← Product Code [NQW](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW) · K243137

# Curiteva Porous PEEK Laminoplasty System (K243137)

_Curiteva, Inc. · NQW · Oct 28, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW/K243137

## Device Facts

- **Applicant:** Curiteva, Inc.
- **Product Code:** [NQW](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW.md)
- **Decision Date:** Oct 28, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Curiteva Porous PEEK Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Curiteva Porous PEEK Laminoplasty System is used to hold or buttess the allografi or autograft material in place in order to prevent the graft material from expulsion or impinging the spinal cord.

## Device Story

Internal spinal fixation system consisting of plates and screws; used in laminoplasty procedures (C3-T3). Plates feature preformed holes for bone screws; hinge plates available for stabilization. Implants feature solid and porous PEEK regions; surface-treated with ~20nm hydroxyapatite (HA) coating. Used by surgeons in clinical settings to buttress bone graft material, preventing expulsion or spinal cord impingement. Adaptable to patient anatomy via various sizes and configurations. Benefits include spinal stabilization and graft containment.

## Clinical Evidence

No clinical data. Bench testing not included as design changes do not present a new worst-case configuration.

## Technological Characteristics

Materials: Titanium alloy (ASTM F136) and PEEK (ASTM F2026). Implants feature solid and porous PEEK regions with ~20nm hydroxyapatite (HA) coating. Components: Plates and screws. Sterilization: Sterile, single-use.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Curiteva Laminoplasty System ([K231232](/device/K231232.md))

## Reference Devices

- Curiteva Porous PEEK Cervical Interbody Fusion System ([K213030](/device/K213030.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 28, 2024

Curiteva, Inc. Eric Linder Chief Technology Officer 25127 Will McComb Drive Tanner, Alabama 35671

Re: K243137

Trade/Device Name: Curiteva Porous PEEK Laminoplasty System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal Interlaminal Fixation Orthosis Regulatory Class: Class II Product Code: NOW Dated: September 30, 2024 Received: September 30, 2024

Dear Eric Linder:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Ethan R. Naylor -S

for Ronald P. Jean, Ph.D. Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K243137

#### Device Name

Curiteva Porous PEEK Laminoplasty System

#### Indications for Use (Describe)

The Curiteva Porous PEEK Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Curiteva Porous PEEK Laminoplasty System is used to hold or buttess the allografi or autograft material in place in order to prevent the graft material from expulsion or impinging the spinal cord.

| Type of Use (Select one or both, as applicable)                                  |                                                                      |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

## A. Submitter Information

| Submitter:      | Curiteva, Inc.<br>25127 Will McComb Drive<br>Tanner, AL 35671<br>Phone: (256) 213-1057<br>Fax: (256) 213-1058 |
|-----------------|---------------------------------------------------------------------------------------------------------------|
| Contact Person: | Eric Linder<br>regulatory@curiteva.com                                                                        |
| Date Prepared:  | September 30th, 2024                                                                                          |

#### B. Device Information

| Trade Name:            | Curiteva Porous PEEK Laminoplasty System                                                                                      |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Common Name:           | Spinal Interlaminal Fixation Orthosis                                                                                         |
| Classification Name:   | Orthrosis, Spine, Plate, Laminoplasty, Metal                                                                                  |
| Device Classification: | Class II (per 21 CFR 888.3050)                                                                                                |
| Product Code:          | NQW                                                                                                                           |
| Classification Panel:  | Division of Orthopedic Devices                                                                                                |
| Predicate Device(s):   | Primary: Curiteva Laminoplasty System (K231232)<br>Reference: Curiteva Porous PEEK Cervical Interbody Fusion System (K213030) |

## C. Device Description

The Curiteva Porous PEEK Laminoplasty System is an internal fixation device for spinal surgery that consists of various configurations of plates and screws. The implant configurations are available in different types and sizes so that adaptations can be made to take into account pathology and individual patient anatomy. The plates come preformed with holes to receive bone screws. Screws are used to attach the plates to bone. System plate configurations may be used with allograft or autograft material. A hinge plate is provided when additional stabilization is necessary.

All system components are manufactured from Titanium alloy (Ti-6A1-4V ELI) that conforms to ASTM F136, or PEEK (Polyetheretherketone) as described by ASTM F2026.

{5}------------------------------------------------

The Curiteva Porous PEEK Laminoplasty implants are sterile, single-use devices and available in a variety of different sizes to accommodate the individual pathology and anatomical conditions of the patient. The implants have solid and porous PEEK regions.

The Curiteva Porous PEEK Laminoplasty implants are manufactured from implant-grade PEEK (per ASTM F2026). Each porous PEEK implant has been surface treated with a hydroxyapatite (HA) coating that is approximately 20nm thick.

#### D. Indications for Use

The Curiteva Porous PEEK Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Curiteva Porous PEEK Laminoplasty System is used to hold or buttress the allograft material in place in order to prevent the graft material from expulsion or impinging the spinal cord.

### E. Comparison of Technological Characteristics

As was established in this submission, the subject Curiteva Porous PEEK Laminoplasty System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and to have similar technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and range of sizes.

### F. Performance Data

Performance data is not included in this submission. The design changes being included for review do not present a new worst-case configuration that would require additional performance testing.

## G. Conclusion

Based on the indications for use, technological characteristics, non-clinical performance testing, and comparison to predicate devices, the subject Curiteva Porous PEEK Laminoplasty System has been shown to be substantially equivalent to legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW/K243137](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW/K243137)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
