← Product Code [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG) · K153386

# Perla Posterior Cervico-Thoracic Fixation system (K153386)

_Spineart · NKG · Jan 25, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K153386

## Device Facts

- **Applicant:** Spineart
- **Product Code:** [NKG](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG.md)
- **Decision Date:** Jan 25, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3075
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The PERLA® posterior cervico-thoracic fixation system is intended to provide immobilization of spinal segments as an adjunct to following acute and chronic instabilities of the cervical spine (C 1 to C7) and the thoracic spine from TI-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radios, and degenerative disease of the facets with instability. The PERLA® posterior cervico-thoracic fixation system is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the PERLA® posterior cervico-thoracic fixation system may be connected to the ROMEO® Posterior Osteosynthesis System with rod connectors. Transition rods may also be used to connect the PERLA® posterior cervico-thoracic fixation system to the ROMEO® Posterior Osteosynthesis System. Refer to the ROMEO® Posterior Osteosynthesis System package insert for a list of the ROMEO® Posterior Osteosynthesis System indications of use.

## Device Story

PERLA® is a posterior cervico-thoracic fixation system; consists of rods, hooks, multi-axial screws, set screws, and connectors. Used by surgeons in clinical settings to immobilize and stabilize spinal segments (C1-T3). Components are rigidly locked in configurations adapted to individual patient anatomy. System can connect to ROMEO® Posterior Osteosynthesis System via rod connectors or transition rods. Provides mechanical stabilization to promote fusion or restore spinal integrity in tumor patients. Benefits include spinal stabilization and deformity correction.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified via static compression bending, static torsion, dynamic compression bending (ASTM F1717), static flexion-extension, static axial/torsion gripping (ASTM F1798), and axial pullout/torque to failure (ASTM F543).

## Technological Characteristics

Materials: Medical grade titanium alloy (ASTM F136) and cobalt chromium (ASTM F1537). Components: Multi-axial screws (3.5mm/4.0mm diameter, 8-52mm length), rods (3.5mm diameter, straight/pre-bent), transition rods (3.5/5.4mm), hooks, and connectors. Sterilization: Gamma radiation. Connectivity: Mechanical assembly; no electronic or software components.

## Regulatory Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.

## Predicate Devices

- Vertex Reconstruction System ([K143471](/device/K143471.md))

## Reference Devices

- Vertex Reconstruction System ([K003780](/device/K003780.md))
- Synthes Cervifix/Axon ([K023675](/device/K023675.md))
- Synthes Cervifix System ([K991089](/device/K991089.md))
- Mountaineer OCT Spinal System ([K110353](/device/K110353.md))
- ROMEO®2 spinal system ([K151695](/device/K151695.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 25, 2016

SPINEART Mr. Franck Pennesi Director of Industry & Quality International Center Cointrin 20 route de pré-bois - CP 1813 1215 Geneva 15 SWITZERLAND

Re: K153386

Trade/Device Name: PERLA® Posterior Cervico-Thoracic Fixation System Regulatory Class: Unclassified Product Code: NKG, KWP Dated: November 19, 2015 Received: November 23, 2015

Dear Mr. Pennesi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K153386

#### Device Name

PERLA® Posterior Cervico-Thoracic Fixation System

## Indications for Use (Describe)

The PERLA® posterior cervico-thoracic fixation system is intended to provide immobilization of spinal segments as an adjunct to following acute and chronic instabilities of the cervical spine (C 1 to C7) and the thoracic spine from TI-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radios, and degenerative disease of the facets with instability.

The PERLA® posterior cervico-thoracic fixation system is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

In order to achieve additional levels of fixation, the PERLA® posterior cervico-thoracic fixation system may be connected to the ROMEO® Posterior Osteosynthesis System with rod connectors. Transition rods may also be used to connect the PERLA® posterior cervico-thoracic fixation system to the ROMEO® Posterior Osteosynthesis System. Refer to the ROMEO® Posterior Osteosynthesis System package insert for a list of the ROMEO® Posterior Osteosynthesis System indications of use.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## TRADITIONAL 510k PERLA® POSTERIOR CERVICO-THORACIC FIXATION SYSTEM

Image /page/3/Picture/2 description: The image shows the logo for Spineart. The logo features the word "spineart" in a stylized font, with the "art" portion in a lighter shade of purple. Above and to the right of the text is a purple butterfly-like design, adding a touch of elegance to the overall design.

