ARENA-C CERVICAL INTERVERTEBRAL BODY FUSION SYSTEM

K113518 · Spinefrontier, Inc. · ODP · Feb 16, 2012 · Orthopedic

Device Facts

Record IDK113518
Device NameARENA-C CERVICAL INTERVERTEBRAL BODY FUSION SYSTEM
ApplicantSpinefrontier, Inc.
Product CodeODP · Orthopedic
Decision DateFeb 16, 2012
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Arena-C™ Cervical Intervertebral Body Fusion Device is a spinal intervertebral body fusion device intended for intervertebral body fusion of the spine of skeletally mature patients, using autogenous bone graft to facilitate fusion. The device is indicated for use in patients with degenerative disc disease (DDD) of the cervical spine at one disc level from the C2-C3 disc to the C7-T1 disc. The SpineFrontier Arena-C™ Cervical Intervertebral Body Fusion Device is intended to be used with supplemental spinal fixation system(s) cleared for use in the cervical spine (example: Anterior Cervical Plate Fixation). Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.

Device Story

Arena-C is a cervical intervertebral body fusion device; constructed of PEEK Optima; designed for implantation into the cervical spine disc space. Device features raised contours on superior/inferior surfaces to resist migration. Implanted by surgeons during spinal fusion procedures; requires use of autogenous bone graft to facilitate fusion; must be used in conjunction with supplemental spinal fixation (e.g., anterior cervical plate). Device provides structural support to the intervertebral space to promote fusion in patients with discogenic pain and degenerative disc disease.

Clinical Evidence

Bench testing only. Mechanical performance assessments conducted per ASTM F2077-03 (static/dynamic axial compression, torsion, shear), ASTM F2267-04 (subsidence), and ASTM Draft Standard F-04.25.02.02 (static expulsion). No clinical data required.

Technological Characteristics

Material: PEEK Optima. Form factor: Cervical intervertebral cage with raised surface contours. Mechanical testing per ASTM F2077-03, ASTM F2267-04, and ASTM Draft Standard F-04.25.02.02.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one level (C2-C3 to C7-T1) who have failed 6 weeks of non-operative treatment. Must be used with supplemental cervical spinal fixation.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 113518 # FFEB 1 6 2012 #### 5.0 510(k) Summary ## 1. Sponsor SpineFrontier, Inc. 500 Cummings Center Suite 3500 Beverly, MA 01915 | Primary Contact: | Fredy H. Varela, RAC | |------------------|----------------------| | Telephone: | 1- 978-232-3990 | November 23, 2011 Date Prepared: ## 2. Device Name and Classification: | Proprietary Name: | Arena-C™ Cervical Intervertebral Body Fusion<br>System, Arena-C™ Intervertebral Body Fusion<br>Device System, Arena-C™ IBFD, Arena-C™<br>IBF, Arena-C™ | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Intervertebral Fusion Device With Bone Graft,<br>Cervical | | Classification Name: | Intervertebral Fusion Device With Bone Graft,<br>Cervical, (21 CFR 888.3080), Class II | | Product Code: | ODP | #### 3. Predicate Devices K090064 Eminent Spine Interbody Fusion System #### 4. Device Description The SpineFrontier Cervical Interbody Fusion Device System (Arena-C™ Cervical Intervertebral) is a spinal intervertebral body fusion device intended for intervertebral body fusion of the spine of skeletally mature patients, using autogenous bone graft to facilitate fusion. The device is indicated for use in patients with degenerative disc disease (DDD) of the cervical spine at one disc level from the C2-C3 disc to the C7-T1 disc. The system is comprised of devices made of Peek Optima®, with various heights to fit the anatomical needs of a wide variety of patients. The device has raised contours on the superior and inferior surfaces that will resist the device movement following implantation. Pg 1 of 2 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a handwritten string of characters. The characters appear to be "K113518". The handwriting is somewhat stylized, but the characters are still legible. The image is simple and only contains the text. #### 5. Intended Use The Arena-C™ Cervical Intervertebral Body Fusion Device is a spinal intervertebral body fusion device intended for intervertebral body fusion of the spine of skeletally mature patients, using autogenous bone graft to facilitate fusion. The device is indicated for use in patients with degenerative disc disease (DDD) of the cervical spine at one disc level from the C2-C3 disc to the C7-T1 disc. The SpineFrontier Arena-C™ Cervical Intervertebral Body Fusion Device is intended to be used with supplemental spinal fixation system(s) cleared for use in the cervical spine (example: Anterior Cervical Plate Fixation). Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage. #### 6. Technological Characteristics The SpineFrontier Arena-C™ Cervical Intervertebral Body Fusion Device was shown to be substantially equivalent to predicate devices through comparison of indications for use, function, operating principles, and materials. #### 7. Basis for Substantial Equivalence The SpineFrontier Arena-C™ Cervical Intervertebral Body Fusion Device was evaluated in accordance with FDA Document, Class II Special Controls, Guidance Document: Intervertebral Fusion Device, June 12, 2007, and has been found to meet criteria defined in the guidance document; and has been demonstrated to be substantially equivalent to predicate devices in terms of indications for use, function, materials, and performance (mechanical testing). Clinical data was not required for this device. Mechanical testing includes performance assessments per the following recognized test methods: - ASTM F2077-03, Static and Dynamic Axial Compression, Static and -Dynamic Torsion, and Static and Dynamic Shear Compression - ASTM F2267-04, Subsidence Under Static Axial Compression - - ASTM Draft Standard F-04.25.02.02, Static Expulsion - {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002 SpineFrontier, Inc. % Mr. Christopher Chang Director of Operations 500 Cummings Center, Suite 3500 Beverly, Massachusetts 01915 FEB 1 6 2012 Re: K113518 Trade/Device Name: Arena-C Cervical Intervertebral Body Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: II Product Code: ODP Dated: November 23, 2011 Received: November 29, 2011 Dear Mr. Chang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device; subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Mr. Christopher Chang CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 00016 K113518 #### 510(k) Number (if Known): Device Name: Arena-C Cervical Intervertebral Body Fusion System #### Indications For Use: The Arena-C Cervical Intervertebral Body Fusion System is a spinal intervertebral body fusion device intended for intervertebral body fusion of the spine of skeletally mature patients, using autogenous bone graft to facilitate fusion. The device is indicated for use in patients with degenerative disc disease (DDD) of the cervical spine at one disc level from the C2-C3 disc to the C7-T1 disc. The SpineFrontier Arena-C Cervical Intervertebral Body Fusion System is intended to be used with supplemental spinal fixation system(s) cleared for use in the cervical spine (example: Antieror Cervical Plate Fixation). Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage. Prescription Use: X OR Over-The-Counter Use: (Part 21 CFR 807.109) PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED #### Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices KIT3518 510(k) Number_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...