← Product Code [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG) · K160756

# PASS OCT Spinal System (K160756)

_Medicrea International SA · NKG · Apr 26, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K160756

## Device Facts

- **Applicant:** Medicrea International SA
- **Product Code:** [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG.md)
- **Decision Date:** Apr 26, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3075
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The PASS OCT Spinal System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radios, and degenerative disease of the facets with instability. The PASS OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the PASS OCT Spinal System may be connected to the PASS LP Spinal System rods with the dual diameter rods or dominos. Refer to the PASS LP Spinal System package insert for a list of the PASS LP Spinal System indications of use.

## Device Story

PASS OCT Spinal System is a posterior spinal fixation system consisting of rods, hooks, polyaxial screws, occipital plates, and connection components. This submission introduces axial and lateral domino connectors to allow connection to PASS LP Spinal System rods. System provides immobilization and stabilization of spinal segments to augment fusion or restore spinal column integrity in tumor cases. Used in surgical settings by orthopedic or neurosurgeons. Implants are rigidly locked in various configurations to provide mechanical stability. Benefits include stabilization of spinal segments and support for fusion in patients with spinal instability or deformity.

## Clinical Evidence

No clinical data or published literature were used to demonstrate substantial equivalence. Bench testing only.

## Technological Characteristics

Materials: Ti-6Al-4V (ASTM F136, ISO 5832-3), CP Titanium (ASTM F67, ISO 5832-2), PEEK OPTIMA LT1 (ASTM F2026), Co-Cr28Mo6 (ISO 5832-12, ASTM F1537). Components: polyaxial screws, hooks, occipital plates, rods (3.5mm), and domino connectors. Mechanical testing performed per ASTM F2706-08.

## Regulatory Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.

## Predicate Devices

- PASS OCT Spinal System ([K150918](/device/K150918.md))
- Vertex Reconstruction System ([K143471](/device/K143471.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, interconnected design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.

April 26, 2016

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Medicrea International S.A. Mr. David Ryan VP Product Development and Marketing 14 Porte du Grand Lyon Neyron 01700 FRANCE

Re: K160756

Trade/Device Name: PASS OCT Spinal System Regulation Number: Unclassified Product Code: NKG, KWP Dated: March 16, 2016 Received: March 18, 2016

Dear Mr. Ryan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160756

Device Name PASS OCT Spinal System

#### Indications for Use (Describe)

The PASS OCT Spinal System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radios, and degenerative disease of the facets with instability. The PASS OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

In order to achieve additional levels of fixation, the PASS OCT Spinal System may be connected to the PASS LP Spinal System rods with the dual diameter rods or dominos. Refer to the PASS LP Spinal System package insert for a list of the PASS LP Spinal System indications of use.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subpart C) |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) SUMMARY

#### MEDICREA INTERNATIONAL's PASS OCT additional components

In accordance with 21 CFR 807.92 of the Federal Code of Requlations the following 510(k) summary is submitted for the PASS OCT Spinal System- Additional Components:

#### Date Prepared: 26 April 2016

#### 1. Submitter: Contact Person: MEDICREA INTERNATIONAL S.A. David RYAN 14 Porte du Grand Lyon MEDICREA INTERNATIONAL S.A. NEYRON 01700 14 Porte du Grand Lyon FRANCE NEYRON 01700 Phone Number : +33 4 72 01 87 87 FRANCE

- PASS OCT Spinal System 2. Trade name:
#### Regulatory Identification/ Classification

Orthosis, Cervical pedicle screw spinal fixation Product Code: NKG Unclassified, Pre-Amendment

Spinal Interlaminal Fixation Orthosis Regulation Number: 21CFR 888.3050 Product Code: KWP Class II

#### 3. Predicate or legally marketed devices which are substantially equivalent:

#### Primary predicate:

- PASS OCT Spinal System, (MEDICREA INTERNATIONAL, K150918) .
#### Additional predicate:

- Vertex Reconstruction System, (MEDTRONIC SOFAMOR DANEK USA; K143471) ●
No reference devices were used in this submission.

#### 4. Description of the device:

The purpose of this submission was to introduce axial and lateral domino connectors to the system.

The PASS OCT Spinal System is a posterior system, which consists of a variety of shapes and sizes of rods, hooks, polyaxial screws, occipital plates, occipital bone screws, and connection components, which can be rigidly locked to the rod in a variety of configurations. See package insert of the system for labeling limitations.

{4}------------------------------------------------

The implants are manufactured in titanium alloy Ti-6AI-4V ELI conforming to ISO 5832-3 specifications and ASTM F136 specifications, in PEEK OPTIMA LT1 conforming to ASTM F2026 specifications and in cobalt-chromium molybdenum alloy Co-Cr28Mo6 that conforms to ISO 5832-12 and ASTM F1537 specifications, and also CP Titanium according to ASTM F67 and ISO 5832-2 specifications.

#### Never use stainless steel and titanium implant components in the same construct.

To achieve best results, do not use any of the PASS OCT Spinal System implant components with components from any other systems or manufacturer unless specifically labeled to do so in this or another MEDICREA® INTERNATIONAL document.

