The Stryker Spine AERO®-C Cervical Cage is indicated for use in cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have six weeks of non-operative therapy. The AERO®-C Cervical Cage System is to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, and is to be implanted via an open, anterior approach. The AERO®-C Cervical Cage System is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine. In addition, the device must be used with the included fixation anchors.
Device Story
AERO®-C Cervical Cage System is an interbody fusion device for cervical spine surgery. Device consists of a hollow, ring-shaped PEEK Optima cage surrounded by a titanium alloy jacket. Implant features two internal pockets for bone graft containment and serrated superior/inferior surfaces for bone engagement. System includes titanium alloy fixation anchors that mate with dovetail channels in the cage to lock into the titanium jacket. Implanted via open, anterior approach by surgeons. Used with autogenous or allogenic bone graft and supplemental spinal fixation systems. Provides structural support to facilitate fusion in DDD patients. Benefits include stabilization of the spinal segment and promotion of interbody fusion.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F2077 (static/dynamic compression, compression shear, torsion), ASTM F04-25-02-02 (expulsion), and ASTM F2267 (subsidence). Additional testing included wear debris assessment, fixation anchor engagement evaluation, and dynamic cadaveric fatigue testing.
Technological Characteristics
Materials: PEEK Optima (ASTM F2026) and titanium alloy (ASTM F136, ISO 5832-3). Design: Hollow, ring-shaped cage with titanium jacket, serrated surfaces, and internal dovetail channels for fixation anchors. Dimensions: Variable heights, widths, depths, and lordotic angles. Sterilization: Not specified.
Indications for Use
Indicated for cervical interbody fusion in skeletally mature patients with degenerative disc disease (DDD) at one level (C2-C3 to C7-T1). Requires history of neck pain of discogenic origin, radiographic confirmation of disc degeneration, and six weeks of failed non-operative therapy.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Stryker Spine AERO®-AL Lumbar Cage System (K133328)
Stryker Spine AERO®-LL Lumbar Cage System (K142066)
Related Devices
K231763 — Stable-C Interbody System · Nexus Spine, LLC · Aug 21, 2023
K171489 — Acapella Cervical Spacer System · Choicespine, LP · Oct 23, 2017
K162760 — Spine Wave Anterior Cervical Spine System · Spine Wave, Inc. · Feb 16, 2017
K142066 — AERO-LL LUMBAR CAGE SYSTEM · Stryker Corporation · Nov 13, 2014
K181295 — Genesys Spine AIS-C Cervical Stand-Alone System · Genesys Spine · Aug 29, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol representing the human form, depicted as three stylized profiles facing to the right. The profiles are connected and flow together, creating a sense of unity and collaboration.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18, 2015
Stryker Spine Ms. Sorava King Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401
Re: K152532
Trade/Device Name: AERO®-C Cervical Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: November 23, 2015 Received: November 24, 2015
Dear Ms. King:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152532
Device Name AERO®-C Cervical Cage System
#### Indications for Use (Describe)
The Stryker Spine AERO®-C Cervical Cage is indicated for use in cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from the C2-C3 disc to the C7-Tl disc.
DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have six weeks of non-operative therapy.
The AERO®-C Cervical Cage System is to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, and is to be implanted via an open, anterior approach.
