K103655 · Exactech, Inc. · ODP · Jun 30, 2011 · Orthopedic
Device Facts
Record ID
K103655
Device Name
OCTANE-C INTERBODY FUSION DEVICE SYSTEM
Applicant
Exactech, Inc.
Product Code
ODP · Orthopedic
Decision Date
Jun 30, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Octane-C Interbody Cervical devices are intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone. This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems). Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with this device.
Device Story
Octane-C Interbody Fusion System provides stability during cervical intervertebral body fusion; implanted via open anterior approach. Device consists of PEEK-Optima cage with 6% barium sulphate, titanium pins, and tantalum beads for radiographic visibility; 6° lordotic configuration. Used by surgeons in clinical settings; requires supplemental anterior plate fixation and autogenous bone graft. Benefits patient by facilitating spinal fusion in degenerative disc disease cases.
Clinical Evidence
Bench testing only. Mechanical performance validated per ASTM F2077 and ASTM F2267, including static and dynamic compression, torsion, and compressive shear, as well as subsidence and expulsion testing.
Indicated for skeletally mature patients with degenerative disc disease (discogenic neck pain confirmed by history/radiography) requiring single-level (C3-C7) anterior cervical intervertebral body fusion. Requires use with supplemental anterior spinal fixation and autogenous bone graft. Patients must have failed at least 6 weeks of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K180674 — EVOL® ha-C Cervical Interbody Fusion System · Cutting Edge Spine, LLC · Jun 25, 2018
K173077 — Cavetto-SA Cervical Cage System · Neurostructures, Inc. · Feb 23, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
KIO 3655 Bac
### Exactech® Octane®-C Interbody Fusion System Traditional 510(k)
## 510(k) Summary
| Company: | Exactech®, Inc<br>2320 Northwest 66th Court<br>Gainesville, FL 32653-1630 |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Date: | January 26, 2011 |
| Contact Person: | Patrick Hughes<br>Regulatory Affairs Specialist<br>Phone: 352-327 4762<br>Fax: 352-378-2617<br>E-mail: patrick.hughes@exac.com |
| Proprietary Name: | Exactech® Octane®-C Interbody Fusion System |
| Common Name: | Intervertebral body fusion system |
### Classification Name:
Intervertebral Fusion Device - Cervical (21 CFR 888.3080, Class II, Product Code ODP)
Legally Marketed Devices to Which Substantial Equivalence Is Claimed Phantom 1 Plus Cage System (#K082801) BAK Cervical Interbody Fusion Device (#P980048)
### Device Description
This submission proposes a new Cervical Interbody Fusion System. The Octane®-C Interbody Fusion System is designed to provide stability during intervertebral body fusion in the cervical spine. Implantable Octane-C devices incorporate a 6° lordotic configuration, are manufactured from PEEK-Optima with 6% barium sulphate, titanium pins, and tantalum beads added for radiographic visibility, are provided sterile, and are designed to be implanted using a direct anterior approach and autogenous bone graft material.
### Indications for Use
The Octane-C Interbody Cervical devices are intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone.
This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems).
{1}------------------------------------------------
## Exactech® Octane®-C Interbody Fusion System Traditional 510(k)
Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with this device.
### Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following characteristics:
#### . Intended Use
The proposed Octane-C Interbody Fusion System devices and cited predicates are intended to be used with autogenous bone graft and an anterior surgical approach for treatment of degenerative disc disease at the levels of C3-C7.
#### . Materials
Octane-C Interbody Fusion System devices and cited predicate devices are made from the same materials, with demonstrable histories of safe and effective use in medical applications, per internationally recognized consensus standards.
#### . Dimensions
Octane-C Interbody Fusion System devices and cited predicate devices have substantially equivalent dimensions and will be available in substantially equivalent size ranges as cited predicate devices.
#### Sterilization processes .
Octane-C Interbody Fusion System devices and cited predicate devices are provided sterile, and are sterilized using equivalent sterilization processes conforming to recognized industry standards.
#### . Performance specifications
Mechanical testing results summarized in this 510(k) premarket notification demonstrate Octane-C Interbody Fusion System devices have substantially equivalent performance characteristics compared to legally marketed predicates based on testing per ASTM F2077, ASTM F2267, and clinically relevant biomechanical loads:
- . Static Compression
- Static Torsion .
- Static Compressive Shear .
- Dynamic Compression .
- Dynamic Torsion .
- Dynamic Compressive Shear .
- . Subsidence
- Expulsion ●
## Substantial Equivalence Conclusion
The Octane-C Interbody Fusion System is substantially equivalent to cited predicates per intended use, materials, dimensions, sterilization processes, and performance characteristics.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the circumference. Inside the circle is a stylized symbol resembling an eagle or bird with three curved lines representing its wings and a wavy line below, possibly representing water or a tail.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 3 0 2011
Exactech®, Inc. % Mr. Patrick Hughes Regulatory Affairs Specialist 2320 Northwest 66th Court Gainesville. Florida 32653-1630
Re: K103655
Trade/Device Name: Exactech® Octane®-C Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: June 22, 2011 Received: June 23, 2011
Dear Mr. Hughes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Mr. Patrick Hughes
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Sincerely yours,
Mark N. Melkerson
Director
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Exactech® Octane@-C Interbody Fusion System Traditional 510(k)
# Indications for Use Statement
510(k) Number: 长103655
Device Name: Exactech® Octane@-C Interbody Fusion System
## INDICATIONS FOR USE:
The Octane-C Interbody Cervical devices are intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neek pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone.
This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems).
Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with this device.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
*(Division Sign-Off)*
Division of Surgical, Orthopedic, and Restorative Devices
5 10(k) Number________________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.