← Product Code [OBT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT) · K180624

# Guided Growth Plate System Plus (K180624)

_Orthofix Srl · OBT · May 15, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT/K180624

## Device Facts

- **Applicant:** Orthofix Srl
- **Product Code:** [OBT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT.md)
- **Decision Date:** May 15, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The Guided Growth Plate System Plus is intended for redirecting the growth of long bones. It is indicated for gradually correcting angular growth deformities in growing children. Specific conditions/diseases for which the device will be indicated include: valgus, varus or flexion, extension deformities of the knee (femur and/or tibia); valgus, varus, or plantar flexion deformities of the ankle; valgus or varus deformities of the elbow (humerus), and radial or ulnar deviation, flexion or extension deformities of the wrist (radius).

## Device Story

Guided Growth Plate System Plus is a metallic bone fixation system used to redirect long bone growth in pediatric patients. The system consists of bone plates and screws; the subject device introduces a 3.5mm thread diameter screw as a dimensional extension to the existing 4.5mm screw predicate. The device is implanted by orthopedic surgeons to treat angular deformities. By acting as a tension band, the system restricts growth on one side of the physis, allowing the opposite side to grow and gradually correct the deformity. The device is manufactured from Ti6Al4V. Clinical benefit includes non-invasive correction of skeletal deformities in growing children.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F564-17, ASTM F1264-16, and ASTM F543-17. Clinical equivalence supported by literature analysis comparing the plate and screw system to traditional staples and the master Orthofix pediatric growth control system (K031493).

## Technological Characteristics

Metallic bone fixation system (bone plate and screws). Material: Ti6Al4V. Standards: ASTM F564-17, ASTM F1264-16, ASTM F543-17. Design: 3.5mm thread diameter screw. Sterilization: Not specified (implied same as predicate).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Guided Growth Plate System Plus ([K172183](/device/K172183.md))
- Fixation staples ([K834513](/device/K834513.md))
- Pediloc Fragment System ([K140431](/device/K140431.md))

## Reference Devices

- Blount Staple Epiphyseal
- Guided Growth plate ([K031493](/device/K031493.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 15, 2018

Orthofix Srl Gianluca Ricadona Sr. Quality & Regulatory Affairs Manager Via delle Nazioni. 9 97012 Bussolengo (VR) Italy

Re: K180624

Trade/Device Name: Guided Growth Plate System Plus Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: OBT Dated: April 24, 2018 Received: April 25, 2018

Dear Gianluca Ricadona:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

## 510(k) Number (if known)

# K180624

Device Name Guided Growth Plate System Plus

#### Indications for Use (Describe)

The Guided Growth Plate System Plus is intended for redirecting the growth of long bones.

It is indicated for gradually correcting angular growth deformities in growing children. Specific conditions/diseases for which the device will be indicated include: valgus, varus or flexion, extension deformities of the knee (femur and/or tibia); valgus, varus, or plantar flexion deformities of the ankle; valgus or varus deformities of the elbow (humerus), and radial or ulnar deviation, flexion or extension deformities of the wrist (radius).

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{3}------------------------------------------------

