← Product Code [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG) · K192014

# SYPMHONY OCT System (K192014)

_Medos International SARL · NKG · Jan 24, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K192014

## Device Facts

- **Applicant:** Medos International SARL
- **Product Code:** [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG.md)
- **Decision Date:** Jan 24, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3075
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The SYMPHONY OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3): - · Traumatic spinal fractures and/or traumatic dislocations; - Instability or deformity; - · Failed previous fusions (e.g. pseudarthrosis); - · Tumors involving the cervical/thoracic spine; · Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and · Degenerative disease of the facets with instability. The SYMPHONY OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage turnors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The SYMPHONY OCT System is compatible with occipital fusion components (plates, rods and clamps) from the SYNAPSE Occipital-Cervical-Thoracic (OCT) System and the MOUNTAINEER OCT Spinal System. Additionally, the SYMPHONY OCT System is compatible with SYNAPSE OCT System hooks and rods. The SONGER Wire/Cable System may be used with the SYMPHONY OCT System to allow for wire/cable attachment to the posterior cervical spine. The SYMPHONY OCT System may be connected to the EXPEDIUM Spine System using connectors and tapered rods. The SYMPHONY OCT System can also be linked to the USS Spinal System and MATRIX Spine System using connectors and tapered rods.

## Device Story

SYMPHONY OCT System is a posterior spinal fixation system; provides immobilization/stabilization of spinal segments. Components include bone anchors (screws), longitudinal rods, interconnection mechanisms (set screws), and transverse connectors. Surgeon assembles implant system to fit patient anatomy. Used in OR by surgeons to treat spinal instabilities/deformities. System compatible with various DePuy Synthes OCT and spine systems (SYNAPSE, MOUNTAINEER, EXPEDIUM, VIPER, USS, MATRIX) via connectors/tapered rods. Provides mechanical stability to spinal column; facilitates fusion or restores integrity in palliative tumor cases.

## Clinical Evidence

Bench testing only. No clinical data presented. Substantial equivalence supported by design analysis and risk assessment of the new sterile tube packaging configuration.

## Technological Characteristics

Posterior spinal fixation system. Materials: Titanium alloy (ASTM F136) for polyaxial screws; Titanium alloy (ASTM F136) and Nitinol (ASTM F2063) for rod-to-rod cross connectors. System includes bone anchors, rods, set screws, and transverse connectors. Sterile tube packaging.

## Regulatory Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.

## Predicate Devices

- SYMPHONY OCT System ([K181949](/device/K181949.md), [K190895](/device/K190895.md))

## Reference Devices

- SYNAPSE Occipital-Cervical-Thoracic (OCT) System
- MOUNTAINEER OCT Spinal System
- SONGER Wire/Cable System
- EXPEDIUM Spine System
- VIPER System
- USS Spinal System
- MATRIX Spine System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 24, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.

Medos International SARL % Sheree Geller Regulatory Affairs Specialist DePuy Synthes Spine 325 Paramount Drive Raynham, Massachusetts 02767

Re: K192014

Trade/Device Name: SYPMHONY™ OCT System Regulation Number: 21 CFR 888.3075 Regulation Name: Posterior Cervical Screw System Regulatory Class: Class II Product Code: NKG, KWP Dated: December 13, 2019 Received: December 16, 2019

Dear Sheree Geller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

## K192014

Device Name SYMPHONY™ OCT System

## Indications for Use (Describe)

The SYMPHONY OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3):

- · Traumatic spinal fractures and/or traumatic dislocations;
- Instability or deformity;
- · Failed previous fusions (e.g. pseudarthrosis);
- · Tumors involving the cervical/thoracic spine;

· Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and

· Degenerative disease of the facets with instability.

The SYMPHONY OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage turnors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

The SYMPHONY OCT System is compatible with occipital fusion components (plates, rods and clamps) from the SYNAPSE Occipital-Cervical-Thoracic (OCT) System and the MOUNTAINEER OCT Spinal System. Additionally, the SYMPHONY OCT System is compatible with SYNAPSE OCT System hooks and rods.

The SONGER Wire/Cable System may be used with the SYMPHONY OCT System to allow for wire/cable attachment to the posterior cervical spine.