MATERIALS: Titanium Alloy (Ti-6Al-4V) according to the ASTM F136-11 & ISO 5832-30; CP Titanium according to the ASTM F67 and ISO 5832-2; PEEK OPTIMA LT1® according to the ASTM F2026-10 and cobalt-chromium molybdenum alloy Co-Cr28Mo6 according to ISO5832-12 and ASTM F1537.

Function: The PASS OCT spinal system was developed as an implant:

- · To provide immobilization and stabilization of posterior spinal segments
- · to auqment the development of a solid spinal fusion
- to provide stability to ease fusion
- to be mechanically resistant to allow the fusion of the operated level

#### 5. Indication for Use

The PASS OCT Spinal System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy. neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The PASS OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

In order to achieve additional levels of fixation, the PASS OCT Spinal System may be connected to the PASS LP Spinal System rods with the dual diameter rods or dominos. Refer to the PASS LP Spinal System package insert for a list of the PASS LP Spinal System indications of use.

#### 6. Substantial equivalence claimed to predicate devices

The PASS OCT Spinal system components are technologically similar to the already cleared PASS OCT Spinal System Components and Vertex Reconstruction System Components in terms of intended use, materials used, mechanical safety and performances

- . PASS OCT Spinal System, (MEDICREA INTERNATIONAL, K150918)
- Vertex Reconstruction System. (MEDTRONIC SOFAMOR DANEK USA: K143471) .

{5}------------------------------------------------

The table below compares the features and characteristics of PASS OCT components to their predicate devices.

| Device                                                        | MEDCREA<br>INTERNATIONAL<br>PASS OCT Spinal<br>System<br>NEW COMPONENT                                                                                                                                               | MEDCREA<br>INTERNATIONAL<br>PASS OCT Spinal<br>System                                                                                                                                                                | MEDTRONIC SOFAMOR<br>DANEK<br>Vertex Reconstruction<br>System                                                                                                                        |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number                                                 | Unknown                                                                                                                                                                                                              | K150918                                                                                                                                                                                                              | K143471                                                                                                                                                                              |
| Intended use                                                  |                                                                                                                                                                                                                      |                                                                                                                                                                                                                      |                                                                                                                                                                                      |
| Occipital                                                     | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                  |
| Cervical                                                      | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                  |
| Thoracic                                                      | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                  |
| Components                                                    |                                                                                                                                                                                                                      |                                                                                                                                                                                                                      |                                                                                                                                                                                      |
| Anchorage<br>means                                            | -Polyaxial screws<br>- Polyaxial hooks<br>- Occipital bone screws                                                                                                                                                    | -Polyaxial screws<br>- Polyaxial hooks<br>- Occipital bone screws                                                                                                                                                    | - Multi-axial screws<br>- Hooks<br>- Occipital bone screws                                                                                                                           |
| Rods diameter                                                 | Ø3.5 mm                                                                                                                                                                                                              | Ø3.5 mm                                                                                                                                                                                                              | Ø3.5mm and Ø3.2mm                                                                                                                                                                    |
| Straight rods                                                 | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                  |
| Pre-bent rods                                                 | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                  | Yes                                                                                                                                                                                  |
| Lateral<br>Domino<br>between<br>Thoracic and<br>cervical rods | Yes                                                                                                                                                                                                                  | No                                                                                                                                                                                                                   | Yes                                                                                                                                                                                  |
| Axial<br>Domino<br>between<br>Thoracic and<br>cervical rods   | Yes                                                                                                                                                                                                                  | No                                                                                                                                                                                                                   | Yes                                                                                                                                                                                  |
| Materials                                                     |                                                                                                                                                                                                                      |                                                                                                                                                                                                                      |                                                                                                                                                                                      |
|                                                               | Titanium Alloy (Ti-6Al-4V) according to ASTM F136 & ISO 5832-3<br>- Grade Titanium according to ASTM F67-13 & ISO 5832-2<br>PEEK OPTIMA LT1 conforming to ASTM F2026<br>Co-Cr 28Mo6 alloy 1 (following the ASTMF1537 | Titanium Alloy (Ti-6Al-4V) according to ASTM F136 & ISO 5832-3<br>- Grade Titanium according to ASTM F67-13 & ISO 5832-2<br>PEEK OPTIMA LT1 conforming to ASTM F2026<br>Co-Cr 28M06 alloy 1 (following the ASTMF1537 | Titanium Alloy (Ti-6Al-4V) according to ASTM F136 & ISO 5832-3<br>- Grade Titanium according to ASTM F67 & ISO 5832-2<br>- Shape Memory Alloy (Nitinol-NiTi) according to ASTM F2063 |

### 7. Non-clinical Test Summary:

The components of the PASS OCT spinal were mechanically evaluated in static and dynamic compression and static torsion tests following the ASTM F2706-08.

{6}------------------------------------------------

## 8. Clinical Test Summary

No clinical data or published literature were used to demonstrate substantial equivalence.

# 9. Conclusions Non clinical and Clinical

The PASS OCT Spinal system components are substantially equivalent to legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K160756](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K160756)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
