← Product Code [NPJ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NPJ) · K042896

# SMITH & NEPHEW HYBRID KNEE FEMORAL (K042896)

_Smith & Nephew, Inc. · NPJ · Jan 12, 2005 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NPJ/K042896

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [NPJ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NPJ.md)
- **Decision Date:** Jan 12, 2005
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Hybrid Knee Femoral Components are intended to be used for those patients whereby conditions exist that can not be solely addressed by a device that treats a single compartment (i.e. unicondylar or patellofemoral prosthesis) of the knee. Indications include: post-traumatic arthritis; degenerative arthritis; and failed osteotomies, hemiarthroplasties; and unicompartmental replacement. These indications will be used for the Hybrid Knee Femoral Components, whereby the medial condyle and patellofemoral regions have been affected by one or more of these conditions. The Hybrid Knee Femoral Components are single use only and are intended for implantation only with bone cement.

## Device Story

Hybrid Knee Femoral Components are orthopedic implants designed to replace medial condyle and patellofemoral regions of the knee joint; intended for patients requiring multi-compartment treatment where single-compartment devices are insufficient. Device is implanted by orthopedic surgeons in a clinical/surgical setting using bone cement. Components integrate with existing Genesis I Unicondylar and Genesis II Total Knee System tibial and patellar components. Mechanical testing confirms ability to withstand in vivo loading. Benefit is restoration of joint function in patients with degenerative or post-traumatic arthritis and failed prior interventions.

## Clinical Evidence

Bench testing only. Mechanical test data indicated the components are equivalent to devices currently used clinically and are capable of withstanding expected in vivo loading without failure.

## Technological Characteristics

Semi-constrained cemented prosthesis; metal/polymer construction. Designed for use with bone cement. Components are modular, intended for integration with Genesis I and Genesis II knee systems.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Genesis I Unicondylar Knee System
- Genesis II Total Knee System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Summary of Safety and Effectiveness

JAN 1 2 2005

Smith & Nephew, Inc.

# Smith & Nephew Hybrid Knee Femoral Components

### Contact Person and Address

Kim Kelly, Project Manager, Clinical and Regulatory Affairs Smith & Nephew, Inc., Orthopaedic Division 1450 East Brooks Road Memphis, TN 38116 (901) 399-6566

## Device Description

The Smith & Nephew Hybrid Knee Femoral Components are designed for use with tibial components of the Genesis I Unicondylar Knee System and patellar components of the Genesis II Total Knee System. The components are used to replace the medial condyle and patellofemoral regions of a femoral knee joint.

#### Device Classification Name

Knee joint femorotibial metal/polymer semi-constrained cemented prosthesis, 21 CFR 888.3530

Knee joint patellofemoral polymer/metal semi-constrained cemented prosthesis, 21 CFR 888.3540

#### Indications for Use

The Hybrid Knee Femoral Components are intended to be used for those patients whereby conditions exist that can not be solely addressed by a device that treats a single compartment (i.e. unicondylar or patellofemoral prosthesis) of the knee.

Indications include:

- post-traumatic arthritis; .
- degenerative arthritis; and ●
- failed osteotomies, hemiarthroplasties; and unicompartmental replacement .

These indications will be used for the Hybrid Knee Femoral Components, whereby the medial condyle and patellofemoral regions have been affected by one or more of these conditions.

The Hybrid Knee Femoral Components are single use only and are intended for implantation only with bone cement.

page 1 of 2

{1}------------------------------------------------

#### Mechanical and Clinical Data

A review of the mechanical test data indicated that the Smith & Nephew Hybrid Knee Femoral Components are equivalent to devices currently used clinically and are capable of withstanding expected in vivo loading without failure.

# Substantial Equivalence Information

The substantial equivalence of the Smith & Nephew Hybrid Knee Femoral Components is substantiated by its similarities in design features, overall indications, and material composition as existing components of the Genesis I Unicondylar Knee System and Genesis II Total Knee Systems distributed by Smith & Nephew, Inc., as well as other commercially available patellofemoral joint prostheses.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized symbol that resembles an abstract caduceus, which is a traditional symbol of medicine.

JAN 1 2 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Kim P. Kelly Project Manager Clinical and Regulatory Affairs Smith & Nephew, Inc., Orthopaedic Division 1450 E. Brooks Road Memphis, Tennessee 38116

Re: K042896

Trade/Device Name: Hybrid Knee Femoral Components Regulation Number: 21 CFR 888.3530 Regulation Name: Knee joint femorotibial metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: NPJ Dated: October 19, 2004 Received: October 20, 2004

Dear Ms. Kelly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mrs. Kelly

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark A. Milleros

Celia M. Witten, Ph.D., M.D.

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K042896

Device Name: Hybrid Knee Femoral Components

Indications for Use:

The Hybrid Knee Femoral Components are intended to be used for those patients whereby conditions exist that can not be solely addressed by a device that treats a single compartment (i.e. unicondylar or patellofemoral prosthesis) of the knee.

Indications include:

- post-traumatic arthritis; .
- degenerative arthritis; and .
- failed osteotomies, hemiarthroplasties; and unicompartmental . replacement

These indications will be used for the Hybrid Knee Femoral Components. whereby the medial condyle and patellofemoral regions have been affected by one or more of these conditions.

The Hybrid Knee Femoral Components are single use only and are intended for implantation only with bone cement

Prescription × AND/OR Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

I. Restorative. and Neurologic

**Page 1 of** 1

**510(k) Number**

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NPJ/K042896](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NPJ/K042896)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
