K181717 · Life Spine, Inc. · NQW · Oct 12, 2018 · Orthopedic
Device Facts
Record ID
K181717
Device Name
Life Spine Laminoplasty System
Applicant
Life Spine, Inc.
Product Code
NQW · Orthopedic
Decision Date
Oct 12, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Life Spine Laminoplasty System is indicated for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Life Spine Laminoplasty System holds or buttresses the allograft in place in order to prevent the allograft from expulsion or impinging on the spinal cord.
Device Story
Life Spine Laminoplasty System consists of titanium alloy plates and screws; used to support laminoplasty procedures in lower cervical/upper thoracic spine (C3-T3). Device functions as mechanical fixation to hold/buttress allograft; prevents allograft expulsion or spinal cord impingement. Intended for single-use only; operated by surgeons in clinical/OR settings. No software or automated processing involved.
Clinical Evidence
Bench testing only. Mechanical performance demonstrated via static and dynamic four-point bending tests (ASTM F2193-02) and static screw pull-out testing (ASTM F543).
Technological Characteristics
Titanium alloy (Ti-6AL-4V-ELI) per ASTM F-136. Components include plates and screws of various diameters/lengths. Mechanical fixation device; non-powered; standalone. Sterilization method not specified.
Indications for Use
Indicated for patients undergoing laminoplasty in the lower cervical and upper thoracic spine (C3-T3) to secure allograft and prevent expulsion or spinal cord impingement.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
K091623 — LAMINOPLASTY FIXATION SYSTEM · Nuvasive, Inc. · Oct 9, 2009
K150469 — Integra Laminoplasty System · Seaspine, Inc. · Mar 23, 2015
K191927 — Hinged Laminoplasty System · Life Spine, Inc. · Sep 20, 2019
K232471 — Vy Spine VyLam Laminoplasty System · Vy Spine, LLC · Oct 10, 2023
K160114 — Xspan Laminoplasty Fixation System · X-Spine Systems, Inc. · Mar 15, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The first logo on the left is the Department of Health & Human Services-USA logo. The second logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
Life Spine Inc. Angela Batker RA/QA Specialist 13951 S Quality Drive Huntley, Illinois 60142
Re: K181717
Trade/Device Name: Life Spine Laminoplasty System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal Interlaminal Fixation Orthosis Regulatory Class: Class II Product Code: NQW Dated: August 28, 2018 Received: August 30, 2018
Dear Angela Batker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
October 12, 2018
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Brent Showalter -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181717
Device Name Life Spine Laminoplasty System
#### Indications for Use (Describe)
The Life Spine Laminoplasty System is indicated for use in the lower thoracic spine (C3-T3) after a laminoplasty has been performed. The Life Spine Laminoplasty System holds or buttresses the allograft in place in order to prevent the allograft from expulsion or impinging on the spinal cord.
Type of Use (Select one or both, as applicable):
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary Life Spine Laminoplasty System
| Submitted By: | Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Angela Batker<br>Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |
| Date Prepared: | June 18, 2018 |
| Trade Name: | Life Spine Laminoplasty System |
| Common Name: | Spinal Interlaminal Fixation Orthosis |
| Classification: | NQW, CFR 888.3050, Class II |
| Primary Predicate: | Medtronic Centerpiece (K050082) |
| Secondary Predicate: | Stryker Escalate (K113802)<br>Synthes Arch (K032534) |
#### Device Description:
The Life Spine Laminoplasty System consists of plates screws in various diameters and lengths to support a laminoplasty procedure. All components are fabricated and manufactured from titanium alloy 6AL-4V-ELI per ASTM F-136.
All implants are intended for single use only and should not be reused under any circumstances. Do not use any of the Life Spine Laminoplasty System components with components from any other system or manufacturer. The Life Spine Laminoplasty System components should never be reused under any circumstances.
{4}------------------------------------------------
#### Intended Use of the Device:
The Life Spine Laminoplasty System is indicated for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Life Spine Laminoplasty System holds or buttresses the allograft in place in order to prevent the allograft from expulsion or impinging on the spinal cord.
#### Technological Characteristics:
The Life Spine Laminoplasty System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing.
## Material:
This submission seeks clearance of a device made from implant grade (Ti-6AL-4V) Titanium alloy according to F136. This this is the same material used in the predicate devices.
## Performance Data:
Testing according to ASTM F2193-02 included Static and Dynamic Four Point Bending Test & Static Screw Pull-Out testing to ASTM F543 was presented to demonstrate the substantial equivalency of the Medtronic Centerpiece (K050082).
#### Substantial Equivalence:
The Life Spine Laminoplasty System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance.
## Conclusion:
The information presented demonstrates the substantial equivalency of The Life Spine Laminoplasty System.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.