← Product Code [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG) · K211993

# Z-LINKPC System (K211993)

_Zavation Medical Products, LLC · NKG · Aug 11, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K211993

## Device Facts

- **Applicant:** Zavation Medical Products, LLC
- **Product Code:** [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG.md)
- **Decision Date:** Aug 11, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3075
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Z-LINKpc System implants are intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1-C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations: instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. These implants are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. To achieve additional levels of fixation, the Z-LINKsc System may be connected to the Zavation Spinal System using rod connectors and tapered rods. When used with the occipital plate, the Z-LINKpc System is also intended to provide immobilization and stabilization for the occipito-cervico-thoracic juction (occiput-T3) in treatment of the instabilities mentioned above, including occipitocervical dislocation.

## Device Story

Z-LINKpc System is a posterior cervical/thoracic spinal fixation system; provides immobilization and stabilization as an adjunct to fusion or for palliative tumor stabilization. Components include polyaxial screws, hooks, rods, cross-connectors, rod connectors, offsets, and occipital plates. Used by surgeons in clinical settings to treat spinal instabilities, fractures, and degenerative conditions. System components are implanted to restore spinal column integrity; can be connected to Zavation Spinal System for extended fixation. Benefits include stabilization of spinal segments and restoration of structural integrity in patients with limited life expectancy. Device is non-sterile and single-use.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated via static compression bending, static torsion, dynamic compression bending, and dynamic torsion testing per ASTM F2706 on worst-case constructs. Results demonstrate performance equivalent or superior to predicate devices.

## Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V ELI, ASTM F136) and cobalt-chromium alloy (Co-28Cr-6Mo, ASTM F1537). Design: Rod-based posterior spinal fixation system including polyaxial screws, hooks, rods, and occipital plates. Non-sterile, single-use components.

## Regulatory Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.

## Predicate Devices

- Nuvasive Viewpoint ([K180198](/device/K180198.md))
- Orthofix Centurion ([K180025](/device/K180025.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 11, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Zavation Medical Products LLC Colby Williams Engineer 220 Lakeland Parkway Flowood, Mississippi 39232

Re: K211993

Trade/Device Name: Z-LINKpc System Regulation Number: 21 CFR 888.3075 Regulation Name: Posterior Cervical Screw System Regulatory Class: Class II Product Code: NKG, KWP Dated: June 25, 2021 Received: June 28, 2021

Dear Colby Williams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K211993

#### Device Name

Z-LINKpc System

#### Indications for Use (Describe)

The Z-LINKpc System implants are intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1-C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations: instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. These implants are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

To achieve additional levels of fixation, the Z-LINKsc System may be connected to the Zavation Spinal System using rod connectors and tapered rods.

When used with the occipital plate, the Z-LINKpc System is also intended to provide immobilization and stabilization for the occipito-cervico-thoracic juction (occiput-T3) in treatment of the instabilities mentioned above, including occipitocervical dislocation.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                         |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <div style="display:inline-block;"> <div style="display:inline-block;">                         ☒ Prescription Use (Part 21 CFR 801 Subpart D)                     </div> <div style="display:inline-block; margin-left: 20px;">                         ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                     </div> </div> </span> |

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

 This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

 The burden time for this collection of information is estimated to average 79 hours per response, including the
 time to review instructions, search existing data sources, gather and maintain the data needed and complete
 and review the collection of information. Send comments regarding this burden estimate or any other aspect
 of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   
Food and Drug Administration
   
Office of Chief Information Officer
   
Paperwork Reduction Act (PRA) Staff
   

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
 information unless it displays a currently valid OMB number."*

{3}------------------------------------------------

# 510K Summary

| Date:                      | June 25, 2021                                                                                                       |
|----------------------------|---------------------------------------------------------------------------------------------------------------------|
| Submitter:                 | Zavation Medical Products LLC<br>220 Lakeland Pkwy<br>Flowood, MS 39232<br>Phone: 601-919-1119<br>Fax: 800-447-1302 |
| Contact Person:            | Colby Williams                                                                                                      |
| Type of 510(k) submission: | Traditional                                                                                                         |
| Trade name:                | Z-LINKPC System                                                                                                     |
| Classification regulation: | 21 CFR 888.3075 Posterior Cervical Screw System<br>21 CFR 888.3050 Spinal Interlaminal Fixation Orthosis            |
| Device classification:     | Class II                                                                                                            |
| Classification Panel:      | Orthopedic Devices                                                                                                  |
| Product code:              | NKG, KWP                                                                                                            |
| Basis for submission:      | Addition of occipital plate.                                                                                        |

#### Device Description:

The Z-LINKsc System is a temporary, titanium alloy (Ti-6AL-4V ELI per ASTM F136) and cobalt chrome alloy (Co-28Cr-6Mo per ASTM F1537), multiple component system comprised of a variety of non-sterile, single use implantable components. The system consist of polyaxial screws, hooks, rods, cross-connectors, rod connectors, offsets, occipital plates, bone screws, and cap screws. The components are available in a variety of lengths and sizes in order to accommodate patient anatomy.

#### Indications for Use:

The Z-LINKoc System implants are intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1-C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. These implants are also intended to restore the integrity of the spinal column even in the absence of fusion for a

{4}------------------------------------------------

limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

To achieve additional levels of fixation, the Z-LINKrc System may be connected to the Zavation Spinal System using rod connectors and tapered rods.

When used with the Occipital Plate, the Z-LINKsc System is also intended to provide immobilization and stabilization for the occipito-cervico-thoracic junction (occiput-T3) in treatment of the instabilities mentioned above, including occipitocervical dislocation.

### Materials:

The Z-LINKec System components are manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136, and cobalt chromium alloy (Co-28Cr-6Mo) as described by ASTM F1537.

## Primary Predicate Device:

K180198 Nuvasive Viewpoint

### Additional Predicate Device:

K180025 Orthofix Centurion

### Technological Characteristics:

The Z-LINKoc system possesses the same technological characteristics as the predicate devices. These include basic design (rod based fixation system having occipital bone screws and various accessories), material (titanium alloy/cobalt chrome alloy), mechanical safety and performance, and intended use.

#### Performance Data:

Static compression bending and torsion, dynamic compression bending, and dynamic torsion tests were performed according to ASTM F2706 on a worst-case construct. The mechanical test results demonstrated the Z-LINKoc system performs as well as or better than the predicate devices.

#### Conclusion:

The Z-LINKsc System is similar to the predicate systems with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K211993](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K211993)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