The SYMPHONY OCT System may be connected to the EXPEDIUM Spine System using connectors and tapered rods. The SYMPHONY OCT System can also be linked to the USS Spinal System and MATRIX Spine System using connectors and tapered rods.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY

| A. | Submitter Information   |                                                                                                                                                       |
|----|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
|    | Manufacturer:           | Medos International SARL<br>Chemin-Blanc 38<br>2400 Le Locle, Switzerland                                                                             |
|    | Submitter:              | DePuy Synthes Spine<br>325 Paramount Drive<br>Raynham, MA 02767                                                                                       |
|    | Contact Person:         | Sheree Geller<br>325 Paramount Drive<br>Raynham, MA 02767<br>(508) 828-3291<br>(508) 828-3797<br>sgeller1@its.jnj.com                                 |
| B. | Date Prepared           | July 25, 2019                                                                                                                                         |
| C. | Device Name             |                                                                                                                                                       |
|    | Trade/Proprietary Name: | SYMPHONY™ OCT System                                                                                                                                  |
|    | Common/Usual Name:      | Posterior Cervical Screw System                                                                                                                       |
|    | Regulatory Class:       | II                                                                                                                                                    |
|    | Review Panel:           | Orthopedic                                                                                                                                            |
|    | Product Codes:          | NKG – Class II – 21 CFR §888.3075<br>Posterior Cervical Screw System<br>KWP – Class II – 21 CFR §888.3050<br>Appliance, Fixation, Spinal Interlaminal |
| D. | Predicate Device Names  |                                                                                                                                                       |
|    | Primary Predicate:      | SYMPHONY OCT System (K181949, K190895)                                                                                                                |

{4}------------------------------------------------

#### E. Device Description

The SYMPHONY OCT System is a posterior spinal fixation system intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion of the craniocervical junction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3). The system is composed of multiple components to allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. The system consists of bone anchors (such as screws) for connection by longitudinal components (such as rods) via an interconnection mechanism (e.g., set screws) with optional transverse connectors (e.g., cross connectors) to link the longitudinal components for additional stability.

#### Intended Use F.

The SYMPHONY OCT System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3):

- · Traumatic spinal fractures and/or traumatic dislocations;
- Instability or deformity;
- · Failed previous fusions (e.g. pseudarthrosis);
- · Tumors involving the cervical/thoracic spine;
- · Degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and
- · Degenerative disease of the facets with instability.

The SYMPHONY OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

The SYMPHONY OCT System is compatible with occipital fusion components (plates, rods and clamps) from the SYNAPSE Occipital-Cervical-Thoracic (OCT) System and the MOUNTAINEER OCT Spinal System. Additionally, the SYMPHONY OCT System is compatible with SYNAPSE OCT System hooks and rods.

The SONGER Wire/Cable System may be used with the SYMPHONY OCT System to allow for wire/cable attachment to the posterior cervical spine.

The SYMPHONY OCT System may be connected to the EXPEDIUM® Spine System and VIPER® System using connectors and tapered rods. The SYMPHONY OCT System can also be linked to the USS Spinal System and MATRIX Spine System using connectors and tapered rods.

{5}------------------------------------------------

### G. Summary of Similarities and Differences in Technological Characteristics, Performance, and Intended Use

The fundamental scientific technology of the subject devices remains unchanged from the predicate devices. This submission presents a sterile tube packaging configuration for polyaxial screw components of the SYMPHONY OCT System. The intended use, technological characteristics, and performance of the SYMPHONY OCT System are equivalent with those of the predicate devices.

#### H. Materials

The SYMPHONY OCT System subject device polyaxial screw components are comprised of Titanium alloy conforming to ASTM F136. The Rod-to-Rod Cross Connectors are comprised of both Titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136 and Nitinol conforming to ASTM F2063.

#### I. Performance Data

Performance data remains unchanged for the subject devices as compared to the predicate devices. Risk analysis was completed for the addition of a sterile tube packaging configuration.

#### J. Conclusion

Evaluation of the subject device intended use, technological characteristics, and performance data demonstrates substantial equivalence with the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K192014](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K192014)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