The AERO®-C Cervical Cage System is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine. In addition, the device must be used with the included fixation anchors.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Section 10: 510k Summary | | |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 510(k) Summary: AERO®- C Cervical Cage System | | |
| Submitter: | Stryker Spine<br>2 Pearl Court<br>Allendale, New Jersey 07401 | |
| Contact Person | Soraya King<br>Regulatory Affairs Specialist<br>Phone: 201-749-8296<br>Fax: 201-962-4296<br>Email: Soraya.King@Stryker.com | |
| Date Prepared | November 23, 2015 | |
| Trade Name | AERO®-C Cervical Cage System | |
| Proposed Class | Class II | |
| Classification Name<br>and Number | Intervertebral body fusion device, 21 CFR 888.3080 | |
| Product Code | ODP | |
| Predicate Devices | The AERO-C® Cervical Cage System was shown to be<br>substantially equivalent to the devices listed below: | |
| | Primary Predicate | |
| | • Stryker Spine AVS® AS PEEK Spacer, K142251 &<br>K120486 | |
| | Additional Predicates | |
| | • Stryker Spine AVS® Anchor-C Cervical Cage System,<br>K102606 | |
| | • ORTHOVITA PEEK Cages, K072981 | |
| | • Surgicraft STALIF C®, K150053 | |
| | • Synthes Zero-P, K072981 | |
| | • LDR Spine ROI-C System, K091088 | |
| | Reference Devices | |
| 510(k) Summary: AERO®- C Cervical Cage System | | |
| | Stryker Spine AERO® -AL Lumber Cage System, K133328 Stryker Spine AERO®-LL Lumbar Cage System, K142066 | |
| Device Description | The AERO®-C Cervical Cage is a hollow, ring-shaped PEEK<br>Optima (per ASTM F2026) cage surrounded by a titanium alloy<br>(per ASTM F136 and ISO 5832-3) jacket. It is intended for use<br>as an interbody fusion device and is offered in a variety of<br>heights, widths, depths, and lordotic angles to adapt to varying<br>patient anatomies. The PEEK Optima cage portion consists of<br>two closed pockets for graft containment and has serrations on<br>the superior and inferior surfaces of the cage. The implant is<br>designed to be used with the internal supplemental fixation<br>provided (AERO®-C Fixation Anchors). The AERO®-C<br>Fixation Anchors are constructed from titanium alloy and<br>possess rails that mate with dovetail channels located within the<br>AERO®-C Cervical Cage. Once fully seated into the channels,<br>the anchors are designed to lock into the titanium jacket. | |
| Indications for Use | The Stryker Spine AERO®-C Cervical Cage is indicated for use<br>in cervical interbody fusion procedures in skeletally mature<br>patients with degenerative disc disease (DDD) at one level from<br>the C2-C3 disc to the C7-T1 disc.<br><br>DDD is defined as neck pain of discogenic origin with<br>degeneration of the disc confirmed by history and radiographic<br>studies. These patients should be skeletally mature and have six<br>weeks of non-operative therapy.<br><br>The AERO®-C Cervical Cage System is to be used with<br>autogenous bone graft and/or allogenic bone graft comprised of<br>cancellous and/or corticocancellous bone graft, and is to be<br>implanted via an open, anterior approach. | |
| 510(k) Summary: AERO®- C Cervical Cage System | | |
| | The AERO®-C Cervical Cage System is intended to be used<br>with supplemental spinal fixation systems that have been cleared<br>for use in the cervical spine. In addition, the device must be used<br>with the included fixation anchors. | |
| Summary of the<br>Technological<br>Characteristics | The subject AERO-C® implant system and the predicates share<br>similar design features:<br>Graft windows for packing autogenous bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft Serrations on the superior and inferior surfaces Comparable heights, widths, depths, and lordotic angles Manufactured from PEEK and Titanium Alloy | |
| 510(k) Summary: AERO®- C Cervical Cage System | | |
| Summary of Non-Clinical Testing | Testing in compliance with FDA's June 12, 2007 "Class II<br>Special Controls Guidance Document: Intervertebral Body<br>Fusion Device" was performed for the AERO-C® implant<br>system and demonstrated substantially equivalent performance<br>to the identified predicate device systems.<br>The following mechanical tests were performed:<br>• Static Compression (per ASTM F2077)<br>• Dynamic Compression (per ASTM F2077)<br>• Static Compression Shear (per ASTM F2077)<br>• Dynamic Compression Shear (per ASTM F2077)<br>• Static Torsion (per ASTM F2077)<br>• Dynamic Torsion (per ASTM F2077)<br>• Expulsion (per ASTM F04-25-02-02 Draft)<br>• Subsidence (per ASTM F2267)<br>• Wear Debris Assessment<br>• Fixation Anchor Engagement Evaluation<br>• Dynamic Cadaveric Testing involving the implantation of the<br>AERO®-C device into multiple functional spinal units and<br>subsequent evaluation of fatigue performance under multiple<br>loading modes. | |
| Conclusion | Based upon a comparison of the design features, indications for<br>use, technological characteristics, the use of established well-<br>known materials, and mechanical performance, the AERO®-C<br>Cervical Cage System has demonstrated substantial equivalence<br>to the identified predicate device systems. | |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.