← Product Code [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG) · K183383

# ANAX OCT Spinal System (K183383)

_U&I Corporation · NKG · Dec 12, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K183383

## Device Facts

- **Applicant:** U&I Corporation
- **Product Code:** [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG.md)
- **Decision Date:** Dec 12, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3075
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

ANAX™ OCT Spinal System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. ANAX™ OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, ANAX™ OCT Spinal System may be connected to Perfix™ Spinal Sytem and ANAX™ 5.5 Spinal System rods with the rod connectors. Transition rods with differing diameters may also be used to connect ANAX™ OCT Spinal System to Perfix™ Spinal System and ANAX™ 5.5 Spinal System.

## Device Story

ANAX™ OCT Spinal System is a posterior spinal fixation system used by surgeons to stabilize spinal segments and promote fusion. Components include polyaxial screws, polyaxial shank screws, hooks, rods, set screws, transverse links, and occipital plates. The system allows for construct assembly to immobilize the craniocervical junction, cervical spine, and upper thoracic spine. It can connect to Perfix™ and ANAX™ 5.5 systems via rod connectors. The device is implanted during surgery; it provides mechanical support to the spinal column. Benefits include stabilization of unstable segments and restoration of spinal integrity in tumor patients. No electronic or software components are involved.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical bench testing and mechanical strength comparisons to the predicate device.

## Technological Characteristics

Constructed from medical grade titanium alloy (ASTM F136). Components include polyaxial screws (self-tapping), hooks, rods, set screws, transverse links, and occipital plates. Mechanical fixation system; non-sterile, single-use. No software, electronics, or energy sources.

## Regulatory Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.

## Predicate Devices

- ANAX™ OCT Spinal System ([K150570](/device/K150570.md))

## Reference Devices

- Perfix™ Spinal System
- ANAX™ 5.5 Spinal System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is a circular seal with an emblem in the center. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 12, 2018

U & I Corporation Ms. Young-Eun Lee Regulatory Affairs Specialist 20. Sandan-ro 76beon-gil (Road) Uijeongbu-si, Gyeonggi-do 11781 KOREA

Re: K183383

Trade/Device Name: ANAX™ OCT Spinal System Regulatory Class: Unclassified Product Code: NKG, KWP Dated: December 3, 2018 Received: December 6, 2018

Dear Ms. Lee:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products, and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ronald P. Jean -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K183383

Device Name ANAXTM OCT Spinal System

#### Indications for Use (Describe)

ANAX™ OCT Spinal System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. ANAX™ OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

In order to achieve additional levels of fixation, ANAX™ OCT Spinal System may be connected to Perfix™ Spinal Sytem and ANAX™ 5.5 Spinal System rods with the rod connectors. Transition rods with differing diameters may also be used to connect ANAX™ OCT Spinal System to Perfix™ Spinal System and ANAX™ 5.5 Spinal System.

| Type of Use (Select one or both, as applicable)                                                                    |                                                                                                                   |
|--------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 6. 510(k) Summary

| Manufacturer:        | U & I Corporation<br>20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,<br>11781, Korea, |
|----------------------|------------------------------------------------------------------------------------------------|
| Sponsor:             | U & I Corporation<br>20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,<br>11781, Korea, |
| Sponsor Contact:     | Young-Eun Lee, RA Specialist<br>+82 31 860 6927<br>youngeun@youic.com                          |
| Date Prepared:       | December 03, 2018                                                                              |
| Trade Name:          | ANAX™ OCT Spinal System                                                                        |
| Classification Name: | Spinal Interlaminal Fixation Orthosis                                                          |
| Classification:      | Unclassified                                                                                   |
| Common Name:         | Occipito-cervico-thoracic Spinal Fixation System                                               |
| Product Code:        | NKG, KWP                                                                                       |
| Predicate Device:    | ANAX™ OCT Spinal System (K150570)                                                              |

### Description of Device:

The ANAX™ OCT Spinal System is manufactured by U&I corporation. The ANAX™ OCT Spinal System is for fixation the cervicocranium (Occiput/C2), the true subaxial region (C3/C6), and the cervicothoracic junction (C7 to T2) by one system. The ANAX™ OCT Spinal System consists of polyaxial screws, polyaxial shank screws, hooks, rods, set screws, transverse (cross) links and occipital plate. Connectors are also provided for surgical convenience. The ANAX™ OCT Spinal System allows surgeons to build a spinal implant construct to stabilize and promote spinal fusion. The single-use ANAXTM OCT Spinal System components are supplied as nonsterile and are fabricated from medical grade titanium alloy (ASTM F136). All polyaxial screws have self-tapping function in the ANAXTM OCT Spinal System.

Specialized instruments made from surgical instrument grade stainless steel are available for the application and removal of the ANAXTM OCT Spinal System implants.

### Indications for Use:

ANAX™ OCT Spinal System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine(C1 to C7) and the thoracic spine from T1-T3:

{4}------------------------------------------------

traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g.pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. ANAX™OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, ANAX™ OCT Spinal System may be connected to Perfix™ Spinal System and ANAX™ 5.5 Spinal System rods with the rod connectors. Transition rods with differing diameters may also be used to connect ANAX™ OCT Spinal System to Perfix™ Spinal System and ANAX™ 5.5 Spinal System.

### Substantial Equivalence:

ANAX™ OCT Spinal System is substantially equivalent to ANAX™ OCT Spinal System (K150570) in design, material, mechanical performance, function and intended use.

The mechanical performance of the ANAX™ OCT Spinal System, made of Ti Alloy, met the acceptance criteria which have been established from the predicate devices.

- 1. Comparison Technological Characteristics
The predicate and proposed devices have the same intended use and basic fundamental scientific technology and share the following similarities;

- The same indications for use
- Similar design features
- Incorporate the same or similar materials
- The equivalent mechanical performance

# 2. Performance Testing

Any mechanical test in non clinical setting (bench testing) was not conducted on the ANAX™ OCT Spinal System. However, the mechanical strength evaluation was conducted to compare data of proposed device of the ANAX™ OCT Spinal System(K150570) and to verify there are no new safety and effectiveness issues were not raised by the proposed device. The data met all acceptance criteria and that verifies performance of the ANAXIM OCT Spinal System is substantially equivalent to predicate devices.

# 3. Conclusion

The data and information provided in this submission support the conclusion that the ANAX™ OCT Spinal System is substantially equivalent to predicate devices with respect to indications for use and technological characteristics.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K183383](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K183383)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