| 510(k) Summary<br>(as required by 21 CFR 807.92)             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                      |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Submitter                                                    | Orthofix Srl                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                      |
|                                                              | Via delle Nazioni, 9                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                      |
|                                                              | 37012 Bussolengo (VR) - Italy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                      |
| Telephone                                                    | + 39 045 6719.000                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                      |
| Fax                                                          | + 39 045 6719.380                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                      |
| Contact Person                                               | Gianluca Ricadona                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                      |
|                                                              | Sr. Quality & Regulatory Affairs Manager                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                      |
| Address                                                      | Via delle Nazioni, 9                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                      |
|                                                              | 37012 Bussolengo (VR) - Italy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                      |
| Telephone                                                    | + 39 045 6719.000                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                      |
| Fax                                                          | + 39 045 6719.380                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                      |
| email                                                        | GianlucaRicadona@orthofix.it                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                      |
| Preparation date                                             | 5/3/2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                      |
| Trade Name                                                   | Guided Growth Plate System Plus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                      |
| Common Name                                                  | Bone plate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                      |
| Panel Code                                                   | 87 / Orthopaedic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                      |
| Classification Name                                          | plate, bone, growth control, pediatric, epiphysiodesis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                      |
| Class                                                        | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                      |
| Regulation Number                                            | 21 CFR 888.3030                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                      |
| Product Code                                                 | OBT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                      |
| Name of Predicate Device                                     | 510(k) #                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Manufacturer         |
| Guided Growth Plate System Plus                              | K172183                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Orthofix Srl         |
| Fixation staples                                             | K834513                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | STRYKER CORP.        |
| - Blount Staple Epiphyseal<br>(Reference device)             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                      |
| Pediloc Fragment System                                      | K140431                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | OrthoPediatrics Corp |
| Description                                                  | The Subject new 3.5mm thread diameter screw is a<br>gamma extension of the existing screws (thread<br>Ø4.5mm), components of the predicate Guided<br>Growth Plate System Plus (K172183).<br>The Subject device is a 3.5mm thread diameter screw,<br>available in three lengths and manufactured by the<br>same material (Ti6Al4V), as the 4.5mm diameter<br>gamma screws, included in the predicate Guided<br>Growth Plate System Plus (K172183).                                                                                                                                      |                      |
| Indications for use                                          | The Guided Growth Plate System Plus is intended for<br>redirecting the growth of long bones.<br>It is indicated for gradually correcting angular growth<br>deformities in growing children. Specific<br>conditions/diseases for which the device will be<br>indicated include: valgus, varus or flexion, extension<br>deformities of the knee (femur and/or<br>tibia); valgus, varus, or plantar flexion deformities of<br>the ankle; valgus or varus deformities of the elbow<br>(humerus), and radial or ulnar deviation, flexion or<br>extension deformities of the wrist (radius). |                      |
| Technological Characteristics and<br>Substantial Equivalence | Documentation was provided to demonstrate that the<br>Subject device is substantially equivalent to the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                      |
|                                                              | (K172183) component(s), in the following fundamental aspects:<br>indications for use, intended patient population, system composition, implant material, method of<br>fixation, packaging configuration, sterilization method, biocompatibility and MRI compatibility.<br>Respect to the predicate (K172183), the Subject device is a design dimensional specification change for                                                                                                                                                                                                      |                      |
|                                                              | the introduction of the new thread diameter 3.5mm screw.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                      |
| Performance data                                             | The design dimensional specification change of the Subject device has been managed by design control activity and Risk Management process. The potential identified risks have been measured and mitigated through mechanical testing and clinical literature analysis, in order to demonstrate that the Subject Device does not introduce additional risks respect to the predicate (K172183).                                                                                                                                                                                        |                      |
|                                                              | The testing activity results demonstrated to meet the established acceptance criteria.<br>The following standards have been followed to perform mechanical test on the Subject device:<br>• ASTM F564 - 17 "Standard Specification and Test Methods for Metallic Bone Staples".<br>• ASTM F1264 -16 "Standard Specifications and Test Methods for Intramedullary Fixation Devices".<br>• ASTM F543 - 17 "Standard Specification and Test Methods for Metallic Medical Bone Screws".                                                                                                    |                      |
|                                                              | The functional mechanical testing results are supported by the clinical literature analysis, conducted to demonstrate the clinical equivalence between the two alternative clinical procedure by:<br>- using the plate and screw system "Guided Growth plate" K031493 - master Orthofix System for pediatric growth control and epiphysiodesis treatment),<br>- using the staples, chosen as reference devices for the performance data.                                                                                                                                               |                      |
| Biocompatibility data                                        | The Subject device is equivalent in its final finished form to the predicate device, Guided Growth Plate System Plus (K172183), concerning the:<br>manufacturing, sterilization, processing, geometry, material, intended population, anatomical location and duration of exposure.<br>Therefore, no additional biocompatibility assessment                                                                                                                                                                                                                                            |                      |
|                                                              | was required for the Subject device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                      |
| Conclusion                                                   | The analysis on the performance data within this Premarket Notification supports the conclusion that the new Subject gamma size 3.5mm thread diameter screw is safe and effective as the predicate (K172183).                                                                                                                                                                                                                                                                                                                                                                          |                      |

# 510(k) Summary
Suired by 21 CER 807

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT/K180624](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT/K180624)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
