← Product Code [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP) · K150521

# Interbody Fusion (IBF)/Vertebral Body Replacement(VBR) System (K150521)

_Rti Surgical, Inc. Dba Rti Biologics · ODP · May 15, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K150521

## Device Facts

- **Applicant:** Rti Surgical, Inc. Dba Rti Biologics
- **Product Code:** [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP.md)
- **Decision Date:** May 15, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

When used as a cervical intervertebral body fusion device (C-Plus), the Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System ("IBF/VBR System") is indicated for intervertebral body fusion of the spine in skeletally mature patients. Cervical IBFs are intended for use at one level in the cervical spine, from the C2-C3 disc to the C7-T1 disc, for the treatment of cervical disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment. IBFs are designed for use with autogenous bone graft and/or allogenic bone graft comprised of cancellous, cortical, and/or corticocancellous bone graft to facilitate fusion. IBFs are intended to be used with supplemental spinal fixation cleared for the implanted level, such as Streamline OCT, SlimFuse, Cequence, PAC, or Aspect Systems. When used as a lumbar intervertebral body fusion device (Rotate, Bullet-Tip, T-Plus, Contact, CrossFuse, and CrossFuse II), the Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System ("IBF/VBR System") is indicated for intervertebral body fusion of the spine in skeletally mature patients. Lumbar IBFs are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Lumbar IBFs are to be used in patients who have had six months of non-operative treatment. IBFs are designed for use with autogenous bone graft to facilitate fusion. IBFs are intended to be used with supplemental spinal fixation cleared for the implanted level, such as Quantum, Streamline TL, Contact ALP, Streamline MIS Systems, or Lat-Fuse Lateral Plate System. When used as a vertebral body replacement (VBR) device (C-Plus, Rotate, Bullet-Tip, T-Plus, Contact, CrossFuse, and CrossFuse II), the Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System ("IBF/VBR System") is intended for use in the thoracolumbar spine (TI-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. VBRs are also indicated for treating fractures of the thoracic and lumbar spine. VBRs are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period of time. The system must be used with supplemental fixation cleared for the conditions listed above (i.e., tumor or trauma of TI-L5) such as the Streamline TL Spinal Fixation System. Streamline MIS Spinal Fixation System or Quantum Spinal Fixation System. Additionally, the VBR device is intended to be used with bone graft.

## Device Story

System comprises implantable PEEK devices with tantalum or titanium alloy radiographic markers; available in various shapes/sizes for patient anatomy. Used in cervical, lumbar, and thoracolumbar spine procedures via open or minimally invasive approaches. Implants function as interbody fusion cages or vertebral body replacements to restore spinal height and biomechanical integrity; require supplemental spinal fixation and bone graft (autogenous or allogenic). Operated by surgeons in clinical/OR settings. Output is structural support for spinal column; facilitates fusion or provides stability post-vertebrectomy. Benefits include decompression of neural tissues and restoration of spinal column integrity.

## Clinical Evidence

Literature analysis of published clinical data for cervical interbody fusion devices similar to the predicate (C-Plus). Outcomes demonstrate that use of allogenic bone graft in anterior cervical interbody fusion for cervical disc disease poses no new risks.

## Technological Characteristics

Materials: PEEK with tantalum or titanium alloy radiographic markers. Form factor: Various shapes/sizes for spinal anatomy. Connectivity: None. Sterilization: Not specified. Software: None.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Pioneer IBF/VBR System ([K133455](/device/K133455.md))
- Pioneer IBF/VBR System ([K043206](/device/K043206.md))
- Pioneer IBF/VBR System ([K061151](/device/K061151.md))
- Pioneer IBF/VBR System ([K073177](/device/K073177.md))
- Pioneer IBF/VBR System ([K112496](/device/K112496.md))
- Pioneer IBF/VBR System ([K133623](/device/K133623.md))
- NuVasive CoRoent Small Contoured Interbody System ([K142050](/device/K142050.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, overlapping each other to create a sense of depth and unity. The faces are simple and abstract, with no distinct features other than the outline of the head and neck.

