← Product Code [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP) · K113559

# LDR SPINE CERVICAL INTERBODY FUSION SYSTME-ROI-C LORDOTIC IMPLANTS (K113559)

_Ldr Spine USA, Inc. · ODP · Dec 29, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K113559

## Device Facts

- **Applicant:** Ldr Spine USA, Inc.
- **Product Code:** [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP.md)
- **Decision Date:** Dec 29, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The LDR Spine Cervical Interbody Fusion System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The LDR Spine ROI-C Cervical Interbody Fusion System is to be used with autogenous bone graft and implanted via an open, anterior approach. Supplemental internal fixation is required to properly utilize this system.

## Device Story

ROI-C Lordotic implant is an interbody fusion device for the cervical spine; extension of previously cleared ROI-C system. Device implanted via open anterior approach by surgeons to treat DDD. Implant acts as a spacer to facilitate fusion; requires autogenous bone graft and supplemental internal fixation. PEEK OPTIMA LT1 material provides structural support; tantalum markers allow radiographic visualization. Device performance validated via static axial compression and torsion testing.

## Clinical Evidence

Bench testing only. Finite Element Analysis performed to simulate static axial compression and static axial torsion per ASTM F2077 standards.

## Technological Characteristics

Materials: PEEK OPTIMA LT1 (ASTM F2026) and tantalum markers (ASTM F136). Form factor: Lordotic interbody fusion implant. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- LDR Spine Cervical Interbody Fusion System ([K091088](/device/K091088.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness
LDR Spine USA Cervical Interbody Fusion System

Owner's Name & Address:

Contact Person:

Date:

Common Name:

Classification:

Proprietary Name:

Legally Marketed
Predicate Device:

Device Description

LDR Spine USA
13785 Research Boulevard, Suite 200
Austin, TX 78750
Phone: (512) 344-3333
Fax: (512) 344-3350

Kiersten Soderman
Regulatory Affairs Specialist

LDR Spine USA
13785 Research Boulevard, Suite 200
Austin, TX 78750
Phone: (512) 344-3370
Fax: (512) 344-3350
Email: kierstensoderman@Idrspine.com

November 29, 2011

Intervertebral Fusion Device with Integrated Fixation,
Cervical (OVE)
Intervertebral Fusion Device with Bone Graft, Cervical
(ODP)

OVE, ODP-21 CFR 888.3080- Intervertebral Body
Fusion Device

LDR Spine Cervical Interbody Fusion System- ROI-C
Lordotic Implants

LDR Spine Cervical Interbody Fusion System
K091088

The ROI-C Lordotic implant is an extension of the
currently cleared ROI-C cervical interbody fusion
system and is intended for use as an interbody fusion
device in the cervical spine. The device is
manufactured from medical grade PEEK OPTIMA®
LT1 in accordance with ASTM F2026 and has
tantalum markers conforming to ASTM F136

K113559
Page 1 of 2

DEC 2 9 2011

153

{1}------------------------------------------------

K113559
Page 2 of 2

## 510(k) Summary of Safety and Effectiveness

LDR Spine USA Cervical Interbody Fusion System

embedded in the implant extremities to facilitate visibility in x-ray imaging. The subject device is designed for placement using an open anterior approach.

The LDR Spine Cervical Interbody Fusion System is Indications for Use: indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The LDR Spine ROI-C Cervical Interbody Fusion System is to be used with autogenous bone graft and implanted via an open, anterior approach. Supplemental internal fixation is required to properly utilize this system.

Non-Clinical

Testing included Finite Element Analysis simulating static axial compression and static axial torsion (per ASTM F2077). The results of this testing demonstrate that the performance of the ROI-C Lordotic implant, when compared with its legally marketed predicates, is substantially equivalent.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC 2 9 2011

LDR Spine USA % Ms. Kiersten Soderman Regulatory Affairs Specialist 13785 Research Boulevard, Suite 200 Austin, Texas 78750

Re: K113559

Trade/Device Name: LDR Spine Cervical Fusion System - ROI-C Lordotic Implants Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP, OVE Dated: November 30, 2011 Received: December 01, 2011

Dear Ms. Soderman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{3}------------------------------------------------

Page 2 - Ms. Kiersten Soderman

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdagov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark A. Millhoss

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE

510(k) Number (if known):

Device Name: LDR Spine Cervical Interbody Fusion System- ROI-C Lordotic Implants

Indications for Use:

The LDR Spine Cervical Interbody Fusion System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The LDR Spine ROI-C Cervical Interbody Fusion System is to be used with autogenous bone graft and implanted via an open, anterior approach. Supplemental internal fixation is required to properly utilize this system.

Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

4

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Viginial Catter

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K113559

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K113559](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K113559)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