# 510(k) SUMMARY

|                                       | SPINEART                                                                          |
|---------------------------------------|-----------------------------------------------------------------------------------|
| Submitted by                          | International Center Cointrin                                                     |
|                                       | 20 route de pré-bois                                                              |
|                                       | CP1813                                                                            |
|                                       | 1215 GENEVA 15                                                                    |
|                                       | SWITZERLAND                                                                       |
| Contacts                              | Franck PENNESI Director of Industry & Quality                                     |
|                                       | Phone : +41 22 570 1246 Fax : +41 22 799 40 26                                    |
|                                       | Mail : fpennesi@spineart.com                                                      |
|                                       | Regulatory contact : Dr Isabelle DRUBAIX (Idée Consulting)                        |
|                                       | Mail : idrubaix@nordnet.fr                                                        |
| Date Prepared                         | January 14th 2016                                                                 |
| Trade Name                            | PERLA® Posterior Cervico-Thoracic Fixation System                                 |
| Common Name                           | Orthosis, Cervical Pedicle Screw Spinal Fixation                                  |
|                                       | Posterior Cervico-Thoracic Fixation system                                        |
| Classification Name                   | Appliance, Fixation, Spinal Interlaminal                                          |
| Product Code / Class<br>CFR section   | Primary: NKG, Unclassified                                                        |
|                                       | CFR Section: Unclassified                                                         |
|                                       | Subsequent: KWP, Class II                                                         |
|                                       | CFR Section: 888.3050 Spinal Interlaminal fixation orthosis                       |
| Device panel                          | Orthopedic                                                                        |
| Legally marketed<br>predicate devices | Primary predicate: Vertex Reconstruction System (K143471) manufactured by         |
|                                       | Medtronic Sofamor Danek USA, Inc                                                  |
|                                       | Reference devices: Vertex Reconstruction System (K003780) manufactured by         |
|                                       | Medtronic Sofamor Danek, Inc Synthes Cervifix/Axon (K023675) manufactured         |
|                                       | by Synthes (USA); Synthes Cervifix System (K991089) manufactured by Synthes       |
|                                       | Spine; Mountaineer OCT Spinal System (K110353) manufactured by Depuy              |
|                                       | Spine, Inc                                                                        |
| Indications for use                   | The PERLA® posterior cervico-thoracic fixation system is intended to provide      |
|                                       | immobilization and stabilization of spinal segments as an adjunct to fusion for   |
|                                       | the following acute and chronic instabilities of the cervical spine (C 1 to C7)   |
|                                       | and the thoracic spine from TI-T3: traumatic spinal fractures and/or traumatic    |
|                                       | dislocations; instability or deformity; failed previous fusions (e.g.,            |
|                                       | pseudarthrosis); tumors involving the cervical spine; and degenerative disease,   |
|                                       | including intractable radiculopathy and/or myelopathy, neck and/or arm pain       |
|                                       | of discogenic origin as confirmed by radiographic studies, and degenerative       |
|                                       | disease of the facets with instability.                                           |
|                                       | The PERLA® posterior cervico-thoracic fixation system is also intended to         |
|                                       | restore the integrity of the spinal column even in the absence of fusion for a    |
|                                       | limited time period in patients with advanced stage tumors involving the          |
|                                       | cervical spine in whom life expectancy is of insufficient duration to permit      |
|                                       | achievement of fusion.                                                            |
|                                       | In order to achieve additional levels of fixation, the PERLA® posterior cervico-  |
|                                       | thoracic fixation system may be connected to the ROMEO® Posterior                 |
|                                       | Osteosynthesis System with rod connectors. Transition rods may also be used       |
|                                       | to connect the PERLA® posterior cervico-thoracic fixation system to the           |
|                                       | ROMEO® Posterior Osteosynthesis System. Refer to the ROMEO® Posterior             |
|                                       | Osteosynthesis System package insert for a list of the ROMEO® Posterior           |
|                                       | Osteosynthesis System indications of use.                                         |
| Description<br>of<br>the<br>device    | The Spineart Perla® system is a posterior cervico-thoracic fixation system        |
|                                       | intended to provide stabilization to promote fusion of the cervical spine and     |
|                                       | the upper thoracic spine. Perla® system consists of a variety of shapes and       |
|                                       | sizes of rods, hooks, multi-axial screws, set screws, rod connectors and          |
|                                       | transverse connectors. These connecting components can be rigidly locked to       |
|                                       | the rod in a variety of configurations to be adapted for the individual case. The |
|                                       | Perla® system can also be linked to the cleared Spineart Romeo®2 spinal           |
|                                       | system (K151695) using the specific Perla® Axial and Parallel Rod to Rod          |
|                                       | connectors 3.5/5.4mm or the Perla® Transition Rods 3.5/5.4mm which are            |
|                                       | part of this submission                                                           |
| Technological<br>Characteristics      | The multi-axial screws are available with diameters 3.5mm and 4.0mm with          |
|                                       | lengths ranging from 8 mm up to 52mm.                                             |
|                                       | The 3.5mm diameter rods come in various lengths and have straight or pre-         |
|                                       | bent designs. The Perla® system comprises also transition rods with dual          |
|                                       | diameters of 3.5/5.4mm, a range of hooks and connectors.                          |
|                                       | The implantable components of the PERLA® posterior cervico-thoracic fixation      |
|                                       | system are made of medical grade titanium alloy and cobalt chromium               |
|                                       | conforming to standards ASTM F136 and F1537, respectively.                        |
|                                       | The implantable components of the PERLA® posterior cervico-thoracic fixation      |
|                                       | system are delivered sterile (gamma sterilization).                               |
|                                       | The PERLA® posterior cervico-thoracic fixation system is supplied with all the    |
|                                       | surgical instruments required for its installation.                               |
|                                       | Published literature and bench testing per ASTM F1717 demonstrate that the        |
| Performance Data                      | PERLA® posterior cervico-thoracic fixation system is substantially equivalent to  |
|                                       | the predicate devices.                                                            |
|                                       | The following non-clinical tests were conducted:                                  |
|                                       | ✓ Static Compression Bending, Static Torsion and Dynamic Compression<br>          |
|                                       | Bending according to ASTM F1717                                                   |
|                                       | ✓ Static flexion-extension testing, Static axial gripping and Static torsion      |
|                                       | gripping according to ASTM F1798                                                  |
|                                       | ✓ Axial pullout strength and Torque to failure according to ASTM F543             |
|                                       | Design comparisons and non-clinical performance testing demonstrate that          |
| Conclusion                            | the PERLA® posterior cervico-thoracic fixation system is substantially            |
|                                       | equivalent to its predicate devices in terms of intended use, material, design,   |
|                                       | mechanical properties and function.                                               |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K153386](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKG/K153386)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