May 15, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

RTI Surgical, Incorporated dba RTI Biologics Ms. Jennifer Bonacci Regulatory Affairs specialist 11621 Research Circle Alachua, Florida 32615

Re: K150521

Trade/Device Name: Interbody Fusion (IBF)/ Vertebral Body Replacement (VBR) System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP, MAX, MQP Dated: February 27, 2015 Received: March 2, 2015

Dear Ms. Bonacci:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K150521

#### Device Name

Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System

#### Indications for Use (Describe)

### CERVICAL INTERBODY FUSION DEVICE

When used as a cervical intervertebral body fusion device (C-Plus), the Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System ("IBF/VBR System") is indicated for intervertebral body fusion of the spine in skeletally mature patients. Cervical IBFs are intended for use at one level in the cervical spine, from the C2-C3 disc to the C7-T1 disc, for the treatment of cervical disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment. IBFs are designed for use with autogenous bone graft and/or allogenic bone graft comprised of cancellous, cortical, and/or corticocancellous bone graft to facilitate fusion. IBFs are intended to be used with supplemental spinal fixation cleared for the implanted level, such as Streamline OCT, SlimFuse, Cequence, PAC, or Aspect Systems.

| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K150521

#### Device Name

Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System

Indications for Use (Describe)

### LUMBAR INTERBODY FUSION DEVICE

When used as a lumbar intervertebral body fusion device (Rotate, Bullet-Tip, T-Plus, Contact, CrossFuse, and CrossFuse II), the Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System ("IBF/VBR System") is indicated for intervertebral body fusion of the spine in skeletally mature patients. Lumbar IBFs are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Lumbar IBFs are to be used in patients who have had six months of non-operative treatment. IBFs are designed for use with autogenous bone graft to facilitate fusion. IBFs are intended to be used with supplemental spinal fixation cleared for the implanted level, such as Quantum, Streamline TL, Contact ALP, Streamline MIS Systems, or Lat-Fuse Lateral Plate System.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K150521

#### Device Name

Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System

Indications for Use (Describe)

#### VERTEBRAL BODY REPLACEMENT

When used as a vertebral body replacement (VBR) device (C-Plus, Rotate, Bullet-Tip, T-Plus, Contact, CrossFuse, and CrossFuse II), the Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System ("IBF/VBR System") is intended for use in the thoracolumbar spine (TI-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. VBRs are also indicated for treating fractures of the thoracic and lumbar spine. VBRs are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period of time. The system must be used with supplemental fixation cleared for the conditions listed above (i.e., tumor or trauma of TI-L5) such as the Streamline TL Spinal Fixation System. Streamline MIS Spinal Fixation System or Quantum Spinal Fixation System. Additionally, the VBR device is intended to be used with bone graft.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{5}------------------------------------------------

## 510(k) Summary Pursuant to 21 CFR 807.92

| Sponsor:                            | Pioneer Surgical Technology, Inc. dba RTI Surgical, Inc.<br>375 River Park Circle<br>Marquette, MI 49855 USA<br>Contact: Jennifer Bonacci, Regulatory Affairs Specialist<br>Kristina Hall, Sr. Manager, Regulatory Affairs<br>Ph: (386) 418-8888<br>Fax: (386) 418-418-1627<br>Prepared: April 16, 2015                                                                                                          |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name:                               | Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System                                                                                                                                                                                                                                                                                                                                                 |
| Trade names:                        | C-Plus, Rotate, Bullet-Tip, T-Plus, Contact, CrossFuse, CrossFuse II,<br>CrossFuse II Coronal Taper, CrossFuse II Hyperlordotic                                                                                                                                                                                                                                                                                  |
| Common name:                        | Intervertebral Body Fusion Device/ Vertebral Body Replacement<br>Device                                                                                                                                                                                                                                                                                                                                          |
| Classifications:                    | 21 CFR 888.3080 – Class II<br>21 CFR 888.3060 – Class II                                                                                                                                                                                                                                                                                                                                                         |
| Product Codes:                      | MAX, ODP, MQP                                                                                                                                                                                                                                                                                                                                                                                                    |
| Panel/ Branch:                      | Orthopedic and Rehabilitation Devices Panel; Panel Code 87<br>Restorative Devices Branch                                                                                                                                                                                                                                                                                                                         |
| Primary Predicate<br>Device:        | Pioneer IBF/VBR System K133455                                                                                                                                                                                                                                                                                                                                                                                   |
| Additional<br>Predicate<br>Devices: | Pioneer IBF/VBR System (K043206, K061151, K073177, K112496, and<br>K133623)<br>NuVasive® CoRoent® Small Contoured Interbody System (K142050)                                                                                                                                                                                                                                                                     |
| Description:                        | The system includes implantable devices manufactured from PEEK with<br>tantalum or titanium alloy radiographic markers that are available in a<br>variety of different shapes and sizes to accommodate varying patient<br>anatomy and surgical approach. The IBF/VBR implants may be<br>implanted via a variety or open or minimally invasive approaches,<br>including anterior, lateral, posterior and oblique. |

