSpine Vu c.POD, Complete Cervical, and Zuma-C                                     |  |  |
| Common Name:         | Intervertebral Fusion device with bone graft, cervical                               |  |  |
| Classification Name: | Intervertebral fusion device with integrated fixation, cervical<br>(21 CFR 888.3080) |  |  |
| Class:               | II                                                                                   |  |  |
| Product Code:        | ODP OVE                                                                              |  |  |

# Legally Marketed Predicate Devices

| 510(k) Number                        | Product Code | Trade Name        | Manufacturer                                                                                                                      |
|--------------------------------------|--------------|-------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| PRIMARY PREDICATE Device             |              |                   |                                                                                                                                   |
| K171046                              | ODP          | Cambria System    | SeaSpine Orthopedics Corporation                                                                                                  |
| Additional PREDICATE Devices         |              |                   |                                                                                                                                   |
| K111675,K101363,<br>K050058, K032064 | ODP          | Vu c-POD          | SeaSpine Orthopedics Corporation<br>(originally manufactured by Integra<br>LifeSciences, now SeaSpine<br>Orthopedics Corporation) |
| K102323                              | ODP          | Complete Cervical | SeaSpine Orthopedics Corporation<br>(originally manufactured by Theken,<br>LLC, now SeaSpine Orthopedics<br>Corporation)          |
| K092521                              | ODP          | Zuma-C            | SeaSpine Orthopedics Corporation                                                                                                  |

{6}------------------------------------------------

# Device Description

The SeaSpine Vu c-POD, Complete Cervical, and Zuma-C Systems are intervertebral fusion devices intended to act as a disc spacer and hold bone graft to promote fusion in the cervical spine. All three system spacers are manufactured from PEEK (ASTM F2026), with tantalum (ASTM F560) radiographic markers. The Complete Cervical spacers also include titanium alloy and screws (ASTM F136), while Zuma-C includes titanium screws, as well as plates and set screws (ASTM F136).

Each system spacer is generally box-shaped with a central canal for receiving autograft bone graft material and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone. The systems are implanted via an anterior approach.

The instruments included with each system facilitate the placement and adjustment of the interbody spacers, and removal if necessary. The instruments are placed in trays and caddies for storage, protection, and organization prior to and during the steam sterilization process.

# Intended Use/Indications for use

# Vu c·POD:

The Vu coPOD Intervertebral Body Fusion Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical degenerative disc disease (DDD) at one level from C2-C3 to C7-T1. Cervical degenerative disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

The Vu c•POD implants are to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and /or corticocancellous bone and implanted via an open, anterior approach. The cervical device is to be used in patients who have had six weeks of non-operative treatment. The Vu coPOD Intervertebral Body Fusion Device is intended for use with supplemental internal fixation systems.

## Complete Cervical:

The Complete Cervical Intervertebral Body Fusion Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

The Complete Cervical implants are to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and /or corticocancellous bone and implanted via an anterior approach. The cervical device is to be used in patients who have had six (6) weeks of non-operative treatment. The cervical device is to be used with two titanium alloy screws which accompany the implant.

# Zuma-C:

Zuma-C is a stand-alone anterior cervical interbody fusion device intended for use as an adjunct to fusion at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as discogenic neck pain with degeneration of the disc confirmed by

{7}------------------------------------------------

history and radiographic studies). Patients should have received at least six weeks of non-operative treatment prior to treatment with the device. Zuma-C is to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and /or corticocancellous bone and implanted via an open, anterior approach. Zuma-C is intended to be used with the bone screw fixation provided and requires no additional fixation.

## Summary of Technological Characteristics

The SeaSpine Vu c-POD, Complete Cervical, and Zuma-C Systems and the predicate devices have the same operational principle; they act as a disc spacer and hold bone graft to promote fusion in the cervical spine. The SeaSpine Vu c POD, Complete Cervical, and Zuma-C Systems are substantially equivalent to the cited predicate devices in areas including intended use/indications for use, technological characteristics (operating principle, design, materials, sterility, manufacturing, etc.) and performance (mechanical safety).

## Non-Clinical Testing

There are no changes to the design, materials, specifications or manufacture of the implants, therefore no mechanical testing was performed for the Vu c Pod and Complete Cervical systems.

Engineering analysis verified that the device modifications in the Zuma-C did not create any new worst cases with respect to mechanical performance. Conclusions from previously performed mechancial testing per ASTM F2077 and F2267 remain valid.

## Clinical Testing

Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.

## Conclusions

The submitted data demonstrate that the SeaSpine Vu c.POD. Complete Cervical, and Zuma-C Systems are substantially equivalent to the cited legally marketed predicate devices

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K172926](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K172926)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