{6}------------------------------------------------

include the use of allogenic bone graft comprised of cancellous, cortical, and/or corticocancellous bone graft as an alternative to autogenous bone graft for use with the predicate C-Plus device in cervical intervertebral body fusion. No changes were made to the C-Plus implants.

Class I and 510k exempt orthopedic manual surgical instruments are also available for use with the System.

#### CERVICAL INTERBODY FUSION DEVICE Indications for

Use:

When used as a cervical intervertebral body fusion device (C-Plus), the Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System ("IBF/VBR System") is indicated for intervertebral body fusion of the spine in skeletally mature patients. Cervical IBFs are intended for use at one level in the cervical spine, from the C2-C3 disc to the C7-T1 disc, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment. IBFs are designed for use with autogenous bone graft and/or allogenic bone graft comprised of cancellous, cortical, and/or corticocancellous bone graft to facilitate fusion. IBFs are intended to be used with supplemental spinal fixation cleared for the implanted level, such as Streamline OCT, SlimFuse, Cequence, PAC, or Aspect Systems.

## LUMBAR INTERBODY FUSION DEVICE

When used as a lumbar intervertebral body fusion device (Rotate. Bullet-Tip, T-Plus, Contact, CrossFuse, and CrossFuse II), the Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System ("IBF/VBR System") is indicated for intervertebral body fusion of the spine in skeletally mature patients. Lumbar IBFs are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Lumbar IBFs are to be used in patients who have had six months of non-operative treatment. IBFs are designed for use with autogenous bone graft to facilitate fusion. IBFs are intended to be used with supplemental spinal fixation cleared for the implanted level, such as Quantum, Streamline TL, Contact ALP, Streamline MIS Systems, or Lat-Fuse Lateral Plate System.

## VERTEBRAL BODY REPLACEMENT

When used as a vertebral body replacement (VBR) device (C-Plus, Rotate, Bullet-Tip, T-Plus, Contact, CrossFuse, and CrossFuse II), the Interbody Fusion (IBF) / Vertebral Body Replacement (VBR) System ("IBF/VBR

{7}------------------------------------------------

System") is intended for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. VBRs are also indicated for treating fractures of the thoracic and lumbar spine. VBRs are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period of time. The system must be used with supplemental fixation cleared for the conditions listed above (i.e., tumor or trauma of T1-L5) such as the Streamline TL Spinal Fixation System, Streamline MIS Spinal Fixation System or Quantum Spinal Fixation System. Additionally, the VBR device is intended to be used with bone graft.

- Summary of The purpose of this 510(k) submission is to seek clearance for the use of Technological allogenic bone graft comprised of cancellous, cortical, and/or Characteristics: corticocancellous bone graft as an alternative to autogenous bone graft for cervical interbody fusion devices previously cleared via K133455. No changes have been made to the actual implants.
Performance Data A literature analysis of published clinical data for the cervical interbody Supporting fusion devices similar to the predicate device, C-Plus (K133455), was Substantial provided in support of the expanded Indications for Use. The published Equivalence clinical outcomes demonstrated that the use of allogenic bone graft Determination: comprised of cancellous, cortical, and/or corticocancellous bone graft, in anterior cervical interbody fusion procedures to treat patients diagnosed with cervical disc disease as defined above poses no new risks to patients. No changes were made to the existing devices, nor were any new components added to the system. Therefore, no additional testing was required or performed.

- The intended use, design features, materials used in manufacturing and Conclusion: sterilization methods are substantially equivalent to the previously cleared, predicate C-Plus device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K150521](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K150521)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
